Data Quality Reports for Session: 116900 User: sherman Completed: 01/08/2009


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D040127.3TWP/SONDE/C2 - Broken temperature sensorstwpsondewnpnC2.b1
D040921.1TWP/SONDE/C2 - RH sensor failure below 9kmtwpsondewnpnC2.b1
D040921.2TWP/SONDE/C2 - Bad RH values alofttwpsondewnpnC2.b1
D050913.3TWP/SONDE/C2 - Bad RH sensortwpsondewnpnC2.b1
D051122.2TWP/SONDE/C2 - Broken temperature sensortwpsondewnpnC2.b1
D051219.2TWP/SONDE/C2 - Broken temperature sensortwpsondewnpnC2.b1


DQRID : D040127.3
Start DateStart TimeEnd DateEnd Time
10/13/2003004010/13/20030203
11/08/2003233511/09/20030215
11/18/2003120511/18/20031340
11/19/2003115611/19/20031346
11/21/2003004411/21/20030113
12/20/2003122512/20/20031408
12/21/2003005612/21/20030126
12/21/2003115512/21/20031344
12/23/2003004212/23/20030212
12/23/2003113712/23/20031259
12/23/2003234212/24/20030132
01/08/2004113501/08/20041309
01/09/2004115701/09/20041342
01/09/2004135901/09/20041537
01/12/2004234201/13/20040214
01/16/2004140901/16/20041548
01/16/2004233701/17/20040109
01/18/2004113501/18/20041256
01/19/2004022301/19/20040346
01/19/2004232901/20/20040157
01/21/2004125201/21/20041400
01/22/2004005201/22/20040227
01/23/2004113701/23/20041319
01/26/2004120201/26/20040146
02/17/2004140602/17/20041536
02/26/2004122102/26/20041350
03/09/2004114503/09/20041315
03/14/2004113103/14/20041300
03/22/2004113403/22/20041300
03/29/2004233203/30/20040047
03/30/2004233403/31/20040059
04/11/2004234804/12/20040120
05/01/2004233505/02/20040110
05/02/2004233605/03/20040109
05/13/2004233405/14/20040112
05/14/2004140605/14/20041411
06/07/2004022406/07/20040351
06/25/2004114606/25/20041328
07/02/2004005907/02/20040217
07/20/2004022007/20/20040358
07/24/2004123507/24/20041355
08/22/2004114408/22/20041207
09/08/2004142509/08/20041605
09/08/2004233809/08/20042345
10/18/2004234610/19/20040128
11/29/2004232111/30/20040105
12/08/2004114312/08/20041306
12/30/2004113012/30/20041239
Subject:
TWP/SONDE/C2 - Broken temperature sensors
DataStreams:twpsondewnpnC2.b1
Description:
The temperature sensors on these radiosondes broke after launch.  All
temperature-related data are incorrect.
Measurements:twpsondewnpnC2.b1:
  • Relative humidity inside the instrument enclosure(rh)
  • Dry bulb temperature(tdry)
  • Surface dew point temperature(dp)


Back To Table of Contents

DQRID : D040921.1
Start DateStart TimeEnd DateEnd Time
09/16/2004113509/16/20041210
Subject:
TWP/SONDE/C2  - RH sensor failure below 9km
DataStreams:twpsondewnpnC2.b1
Description:
It appears that the RH sensor heating circuit failed during this sounding.  The RH 
oscillates rapidly between low (~0 %RH) and not so low (ambient %RH) and the oscillation seems 
to stop when the temperature is below ~35 degC.  According to Vaisala these are symptoms 
of failure in the ASIC (application specific integrated circuit) that controls the heating.
Measurements:twpsondewnpnC2.b1:
  • Relative humidity inside the instrument enclosure(rh)
  • Surface dew point temperature(dp)


Back To Table of Contents

DQRID : D040921.2
Start DateStart TimeEnd DateEnd Time
09/17/2004114509/17/20041216
09/18/2004005809/18/20040133
09/19/2004140809/19/20041447
Subject:
TWP/SONDE/C2 - Bad RH values aloft
DataStreams:twpsondewnpnC2.b1
Description:
It appears that the RH sensor heating circuit on these sondes failed after launch but 
resume proper operation when the ambient temperature dropped below ~40 degC.
Measurements:twpsondewnpnC2.b1:
  • Relative humidity inside the instrument enclosure(rh)
  • Surface dew point temperature(dp)


Back To Table of Contents

DQRID : D050913.3
Start DateStart TimeEnd DateEnd Time
07/03/2005235007/04/20050024
07/10/2005234007/11/20050021
07/19/2005233207/20/20050016
Subject:
TWP/SONDE/C2 - Bad RH sensor
DataStreams:twpsondewnpnC2.b1
Description:
It appears that the RH sensor heating circuit failed during this sounding.  The RH 
oscillates rapidly between low (~0 %RH) and not so low (ambient %RH) and the oscillation seems 
to stop when the temperature is below -50 degC.  According to Vaisala these are symptoms 
of failure in the ASIC (application specific integrated circuit) that controls the heating.
Measurements:twpsondewnpnC2.b1:
  • Relative humidity inside the instrument enclosure(rh)
  • Surface dew point temperature(dp)


Back To Table of Contents

DQRID : D051122.2
Start DateStart TimeEnd DateEnd Time
11/19/2005233411/20/20050153
11/21/2005233211/22/20050140
Subject:
TWP/SONDE/C2 - Broken temperature sensor
DataStreams:twpsondewnpnC2.b1
Description:
The temperature sensor on this radiosonde broke after launch.  All the temperature-related 
data are incorrect.
Measurements:twpsondewnpnC2.b1:
  • Relative humidity inside the instrument enclosure(rh)
  • Dry bulb temperature(tdry)
  • Surface dew point temperature(dp)


Back To Table of Contents

DQRID : D051219.2
Start DateStart TimeEnd DateEnd Time
12/07/2005115312/07/20051323
Subject:
TWP/SONDE/C2  - Broken temperature sensor
DataStreams:twpsondewnpnC2.b1
Description:
The temperature sensor on this radiosonde broke after launch.  All the temperature-related 
data are incorrect.
Measurements:twpsondewnpnC2.b1:
  • Relative humidity inside the instrument enclosure(rh)
  • Dry bulb temperature(tdry)
  • Surface dew point temperature(dp)


Back To Table of Contents



END OF DATA