Data Quality Reports for Session: 116849 User: koontz Completed: 01/07/2009


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D050517.2SGP/MFRSR/E24 - False signalsgpmfrsrE24.00, sgpmfrsrE24.a0, sgpmfrsrE24.b1


DQRID : D050517.2
Start DateStart TimeEnd DateEnd Time
02/01/2005000006/25/20050000
Subject:
SGP/MFRSR/E24 - False signal
DataStreams:sgpmfrsrE24.00, sgpmfrsrE24.a0, sgpmfrsrE24.b1
Description:
It looks like an irradiance signal is coming through in the voltage data.  It has also 
been apparent that an inverse relationship in the temp and voltage data exists.  Several 
people have looked at this and there doesn't seem to be an adverse effect on the irradiance 
data, so this is being recorded as a transparent DQR.
Measurements:sgpmfrsrE24.b1:
  • Data Logger Supply Voltage(logger_volt)
  • Detector Temperature(head_temp)

sgpmfrsrE24.00:
  • null(Raw data stream - documentation not supported)

sgpmfrsrE24.a0:
  • Detector Temperature(head_temp)
  • Data Logger Supply Voltage(logger_volt)


Back To Table of Contents



END OF DATA