Data Quality Reports for Session: 116665 User: renata Completed: 12/23/2008


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D011102.3SGP EBBR E19 - E19 EBBR Soil Temperature 3 Intermittenmtly Incorrectsgp15ebbrE19.b1, sgp30ebbrE19.b1
D011127.30SGP EBBR E19 - E19 EBBR Data Incorrect During 6 Month Checkssgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.a1
D020716.15SGP/EBBR/E19 - 6 Month Checkssgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.a1
D020717.29SGP/EBBR/E19 - Most Measurements Incorrectsgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.00,
sgp5ebbrE19.a1
D020717.30SGP/EBBR/E19 - All Measurements Zerosgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.00,
sgp5ebbrE19.a1
D020717.32SGP/EBBR/E19 - Measurements Incorrect After CR10 Replacementsgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.00,
sgp5ebbrE19.a1
D020717.33SGP/EBBR/E19 - Low Battery Voltagesgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.00,
sgp5ebbrE19.a1
D030110.24SGP/EBBR/E19 - Data Incorrect during 6 month checkssgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.a1
D030116.6SGP/EBBR/E19 - Incorrect data after power interruptionsgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.a1
D030116.7SGP/EBBR/E19 - Most Measurements Zero following Restartsgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.a1
D030117.6SGP/EBBR/E19 - Measurements zero following logger restartsgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.a1
D030711.11SGP/EBBR/E19 - Soil Temperature #2 Incorrectsgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1
D030711.12SGP/EBBR/E19 - All Data Zerossgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.a1
D030711.13SGP/EBBR/E19 - Several Soil Measurements Incorrectsgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1
D030711.14SGP/EBBR/E19 -Several Soil Measurements Were Incorrectsgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1
D030711.15SGP/EBBR/E19 - Soil Temperature #3 Incorrectsgp15ebbrE19.a1, sgp15ebbrE19.b1
D030711.17SGP/EBBR/E19 - Soil Heat Flow #3 Incorrectsgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1
D030711.18SGP/EBBR/E19 - Soil Temperature #2 Incorrectsgp15ebbrE19.a1, sgp15ebbrE19.b1
D030711.19SGP/EBBR/E19 - Soil Heat Flow#1 Incorrectsgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1
D030711.20SGP/EBBR/E19 - Many Measurements Incorrectsgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.a1
D030712.20SGP/EBBR/E19 - 6 Month Checkssgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.a1
D030919.10SGP/EBBR/E19 - Water in Net Radiometer; Surface Fluxes are Incorrectsgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.a1
D030922.6SGP/EBBR/E19 - Tair Right Offscalesgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.a1
D031110.8SGP/EBBR/E19 - Battery Voltage Too Low, Most Measurements Incorrectsgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1
D031117.2SGP/EBBR/E19 - AEM Stuck In One Position, Surface Fluxes Incorrectsgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1
D031124.11SGP/EBBR/E19 - Data Missing During Recalibration Changeoutsgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30baebbrE19.c1, sgp30ebbrE19.a1, sgp30ebbrE19.b1,
sgp5ebbrE19.a1, sgpebbrE19.00, sgpebbrE19.01
D031204.17SGP/EBBR/E19 - 6 Month Checkssgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.a1
D031204.18SGP/EBBR/E19 - 6 Month Checkssgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1
D031204.2SGP/EBBR/E19 - Data All Zerossgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.a1
D040106.1SGP/EBBR/E19 - Wind Speed Sensor Frozen Stopped by Ice Stormsgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1
D040206.4SGP/EBBR/E19 - Low Battery Voltage Caused Incorrect Datasgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1
D040304.26SGP/EBBR/E19 - Wind Speed Frozen Stopped By Icesgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1
D040525.4SGP/EBBR/E19 - Metadata errors sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.a1
D040608.7SGP/EBBR/E19 - AEM Hung Inbetween Home Positionssgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1
D041222.15SGP/EBBR/E19 - 6 Month Checkssgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1
D050206.13SGP/EBBR/E19 - Wind Speed Frozen Stopped By Ice Stormsgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1
D050207.9SGP/EBBR/E19 - Wind Direction Frozen Stopped By Ice Stormsgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1
D050215.20SGP/EBBR/E19 - Wind Speed Constantsgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1
D050624.3SGP/EBBR/E19 - Net Radiation Incorrect - Cable Chewed by Rodentssgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1
D051112.1SGP/EBBR/E19 - metadata correctionssgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1
D051214.2SGP/EBBR/E19 - Net Radiation, Sensible and Latent Heat Fluxes Incorrectsgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1
D051214.3SGP/EBBR/E19 - Incorrect Sensible and Latent Heat Fluxessgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1
D080527.2SGP/EBBR/E19 - Sensible and latent heat flux and net radiation incorrectsgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1


DQRID : D011102.3
Start DateStart TimeEnd DateEnd Time
05/07/2001000005/07/20012356
05/21/2001000005/21/20012356
Subject:
SGP EBBR E19 - E19 EBBR Soil Temperature 3 Intermittenmtly Incorrect
DataStreams:sgp15ebbrE19.b1, sgp30ebbrE19.b1
Description:
Soil Temperature # 3 was incorrect at times during these periods.
Compare G3 with G1, G2, G4, and G5 to determine when TS3 was incorrect.
Measurements:sgp15ebbrE19.b1:
  • Soil temperature 3(rr_ts3)

sgp30ebbrE19.b1:
  • soil heat flow, site 3(g3)
  • average surface soil heat flow(ave_shf)
  • h(h)
  • latent heat flux(e)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 0-5 cm integrated soil temperature, site 3(ts3)


Back To Table of Contents

DQRID : D011127.30
Start DateStart TimeEnd DateEnd Time
04/19/2001170004/19/20012001
10/03/2001141310/03/20011506
Subject:
SGP EBBR E19 - E19 EBBR Data Incorrect During 6 Month Checks
DataStreams:sgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.a1
Description:
The data was incorrect during the 6 month checks.
Measurements:sgp15ebbrE19.a1:
  • Right air temperature(tair_r)
  • Reference temperature(rr_tref)
  • Soil temperature 5(rr_ts5)
  • scalar wind speed(wind_s)
  • Soil temperature 4(rr_ts4)
  • Wind direction (relative to true north)(mv_wind_d)
  • Soil moisture 1(r_sm1)
  • Soil moisture 3(r_sm3)
  • Soil moisture 2(r_sm2)
  • Soil heat flow 1(mv_hft1)
  • Soil heat flow 2(mv_hft2)
  • Left air temperature(tair_l)
  • Left relative humidity(mv_hum_l)
  • Soil heat flow 4(mv_hft4)
  • Right relative humidity(mv_hum_r)
  • Soil heat flow 5(mv_hft5)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Soil heat flow 3(mv_hft3)
  • Soil moisture 4(r_sm4)
  • Soil temperature 1(rr_ts1)
  • Soil temperature 2(rr_ts2)
  • Soil moisture 5(r_sm5)
  • Soil temperature 3(rr_ts3)
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Atmospheric pressure(mv_pres)
  • Net radiation(mv_q)

sgp15ebbrE19.b1:
  • Right relative humidity(mv_hum_r)
  • Soil heat flow 1(mv_hft1)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Soil temperature 1(rr_ts1)
  • Soil heat flow 3(mv_hft3)
  • Atmospheric pressure(mv_pres)
  • Soil temperature 2(rr_ts2)
  • Soil heat flow 2(mv_hft2)
  • Left air temperature(tair_l)
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Soil moisture 5(r_sm5)
  • Soil temperature 4(rr_ts4)
  • Net radiation(mv_q)
  • scalar wind speed(wind_s)
  • Reference temperature(rr_tref)
  • Soil moisture 2(r_sm2)
  • Soil temperature 5(rr_ts5)
  • Left relative humidity(mv_hum_l)
  • Soil temperature 3(rr_ts3)
  • Wind direction (relative to true north)(mv_wind_d)
  • Soil moisture 4(r_sm4)
  • Right air temperature(tair_r)
  • Soil moisture 3(r_sm3)
  • Soil heat flow 4(mv_hft4)
  • Soil heat flow 5(mv_hft5)
  • Soil moisture 1(r_sm1)

sgp5ebbrE19.a1:
  • bottom vapor pressure(vp_bot)
  • Temperature of the top humidity chamber(thum_top)
  • Bottom humidity(hum_bot)
  • Top humidity(hum_top)
  • vector wind speed(res_ws)
  • wind direction (relative to true north)(wind_d)
  • bottom air temperature(tair_bot)
  • Reference Thermistor Temperature(tref)
  • top air temperature(tair_top)
  • scalar wind speed(wind_s)
  • Retrieved pressure profile(pres)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • top vapor pressure(vp_top)
  • net radiation(q)

sgp30ebbrE19.a1:
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • vector wind speed(res_ws)
  • 5 cm soil heat flow, site 1(shf1)
  • soil heat flow, site 2(g2)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • soil heat flow, site 1(g1)
  • 5 cm soil heat flow, site 4(shf4)
  • net radiation(q)
  • Reference Thermistor Temperature(tref)
  • volumetric soil moisture, site 3(sm3)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • 5 cm soil heat flow, site 5(shf5)
  • average surface soil heat flow(ave_shf)
  • top vapor pressure(vp_top)
  • volumetric soil moisture, site 4(sm4)
  • Soil heat capacity 3(cs3)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • Soil heat capacity 5(cs5)
  • Retrieved pressure profile(pres)
  • Temperature of the top humidity chamber(thum_top)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 5 cm soil heat flow, site 2(shf2)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • Soil heat capacity 2(cs2)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • soil heat flow, site 4(g4)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • volumetric soil moisture, site 2(sm2)
  • Soil heat capacity 1(cs1)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • latent heat flux(e)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • Soil heat capacity 4(cs4)
  • h(h)
  • bottom vapor pressure(vp_bot)
  • wind direction (relative to true north)(wind_d)
  • soil heat flow, site 3(g3)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Top humidity(hum_top)
  • volumetric soil moisture, site 5(sm5)
  • Bottom humidity(hum_bot)
  • soil heat flow, site 5(g5)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • 5 cm soil heat flow, site 3(shf3)
  • scalar wind speed(wind_s)
  • top air temperature(tair_top)
  • bottom air temperature(tair_bot)
  • volumetric soil moisture, site 1(sm1)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)

sgp30ebbrE19.b1:
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • Reference Thermistor Temperature(tref)
  • 5 cm soil heat flow, site 2(shf2)
  • volumetric soil moisture, site 1(sm1)
  • volumetric soil moisture, site 2(sm2)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)
  • Soil heat capacity 5(cs5)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)
  • volumetric soil moisture, site 4(sm4)
  • Soil heat capacity 3(cs3)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • h(h)
  • Temperature of the top humidity chamber(thum_top)
  • scalar wind speed(wind_s)
  • top air temperature(tair_top)
  • Soil heat capacity 4(cs4)
  • Bottom humidity(hum_bot)
  • soil heat flow, site 4(g4)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • top vapor pressure(vp_top)
  • vector wind speed(res_ws)
  • average surface soil heat flow(ave_shf)
  • volumetric soil moisture, site 5(sm5)
  • latent heat flux(e)
  • wind direction (relative to true north)(wind_d)
  • soil heat flow, site 2(g2)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Soil heat capacity 1(cs1)
  • volumetric soil moisture, site 3(sm3)
  • soil heat flow, site 5(g5)
  • soil heat flow, site 1(g1)
  • soil heat flow, site 3(g3)
  • bottom vapor pressure(vp_bot)
  • 5 cm soil heat flow, site 1(shf1)
  • Soil heat capacity 2(cs2)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 5 cm soil heat flow, site 5(shf5)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • Top humidity(hum_top)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • bottom air temperature(tair_bot)
  • net radiation(q)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • 5 cm soil heat flow, site 3(shf3)
  • 5 cm soil heat flow, site 4(shf4)
  • Retrieved pressure profile(pres)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)


Back To Table of Contents

DQRID : D020716.15
Start DateStart TimeEnd DateEnd Time
05/16/2002143005/16/20021501
Subject:
SGP/EBBR/E19 - 6 Month Checks
DataStreams:sgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.a1
Description:
The indicated measurements were incorrect during the 6 month checks.
Measurements:sgp15ebbrE19.a1:
  • Right air temperature(tair_r)
  • Left air temperature(tair_l)

sgp15ebbrE19.b1:
  • Right air temperature(tair_r)
  • Left air temperature(tair_l)

sgp5ebbrE19.a1:
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • bottom vapor pressure(vp_bot)
  • top vapor pressure(vp_top)
  • Temperature of the top humidity chamber(thum_top)
  • Bottom humidity(hum_bot)
  • bottom air temperature(tair_bot)
  • Top humidity(hum_top)
  • top air temperature(tair_top)

sgp30ebbrE19.a1:
  • top vapor pressure(vp_top)
  • latent heat flux(e)
  • h(h)
  • bottom vapor pressure(vp_bot)
  • Temperature of the top humidity chamber(thum_top)
  • top air temperature(tair_top)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • bottom air temperature(tair_bot)
  • Top humidity(hum_top)
  • Bottom humidity(hum_bot)

sgp30ebbrE19.b1:
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • h(h)
  • Temperature of the top humidity chamber(thum_top)
  • top air temperature(tair_top)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Top humidity(hum_top)
  • Bottom humidity(hum_bot)
  • bottom air temperature(tair_bot)
  • top vapor pressure(vp_top)
  • bottom vapor pressure(vp_bot)
  • latent heat flux(e)


Back To Table of Contents

DQRID : D020717.29
Start DateStart TimeEnd DateEnd Time
04/08/2002043004/08/20020431
05/19/2002040005/19/20020401
Subject:
SGP/EBBR/E19 - Most Measurements Incorrect
DataStreams:sgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.00,
sgp5ebbrE19.a1
Description:
Most measurements were incorrect for the indicated periods.
Measurements:sgp5ebbrE19.00:
  • null(Raw data stream - documentation not supported)

sgp15ebbrE19.a1:
  • Right air temperature(tair_r)
  • Reference temperature(rr_tref)
  • Soil temperature 5(rr_ts5)
  • scalar wind speed(wind_s)
  • Soil temperature 4(rr_ts4)
  • Wind direction (relative to true north)(mv_wind_d)
  • Soil moisture 1(r_sm1)
  • Home signal(mv_home)
  • Soil moisture 3(r_sm3)
  • Soil moisture 2(r_sm2)
  • Soil heat flow 1(mv_hft1)
  • Soil heat flow 2(mv_hft2)
  • Left air temperature(tair_l)
  • Left relative humidity(mv_hum_l)
  • Battery(bat)
  • Soil heat flow 4(mv_hft4)
  • Right relative humidity(mv_hum_r)
  • Soil heat flow 5(mv_hft5)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Soil heat flow 3(mv_hft3)
  • Soil moisture 4(r_sm4)
  • Soil temperature 1(rr_ts1)
  • Soil temperature 2(rr_ts2)
  • Soil moisture 5(r_sm5)
  • Soil temperature 3(rr_ts3)
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Atmospheric pressure(mv_pres)
  • Net radiation(mv_q)

sgp15ebbrE19.b1:
  • Right relative humidity(mv_hum_r)
  • Soil heat flow 1(mv_hft1)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Soil temperature 1(rr_ts1)
  • Soil heat flow 3(mv_hft3)
  • Atmospheric pressure(mv_pres)
  • Soil temperature 2(rr_ts2)
  • Soil heat flow 2(mv_hft2)
  • Left air temperature(tair_l)
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Soil moisture 5(r_sm5)
  • Soil temperature 4(rr_ts4)
  • Net radiation(mv_q)
  • scalar wind speed(wind_s)
  • Reference temperature(rr_tref)
  • Soil moisture 2(r_sm2)
  • Soil temperature 5(rr_ts5)
  • Battery(bat)
  • Left relative humidity(mv_hum_l)
  • Soil temperature 3(rr_ts3)
  • Home signal(mv_home)
  • Wind direction (relative to true north)(mv_wind_d)
  • Soil moisture 4(r_sm4)
  • Right air temperature(tair_r)
  • Soil moisture 3(r_sm3)
  • Soil heat flow 4(mv_hft4)
  • Soil heat flow 5(mv_hft5)
  • Soil moisture 1(r_sm1)

sgp5ebbrE19.a1:
  • bottom vapor pressure(vp_bot)
  • Temperature of the top humidity chamber(thum_top)
  • Bottom humidity(hum_bot)
  • Top humidity(hum_top)
  • vector wind speed(res_ws)
  • wind direction (relative to true north)(wind_d)
  • bottom air temperature(tair_bot)
  • Reference Thermistor Temperature(tref)
  • top air temperature(tair_top)
  • scalar wind speed(wind_s)
  • Retrieved pressure profile(pres)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Home signal(home)
  • top vapor pressure(vp_top)
  • net radiation(q)

sgp30ebbrE19.a1:
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • vector wind speed(res_ws)
  • 5 cm soil heat flow, site 1(shf1)
  • soil heat flow, site 2(g2)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • soil heat flow, site 1(g1)
  • 5 cm soil heat flow, site 4(shf4)
  • net radiation(q)
  • Exchange mechanism position indicator (15 to 30 min)(home_30)
  • Exchange mechanism position indicator (0 to 15 min)(home_15)
  • Reference Thermistor Temperature(tref)
  • volumetric soil moisture, site 3(sm3)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • 5 cm soil heat flow, site 5(shf5)
  • average surface soil heat flow(ave_shf)
  • top vapor pressure(vp_top)
  • volumetric soil moisture, site 4(sm4)
  • Soil heat capacity 3(cs3)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • Soil heat capacity 5(cs5)
  • Retrieved pressure profile(pres)
  • Temperature of the top humidity chamber(thum_top)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 5 cm soil heat flow, site 2(shf2)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • Soil heat capacity 2(cs2)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • soil heat flow, site 4(g4)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • volumetric soil moisture, site 2(sm2)
  • Soil heat capacity 1(cs1)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • latent heat flux(e)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • Soil heat capacity 4(cs4)
  • h(h)
  • bottom vapor pressure(vp_bot)
  • wind direction (relative to true north)(wind_d)
  • soil heat flow, site 3(g3)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Top humidity(hum_top)
  • volumetric soil moisture, site 5(sm5)
  • Bottom humidity(hum_bot)
  • soil heat flow, site 5(g5)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • 5 cm soil heat flow, site 3(shf3)
  • scalar wind speed(wind_s)
  • top air temperature(tair_top)
  • bottom air temperature(tair_bot)
  • volumetric soil moisture, site 1(sm1)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)

sgp30ebbrE19.b1:
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • Reference Thermistor Temperature(tref)
  • 5 cm soil heat flow, site 2(shf2)
  • volumetric soil moisture, site 1(sm1)
  • volumetric soil moisture, site 2(sm2)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • Exchange mechanism position indicator (15 to 30 min)(home_30)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)
  • Soil heat capacity 5(cs5)
  • Exchange mechanism position indicator (0 to 15 min)(home_15)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)
  • volumetric soil moisture, site 4(sm4)
  • Soil heat capacity 3(cs3)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • h(h)
  • Temperature of the top humidity chamber(thum_top)
  • scalar wind speed(wind_s)
  • top air temperature(tair_top)
  • Soil heat capacity 4(cs4)
  • Bottom humidity(hum_bot)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • soil heat flow, site 4(g4)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • top vapor pressure(vp_top)
  • vector wind speed(res_ws)
  • average surface soil heat flow(ave_shf)
  • volumetric soil moisture, site 5(sm5)
  • latent heat flux(e)
  • wind direction (relative to true north)(wind_d)
  • soil heat flow, site 2(g2)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Soil heat capacity 1(cs1)
  • volumetric soil moisture, site 3(sm3)
  • soil heat flow, site 5(g5)
  • soil heat flow, site 1(g1)
  • soil heat flow, site 3(g3)
  • bottom vapor pressure(vp_bot)
  • 5 cm soil heat flow, site 1(shf1)
  • Soil heat capacity 2(cs2)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 5 cm soil heat flow, site 5(shf5)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • Top humidity(hum_top)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • bottom air temperature(tair_bot)
  • net radiation(q)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • 5 cm soil heat flow, site 3(shf3)
  • 5 cm soil heat flow, site 4(shf4)
  • Retrieved pressure profile(pres)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)


Back To Table of Contents

DQRID : D020717.30
Start DateStart TimeEnd DateEnd Time
04/18/2002173004/18/20021731
05/05/2002140005/05/20021401
06/27/2002150006/27/20021501
Subject:
SGP/EBBR/E19 - All Measurements Zero
DataStreams:sgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.00,
sgp5ebbrE19.a1
Description:
Measurements are often zero the first period after a CR10 restart.
Measurements:sgp5ebbrE19.00:
  • null(Raw data stream - documentation not supported)

sgp15ebbrE19.a1:
  • Right air temperature(tair_r)
  • Reference temperature(rr_tref)
  • Soil temperature 5(rr_ts5)
  • scalar wind speed(wind_s)
  • Soil temperature 4(rr_ts4)
  • Wind direction (relative to true north)(mv_wind_d)
  • Soil moisture 1(r_sm1)
  • Home signal(mv_home)
  • Soil moisture 3(r_sm3)
  • Soil moisture 2(r_sm2)
  • Soil heat flow 1(mv_hft1)
  • Soil heat flow 2(mv_hft2)
  • Left air temperature(tair_l)
  • Left relative humidity(mv_hum_l)
  • Battery(bat)
  • Soil heat flow 4(mv_hft4)
  • Right relative humidity(mv_hum_r)
  • Soil heat flow 5(mv_hft5)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Soil heat flow 3(mv_hft3)
  • Soil moisture 4(r_sm4)
  • Soil temperature 1(rr_ts1)
  • Soil temperature 2(rr_ts2)
  • Soil moisture 5(r_sm5)
  • Soil temperature 3(rr_ts3)
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Atmospheric pressure(mv_pres)
  • Net radiation(mv_q)

sgp15ebbrE19.b1:
  • Right relative humidity(mv_hum_r)
  • Soil heat flow 1(mv_hft1)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Soil temperature 1(rr_ts1)
  • Soil heat flow 3(mv_hft3)
  • Atmospheric pressure(mv_pres)
  • Soil temperature 2(rr_ts2)
  • Soil heat flow 2(mv_hft2)
  • Left air temperature(tair_l)
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Soil moisture 5(r_sm5)
  • Soil temperature 4(rr_ts4)
  • Net radiation(mv_q)
  • scalar wind speed(wind_s)
  • Reference temperature(rr_tref)
  • Soil moisture 2(r_sm2)
  • Soil temperature 5(rr_ts5)
  • Battery(bat)
  • Left relative humidity(mv_hum_l)
  • Soil temperature 3(rr_ts3)
  • Home signal(mv_home)
  • Wind direction (relative to true north)(mv_wind_d)
  • Soil moisture 4(r_sm4)
  • Right air temperature(tair_r)
  • Soil moisture 3(r_sm3)
  • Soil heat flow 4(mv_hft4)
  • Soil heat flow 5(mv_hft5)
  • Soil moisture 1(r_sm1)

sgp5ebbrE19.a1:
  • bottom vapor pressure(vp_bot)
  • Temperature of the top humidity chamber(thum_top)
  • Bottom humidity(hum_bot)
  • Top humidity(hum_top)
  • vector wind speed(res_ws)
  • wind direction (relative to true north)(wind_d)
  • bottom air temperature(tair_bot)
  • Reference Thermistor Temperature(tref)
  • top air temperature(tair_top)
  • scalar wind speed(wind_s)
  • Retrieved pressure profile(pres)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Home signal(home)
  • top vapor pressure(vp_top)
  • net radiation(q)

sgp30ebbrE19.a1:
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • vector wind speed(res_ws)
  • 5 cm soil heat flow, site 1(shf1)
  • soil heat flow, site 2(g2)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • soil heat flow, site 1(g1)
  • 5 cm soil heat flow, site 4(shf4)
  • net radiation(q)
  • Exchange mechanism position indicator (15 to 30 min)(home_30)
  • Exchange mechanism position indicator (0 to 15 min)(home_15)
  • Reference Thermistor Temperature(tref)
  • volumetric soil moisture, site 3(sm3)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • 5 cm soil heat flow, site 5(shf5)
  • average surface soil heat flow(ave_shf)
  • top vapor pressure(vp_top)
  • volumetric soil moisture, site 4(sm4)
  • Soil heat capacity 3(cs3)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • Soil heat capacity 5(cs5)
  • Retrieved pressure profile(pres)
  • Temperature of the top humidity chamber(thum_top)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 5 cm soil heat flow, site 2(shf2)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • Soil heat capacity 2(cs2)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • soil heat flow, site 4(g4)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • volumetric soil moisture, site 2(sm2)
  • Soil heat capacity 1(cs1)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • latent heat flux(e)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • Soil heat capacity 4(cs4)
  • h(h)
  • bottom vapor pressure(vp_bot)
  • wind direction (relative to true north)(wind_d)
  • soil heat flow, site 3(g3)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Top humidity(hum_top)
  • volumetric soil moisture, site 5(sm5)
  • Bottom humidity(hum_bot)
  • soil heat flow, site 5(g5)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • 5 cm soil heat flow, site 3(shf3)
  • scalar wind speed(wind_s)
  • top air temperature(tair_top)
  • bottom air temperature(tair_bot)
  • volumetric soil moisture, site 1(sm1)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)

sgp30ebbrE19.b1:
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • Reference Thermistor Temperature(tref)
  • 5 cm soil heat flow, site 2(shf2)
  • volumetric soil moisture, site 1(sm1)
  • volumetric soil moisture, site 2(sm2)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • Exchange mechanism position indicator (15 to 30 min)(home_30)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)
  • Soil heat capacity 5(cs5)
  • Exchange mechanism position indicator (0 to 15 min)(home_15)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)
  • volumetric soil moisture, site 4(sm4)
  • Soil heat capacity 3(cs3)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • h(h)
  • Temperature of the top humidity chamber(thum_top)
  • scalar wind speed(wind_s)
  • top air temperature(tair_top)
  • Soil heat capacity 4(cs4)
  • Bottom humidity(hum_bot)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • soil heat flow, site 4(g4)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • top vapor pressure(vp_top)
  • vector wind speed(res_ws)
  • average surface soil heat flow(ave_shf)
  • volumetric soil moisture, site 5(sm5)
  • latent heat flux(e)
  • wind direction (relative to true north)(wind_d)
  • soil heat flow, site 2(g2)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Soil heat capacity 1(cs1)
  • volumetric soil moisture, site 3(sm3)
  • soil heat flow, site 5(g5)
  • soil heat flow, site 1(g1)
  • soil heat flow, site 3(g3)
  • bottom vapor pressure(vp_bot)
  • 5 cm soil heat flow, site 1(shf1)
  • Soil heat capacity 2(cs2)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 5 cm soil heat flow, site 5(shf5)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • Top humidity(hum_top)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • bottom air temperature(tair_bot)
  • net radiation(q)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • 5 cm soil heat flow, site 3(shf3)
  • 5 cm soil heat flow, site 4(shf4)
  • Retrieved pressure profile(pres)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)


Back To Table of Contents

DQRID : D020717.32
Start DateStart TimeEnd DateEnd Time
05/16/2002200005/17/20020531
Subject:
SGP/EBBR/E19 - Measurements Incorrect After CR10 Replacement
DataStreams:sgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.00,
sgp5ebbrE19.a1
Description:
Replacement of the CR10 datalogger caused most measurements to be incorrect, reason 
unknown.
A different logger was installed at the end of the specified period and this resolved the 
problem.
Measurements:sgp5ebbrE19.00:
  • null(Raw data stream - documentation not supported)

sgp15ebbrE19.a1:
  • Right air temperature(tair_r)
  • Reference temperature(rr_tref)
  • Soil temperature 5(rr_ts5)
  • scalar wind speed(wind_s)
  • Soil temperature 4(rr_ts4)
  • Wind direction (relative to true north)(mv_wind_d)
  • Soil moisture 1(r_sm1)
  • Home signal(mv_home)
  • Soil moisture 3(r_sm3)
  • Soil moisture 2(r_sm2)
  • Soil heat flow 1(mv_hft1)
  • Soil heat flow 2(mv_hft2)
  • Left air temperature(tair_l)
  • Left relative humidity(mv_hum_l)
  • Battery(bat)
  • Soil heat flow 4(mv_hft4)
  • Right relative humidity(mv_hum_r)
  • Soil heat flow 5(mv_hft5)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Soil heat flow 3(mv_hft3)
  • Soil moisture 4(r_sm4)
  • Soil temperature 1(rr_ts1)
  • Soil temperature 2(rr_ts2)
  • Soil moisture 5(r_sm5)
  • Soil temperature 3(rr_ts3)
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Atmospheric pressure(mv_pres)
  • Net radiation(mv_q)

sgp15ebbrE19.b1:
  • Right relative humidity(mv_hum_r)
  • Soil heat flow 1(mv_hft1)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Soil temperature 1(rr_ts1)
  • Soil heat flow 3(mv_hft3)
  • Atmospheric pressure(mv_pres)
  • Soil temperature 2(rr_ts2)
  • Soil heat flow 2(mv_hft2)
  • Left air temperature(tair_l)
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Soil moisture 5(r_sm5)
  • Soil temperature 4(rr_ts4)
  • Net radiation(mv_q)
  • scalar wind speed(wind_s)
  • Reference temperature(rr_tref)
  • Soil moisture 2(r_sm2)
  • Soil temperature 5(rr_ts5)
  • Battery(bat)
  • Left relative humidity(mv_hum_l)
  • Soil temperature 3(rr_ts3)
  • Home signal(mv_home)
  • Wind direction (relative to true north)(mv_wind_d)
  • Soil moisture 4(r_sm4)
  • Right air temperature(tair_r)
  • Soil moisture 3(r_sm3)
  • Soil heat flow 4(mv_hft4)
  • Soil heat flow 5(mv_hft5)
  • Soil moisture 1(r_sm1)

sgp5ebbrE19.a1:
  • bottom vapor pressure(vp_bot)
  • Temperature of the top humidity chamber(thum_top)
  • Bottom humidity(hum_bot)
  • Top humidity(hum_top)
  • vector wind speed(res_ws)
  • wind direction (relative to true north)(wind_d)
  • bottom air temperature(tair_bot)
  • Reference Thermistor Temperature(tref)
  • top air temperature(tair_top)
  • scalar wind speed(wind_s)
  • Retrieved pressure profile(pres)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Home signal(home)
  • top vapor pressure(vp_top)
  • net radiation(q)

sgp30ebbrE19.a1:
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • vector wind speed(res_ws)
  • 5 cm soil heat flow, site 1(shf1)
  • soil heat flow, site 2(g2)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • soil heat flow, site 1(g1)
  • 5 cm soil heat flow, site 4(shf4)
  • net radiation(q)
  • Exchange mechanism position indicator (15 to 30 min)(home_30)
  • Exchange mechanism position indicator (0 to 15 min)(home_15)
  • Reference Thermistor Temperature(tref)
  • volumetric soil moisture, site 3(sm3)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • 5 cm soil heat flow, site 5(shf5)
  • average surface soil heat flow(ave_shf)
  • top vapor pressure(vp_top)
  • volumetric soil moisture, site 4(sm4)
  • Soil heat capacity 3(cs3)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • Soil heat capacity 5(cs5)
  • Retrieved pressure profile(pres)
  • Temperature of the top humidity chamber(thum_top)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 5 cm soil heat flow, site 2(shf2)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • Soil heat capacity 2(cs2)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • soil heat flow, site 4(g4)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • volumetric soil moisture, site 2(sm2)
  • Soil heat capacity 1(cs1)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • latent heat flux(e)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • Soil heat capacity 4(cs4)
  • h(h)
  • bottom vapor pressure(vp_bot)
  • wind direction (relative to true north)(wind_d)
  • soil heat flow, site 3(g3)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Top humidity(hum_top)
  • volumetric soil moisture, site 5(sm5)
  • Bottom humidity(hum_bot)
  • soil heat flow, site 5(g5)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • 5 cm soil heat flow, site 3(shf3)
  • scalar wind speed(wind_s)
  • top air temperature(tair_top)
  • bottom air temperature(tair_bot)
  • volumetric soil moisture, site 1(sm1)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)

sgp30ebbrE19.b1:
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • Reference Thermistor Temperature(tref)
  • 5 cm soil heat flow, site 2(shf2)
  • volumetric soil moisture, site 1(sm1)
  • volumetric soil moisture, site 2(sm2)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • Exchange mechanism position indicator (15 to 30 min)(home_30)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)
  • Soil heat capacity 5(cs5)
  • Exchange mechanism position indicator (0 to 15 min)(home_15)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)
  • volumetric soil moisture, site 4(sm4)
  • Soil heat capacity 3(cs3)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • h(h)
  • Temperature of the top humidity chamber(thum_top)
  • scalar wind speed(wind_s)
  • top air temperature(tair_top)
  • Soil heat capacity 4(cs4)
  • Bottom humidity(hum_bot)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • soil heat flow, site 4(g4)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • top vapor pressure(vp_top)
  • vector wind speed(res_ws)
  • average surface soil heat flow(ave_shf)
  • volumetric soil moisture, site 5(sm5)
  • latent heat flux(e)
  • wind direction (relative to true north)(wind_d)
  • soil heat flow, site 2(g2)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Soil heat capacity 1(cs1)
  • volumetric soil moisture, site 3(sm3)
  • soil heat flow, site 5(g5)
  • soil heat flow, site 1(g1)
  • soil heat flow, site 3(g3)
  • bottom vapor pressure(vp_bot)
  • 5 cm soil heat flow, site 1(shf1)
  • Soil heat capacity 2(cs2)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 5 cm soil heat flow, site 5(shf5)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • Top humidity(hum_top)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • bottom air temperature(tair_bot)
  • net radiation(q)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • 5 cm soil heat flow, site 3(shf3)
  • 5 cm soil heat flow, site 4(shf4)
  • Retrieved pressure profile(pres)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)


Back To Table of Contents

DQRID : D020717.33
Start DateStart TimeEnd DateEnd Time
05/27/2002000006/02/20021031
Subject:
SGP/EBBR/E19 - Low Battery Voltage
DataStreams:sgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.00,
sgp5ebbrE19.a1
Description:
A battery voltage of less than 10.5 causes many of the sensors to produce incorrect 
measurements.
The 15 minute data contains the battery voltage and should be used to determine whether 
the measurements are usable, based on battery voltage.
Measurements:sgp5ebbrE19.00:
  • null(Raw data stream - documentation not supported)

sgp15ebbrE19.a1:
  • Right air temperature(tair_r)
  • Reference temperature(rr_tref)
  • Soil temperature 5(rr_ts5)
  • scalar wind speed(wind_s)
  • Soil temperature 4(rr_ts4)
  • Wind direction (relative to true north)(mv_wind_d)
  • Soil moisture 1(r_sm1)
  • Home signal(mv_home)
  • Soil moisture 3(r_sm3)
  • Soil moisture 2(r_sm2)
  • Soil heat flow 1(mv_hft1)
  • Soil heat flow 2(mv_hft2)
  • Left air temperature(tair_l)
  • Left relative humidity(mv_hum_l)
  • Soil heat flow 4(mv_hft4)
  • Right relative humidity(mv_hum_r)
  • Soil heat flow 5(mv_hft5)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Soil heat flow 3(mv_hft3)
  • Soil moisture 4(r_sm4)
  • Soil temperature 1(rr_ts1)
  • Soil temperature 2(rr_ts2)
  • Soil moisture 5(r_sm5)
  • Soil temperature 3(rr_ts3)
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Atmospheric pressure(mv_pres)
  • Net radiation(mv_q)

sgp15ebbrE19.b1:
  • Right relative humidity(mv_hum_r)
  • Soil heat flow 1(mv_hft1)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Soil temperature 1(rr_ts1)
  • Soil heat flow 3(mv_hft3)
  • Atmospheric pressure(mv_pres)
  • Soil temperature 2(rr_ts2)
  • Soil heat flow 2(mv_hft2)
  • Left air temperature(tair_l)
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Soil moisture 5(r_sm5)
  • Soil temperature 4(rr_ts4)
  • Net radiation(mv_q)
  • scalar wind speed(wind_s)
  • Reference temperature(rr_tref)
  • Soil moisture 2(r_sm2)
  • Soil temperature 5(rr_ts5)
  • Left relative humidity(mv_hum_l)
  • Soil temperature 3(rr_ts3)
  • Home signal(mv_home)
  • Wind direction (relative to true north)(mv_wind_d)
  • Soil moisture 4(r_sm4)
  • Right air temperature(tair_r)
  • Soil moisture 3(r_sm3)
  • Soil heat flow 4(mv_hft4)
  • Soil heat flow 5(mv_hft5)
  • Soil moisture 1(r_sm1)

sgp5ebbrE19.a1:
  • bottom vapor pressure(vp_bot)
  • Temperature of the top humidity chamber(thum_top)
  • Bottom humidity(hum_bot)
  • Top humidity(hum_top)
  • vector wind speed(res_ws)
  • wind direction (relative to true north)(wind_d)
  • bottom air temperature(tair_bot)
  • Reference Thermistor Temperature(tref)
  • top air temperature(tair_top)
  • scalar wind speed(wind_s)
  • Retrieved pressure profile(pres)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Home signal(home)
  • top vapor pressure(vp_top)
  • net radiation(q)

sgp30ebbrE19.a1:
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • vector wind speed(res_ws)
  • 5 cm soil heat flow, site 1(shf1)
  • soil heat flow, site 2(g2)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • soil heat flow, site 1(g1)
  • 5 cm soil heat flow, site 4(shf4)
  • net radiation(q)
  • Exchange mechanism position indicator (15 to 30 min)(home_30)
  • Exchange mechanism position indicator (0 to 15 min)(home_15)
  • Reference Thermistor Temperature(tref)
  • volumetric soil moisture, site 3(sm3)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • 5 cm soil heat flow, site 5(shf5)
  • average surface soil heat flow(ave_shf)
  • top vapor pressure(vp_top)
  • volumetric soil moisture, site 4(sm4)
  • Soil heat capacity 3(cs3)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • Soil heat capacity 5(cs5)
  • Retrieved pressure profile(pres)
  • Temperature of the top humidity chamber(thum_top)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 5 cm soil heat flow, site 2(shf2)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • Soil heat capacity 2(cs2)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • soil heat flow, site 4(g4)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • volumetric soil moisture, site 2(sm2)
  • Soil heat capacity 1(cs1)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • latent heat flux(e)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • Soil heat capacity 4(cs4)
  • h(h)
  • bottom vapor pressure(vp_bot)
  • wind direction (relative to true north)(wind_d)
  • soil heat flow, site 3(g3)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Top humidity(hum_top)
  • volumetric soil moisture, site 5(sm5)
  • Bottom humidity(hum_bot)
  • soil heat flow, site 5(g5)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • 5 cm soil heat flow, site 3(shf3)
  • scalar wind speed(wind_s)
  • top air temperature(tair_top)
  • bottom air temperature(tair_bot)
  • volumetric soil moisture, site 1(sm1)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)

sgp30ebbrE19.b1:
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • Reference Thermistor Temperature(tref)
  • 5 cm soil heat flow, site 2(shf2)
  • volumetric soil moisture, site 1(sm1)
  • volumetric soil moisture, site 2(sm2)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • Exchange mechanism position indicator (15 to 30 min)(home_30)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)
  • Soil heat capacity 5(cs5)
  • Exchange mechanism position indicator (0 to 15 min)(home_15)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)
  • volumetric soil moisture, site 4(sm4)
  • Soil heat capacity 3(cs3)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • h(h)
  • Temperature of the top humidity chamber(thum_top)
  • scalar wind speed(wind_s)
  • top air temperature(tair_top)
  • Soil heat capacity 4(cs4)
  • Bottom humidity(hum_bot)
  • soil heat flow, site 4(g4)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • top vapor pressure(vp_top)
  • vector wind speed(res_ws)
  • average surface soil heat flow(ave_shf)
  • volumetric soil moisture, site 5(sm5)
  • latent heat flux(e)
  • wind direction (relative to true north)(wind_d)
  • soil heat flow, site 2(g2)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Soil heat capacity 1(cs1)
  • volumetric soil moisture, site 3(sm3)
  • soil heat flow, site 5(g5)
  • soil heat flow, site 1(g1)
  • soil heat flow, site 3(g3)
  • bottom vapor pressure(vp_bot)
  • 5 cm soil heat flow, site 1(shf1)
  • Soil heat capacity 2(cs2)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 5 cm soil heat flow, site 5(shf5)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • Top humidity(hum_top)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • bottom air temperature(tair_bot)
  • net radiation(q)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • 5 cm soil heat flow, site 3(shf3)
  • 5 cm soil heat flow, site 4(shf4)
  • Retrieved pressure profile(pres)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)


Back To Table of Contents

DQRID : D030110.24
Start DateStart TimeEnd DateEnd Time
11/14/2002160011/14/20021630
Subject:
SGP/EBBR/E19 - Data Incorrect during 6 month checks
DataStreams:sgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.a1
Description:
Some of the measurements are incorrect during the 6 month checks.
Measurements:sgp15ebbrE19.a1:
  • Right air temperature(tair_r)
  • Left air temperature(tair_l)

sgp15ebbrE19.b1:
  • Right air temperature(tair_r)
  • Left air temperature(tair_l)

sgp5ebbrE19.a1:
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • bottom vapor pressure(vp_bot)
  • top vapor pressure(vp_top)
  • Temperature of the top humidity chamber(thum_top)
  • Bottom humidity(hum_bot)
  • bottom air temperature(tair_bot)
  • Top humidity(hum_top)
  • top air temperature(tair_top)

sgp30ebbrE19.a1:
  • top vapor pressure(vp_top)
  • latent heat flux(e)
  • h(h)
  • bottom vapor pressure(vp_bot)
  • Temperature of the top humidity chamber(thum_top)
  • top air temperature(tair_top)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • bottom air temperature(tair_bot)
  • Top humidity(hum_top)
  • Bottom humidity(hum_bot)

sgp30ebbrE19.b1:
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • h(h)
  • Temperature of the top humidity chamber(thum_top)
  • top air temperature(tair_top)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Top humidity(hum_top)
  • Bottom humidity(hum_bot)
  • bottom air temperature(tair_bot)
  • top vapor pressure(vp_top)
  • bottom vapor pressure(vp_bot)
  • latent heat flux(e)


Back To Table of Contents

DQRID : D030116.6
Start DateStart TimeEnd DateEnd Time
07/11/2002160007/11/20021629
Subject:
SGP/EBBR/E19 - Incorrect data after power interruption
DataStreams:sgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.a1
Description:
Most measurements are suspect after a power outage, caused by
replacement of the solar panel regulator.
Measurements:sgp15ebbrE19.a1:
  • Right air temperature(tair_r)
  • Reference temperature(rr_tref)
  • Soil temperature 5(rr_ts5)
  • scalar wind speed(wind_s)
  • Soil temperature 4(rr_ts4)
  • Wind direction (relative to true north)(mv_wind_d)
  • Soil moisture 1(r_sm1)
  • Home signal(mv_home)
  • Soil moisture 3(r_sm3)
  • Soil moisture 2(r_sm2)
  • Soil heat flow 1(mv_hft1)
  • Soil heat flow 2(mv_hft2)
  • Left air temperature(tair_l)
  • Left relative humidity(mv_hum_l)
  • Soil heat flow 4(mv_hft4)
  • Right relative humidity(mv_hum_r)
  • Soil heat flow 5(mv_hft5)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Soil heat flow 3(mv_hft3)
  • Soil moisture 4(r_sm4)
  • Soil temperature 1(rr_ts1)
  • Soil temperature 2(rr_ts2)
  • Soil moisture 5(r_sm5)
  • Soil temperature 3(rr_ts3)
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Atmospheric pressure(mv_pres)
  • Net radiation(mv_q)

sgp15ebbrE19.b1:
  • Right relative humidity(mv_hum_r)
  • Soil heat flow 1(mv_hft1)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Soil temperature 1(rr_ts1)
  • Soil heat flow 3(mv_hft3)
  • Atmospheric pressure(mv_pres)
  • Soil temperature 2(rr_ts2)
  • Soil heat flow 2(mv_hft2)
  • Left air temperature(tair_l)
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Soil moisture 5(r_sm5)
  • Soil temperature 4(rr_ts4)
  • Net radiation(mv_q)
  • scalar wind speed(wind_s)
  • Reference temperature(rr_tref)
  • Soil moisture 2(r_sm2)
  • Soil temperature 5(rr_ts5)
  • Left relative humidity(mv_hum_l)
  • Soil temperature 3(rr_ts3)
  • Home signal(mv_home)
  • Wind direction (relative to true north)(mv_wind_d)
  • Soil moisture 4(r_sm4)
  • Right air temperature(tair_r)
  • Soil moisture 3(r_sm3)
  • Soil heat flow 4(mv_hft4)
  • Soil heat flow 5(mv_hft5)
  • Soil moisture 1(r_sm1)

sgp5ebbrE19.a1:
  • bottom vapor pressure(vp_bot)
  • Temperature of the top humidity chamber(thum_top)
  • Bottom humidity(hum_bot)
  • Top humidity(hum_top)
  • vector wind speed(res_ws)
  • wind direction (relative to true north)(wind_d)
  • bottom air temperature(tair_bot)
  • Reference Thermistor Temperature(tref)
  • top air temperature(tair_top)
  • Retrieved pressure profile(pres)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Home signal(home)
  • top vapor pressure(vp_top)
  • net radiation(q)

sgp30ebbrE19.a1:
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • vector wind speed(res_ws)
  • 5 cm soil heat flow, site 1(shf1)
  • soil heat flow, site 2(g2)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • soil heat flow, site 1(g1)
  • 5 cm soil heat flow, site 4(shf4)
  • net radiation(q)
  • Exchange mechanism position indicator (15 to 30 min)(home_30)
  • Exchange mechanism position indicator (0 to 15 min)(home_15)
  • Reference Thermistor Temperature(tref)
  • volumetric soil moisture, site 3(sm3)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • 5 cm soil heat flow, site 5(shf5)
  • average surface soil heat flow(ave_shf)
  • top vapor pressure(vp_top)
  • volumetric soil moisture, site 4(sm4)
  • Soil heat capacity 3(cs3)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • Soil heat capacity 5(cs5)
  • Retrieved pressure profile(pres)
  • Temperature of the top humidity chamber(thum_top)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 5 cm soil heat flow, site 2(shf2)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • Soil heat capacity 2(cs2)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • soil heat flow, site 4(g4)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • volumetric soil moisture, site 2(sm2)
  • Soil heat capacity 1(cs1)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • latent heat flux(e)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • Soil heat capacity 4(cs4)
  • h(h)
  • bottom vapor pressure(vp_bot)
  • wind direction (relative to true north)(wind_d)
  • soil heat flow, site 3(g3)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Top humidity(hum_top)
  • volumetric soil moisture, site 5(sm5)
  • Bottom humidity(hum_bot)
  • soil heat flow, site 5(g5)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • 5 cm soil heat flow, site 3(shf3)
  • scalar wind speed(wind_s)
  • top air temperature(tair_top)
  • bottom air temperature(tair_bot)
  • volumetric soil moisture, site 1(sm1)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)

sgp30ebbrE19.b1:
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • Reference Thermistor Temperature(tref)
  • 5 cm soil heat flow, site 2(shf2)
  • volumetric soil moisture, site 1(sm1)
  • volumetric soil moisture, site 2(sm2)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • Exchange mechanism position indicator (15 to 30 min)(home_30)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)
  • Soil heat capacity 5(cs5)
  • Exchange mechanism position indicator (0 to 15 min)(home_15)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)
  • volumetric soil moisture, site 4(sm4)
  • Soil heat capacity 3(cs3)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • h(h)
  • Temperature of the top humidity chamber(thum_top)
  • scalar wind speed(wind_s)
  • top air temperature(tair_top)
  • Soil heat capacity 4(cs4)
  • Bottom humidity(hum_bot)
  • soil heat flow, site 4(g4)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • top vapor pressure(vp_top)
  • vector wind speed(res_ws)
  • average surface soil heat flow(ave_shf)
  • volumetric soil moisture, site 5(sm5)
  • latent heat flux(e)
  • wind direction (relative to true north)(wind_d)
  • soil heat flow, site 2(g2)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Soil heat capacity 1(cs1)
  • volumetric soil moisture, site 3(sm3)
  • soil heat flow, site 5(g5)
  • soil heat flow, site 1(g1)
  • soil heat flow, site 3(g3)
  • bottom vapor pressure(vp_bot)
  • 5 cm soil heat flow, site 1(shf1)
  • Soil heat capacity 2(cs2)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 5 cm soil heat flow, site 5(shf5)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • Top humidity(hum_top)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • bottom air temperature(tair_bot)
  • net radiation(q)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • 5 cm soil heat flow, site 3(shf3)
  • 5 cm soil heat flow, site 4(shf4)
  • Retrieved pressure profile(pres)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)


Back To Table of Contents

DQRID : D030116.7
Start DateStart TimeEnd DateEnd Time
07/11/2002150007/11/20021529
Subject:
SGP/EBBR/E19 - Most Measurements Zero following Restart
DataStreams:sgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.a1
Description:
Most of the measurements were zero, which normally happens after a 
datalogger restart.
Measurements:sgp15ebbrE19.a1:
  • Right air temperature(tair_r)
  • Reference temperature(rr_tref)
  • Soil temperature 5(rr_ts5)
  • scalar wind speed(wind_s)
  • Soil temperature 4(rr_ts4)
  • Wind direction (relative to true north)(mv_wind_d)
  • Soil moisture 1(r_sm1)
  • Home signal(mv_home)
  • Soil moisture 3(r_sm3)
  • Soil moisture 2(r_sm2)
  • Soil heat flow 1(mv_hft1)
  • Soil heat flow 2(mv_hft2)
  • Left air temperature(tair_l)
  • Left relative humidity(mv_hum_l)
  • Battery(bat)
  • Soil heat flow 4(mv_hft4)
  • Right relative humidity(mv_hum_r)
  • Soil heat flow 5(mv_hft5)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Soil heat flow 3(mv_hft3)
  • Soil moisture 4(r_sm4)
  • Soil temperature 1(rr_ts1)
  • Soil temperature 2(rr_ts2)
  • Soil moisture 5(r_sm5)
  • Soil temperature 3(rr_ts3)
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Atmospheric pressure(mv_pres)
  • Net radiation(mv_q)

sgp15ebbrE19.b1:
  • Right relative humidity(mv_hum_r)
  • Soil heat flow 1(mv_hft1)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Soil temperature 1(rr_ts1)
  • Soil heat flow 3(mv_hft3)
  • Atmospheric pressure(mv_pres)
  • Soil temperature 2(rr_ts2)
  • Soil heat flow 2(mv_hft2)
  • Left air temperature(tair_l)
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Soil moisture 5(r_sm5)
  • Soil temperature 4(rr_ts4)
  • Net radiation(mv_q)
  • scalar wind speed(wind_s)
  • Reference temperature(rr_tref)
  • Soil moisture 2(r_sm2)
  • Soil temperature 5(rr_ts5)
  • Battery(bat)
  • Left relative humidity(mv_hum_l)
  • Soil temperature 3(rr_ts3)
  • Home signal(mv_home)
  • Wind direction (relative to true north)(mv_wind_d)
  • Soil moisture 4(r_sm4)
  • Right air temperature(tair_r)
  • Soil moisture 3(r_sm3)
  • Soil heat flow 4(mv_hft4)
  • Soil heat flow 5(mv_hft5)
  • Soil moisture 1(r_sm1)

sgp5ebbrE19.a1:
  • bottom vapor pressure(vp_bot)
  • Temperature of the top humidity chamber(thum_top)
  • Bottom humidity(hum_bot)
  • Top humidity(hum_top)
  • vector wind speed(res_ws)
  • wind direction (relative to true north)(wind_d)
  • bottom air temperature(tair_bot)
  • Reference Thermistor Temperature(tref)
  • top air temperature(tair_top)
  • scalar wind speed(wind_s)
  • Retrieved pressure profile(pres)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Home signal(home)
  • top vapor pressure(vp_top)
  • net radiation(q)

sgp30ebbrE19.a1:
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • vector wind speed(res_ws)
  • 5 cm soil heat flow, site 1(shf1)
  • soil heat flow, site 2(g2)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • soil heat flow, site 1(g1)
  • 5 cm soil heat flow, site 4(shf4)
  • net radiation(q)
  • Exchange mechanism position indicator (15 to 30 min)(home_30)
  • Exchange mechanism position indicator (0 to 15 min)(home_15)
  • Reference Thermistor Temperature(tref)
  • volumetric soil moisture, site 3(sm3)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • 5 cm soil heat flow, site 5(shf5)
  • average surface soil heat flow(ave_shf)
  • top vapor pressure(vp_top)
  • volumetric soil moisture, site 4(sm4)
  • Soil heat capacity 3(cs3)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • Soil heat capacity 5(cs5)
  • Retrieved pressure profile(pres)
  • Temperature of the top humidity chamber(thum_top)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 5 cm soil heat flow, site 2(shf2)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • Soil heat capacity 2(cs2)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • soil heat flow, site 4(g4)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • volumetric soil moisture, site 2(sm2)
  • Soil heat capacity 1(cs1)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • latent heat flux(e)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • Soil heat capacity 4(cs4)
  • h(h)
  • bottom vapor pressure(vp_bot)
  • wind direction (relative to true north)(wind_d)
  • soil heat flow, site 3(g3)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Top humidity(hum_top)
  • volumetric soil moisture, site 5(sm5)
  • Bottom humidity(hum_bot)
  • soil heat flow, site 5(g5)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • 5 cm soil heat flow, site 3(shf3)
  • scalar wind speed(wind_s)
  • top air temperature(tair_top)
  • bottom air temperature(tair_bot)
  • volumetric soil moisture, site 1(sm1)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)

sgp30ebbrE19.b1:
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • Reference Thermistor Temperature(tref)
  • 5 cm soil heat flow, site 2(shf2)
  • volumetric soil moisture, site 1(sm1)
  • volumetric soil moisture, site 2(sm2)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • Exchange mechanism position indicator (15 to 30 min)(home_30)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)
  • Soil heat capacity 5(cs5)
  • Exchange mechanism position indicator (0 to 15 min)(home_15)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)
  • volumetric soil moisture, site 4(sm4)
  • Soil heat capacity 3(cs3)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • h(h)
  • Temperature of the top humidity chamber(thum_top)
  • scalar wind speed(wind_s)
  • top air temperature(tair_top)
  • Soil heat capacity 4(cs4)
  • Bottom humidity(hum_bot)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • soil heat flow, site 4(g4)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • top vapor pressure(vp_top)
  • vector wind speed(res_ws)
  • average surface soil heat flow(ave_shf)
  • volumetric soil moisture, site 5(sm5)
  • latent heat flux(e)
  • wind direction (relative to true north)(wind_d)
  • soil heat flow, site 2(g2)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Soil heat capacity 1(cs1)
  • volumetric soil moisture, site 3(sm3)
  • soil heat flow, site 5(g5)
  • soil heat flow, site 1(g1)
  • soil heat flow, site 3(g3)
  • bottom vapor pressure(vp_bot)
  • 5 cm soil heat flow, site 1(shf1)
  • Soil heat capacity 2(cs2)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 5 cm soil heat flow, site 5(shf5)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • Top humidity(hum_top)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • bottom air temperature(tair_bot)
  • net radiation(q)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • 5 cm soil heat flow, site 3(shf3)
  • 5 cm soil heat flow, site 4(shf4)
  • Retrieved pressure profile(pres)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)


Back To Table of Contents

DQRID : D030117.6
Start DateStart TimeEnd DateEnd Time
01/09/2003170001/09/20031729
Subject:
SGP/EBBR/E19 - Measurements zero following logger restart
DataStreams:sgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.a1
Description:
Most of the measurements were zero, which normally happens after a 
datalogger restart.  The restart on 01/09/03 occurred after J-panel
replacements.
Measurements:sgp15ebbrE19.a1:
  • Right air temperature(tair_r)
  • Reference temperature(rr_tref)
  • Soil temperature 5(rr_ts5)
  • scalar wind speed(wind_s)
  • Soil temperature 4(rr_ts4)
  • Wind direction (relative to true north)(mv_wind_d)
  • Soil moisture 1(r_sm1)
  • Home signal(mv_home)
  • Soil moisture 3(r_sm3)
  • Soil moisture 2(r_sm2)
  • Soil heat flow 1(mv_hft1)
  • Soil heat flow 2(mv_hft2)
  • Left air temperature(tair_l)
  • Left relative humidity(mv_hum_l)
  • Soil heat flow 4(mv_hft4)
  • Right relative humidity(mv_hum_r)
  • Soil heat flow 5(mv_hft5)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Soil heat flow 3(mv_hft3)
  • Soil moisture 4(r_sm4)
  • Soil temperature 1(rr_ts1)
  • Soil temperature 2(rr_ts2)
  • Soil moisture 5(r_sm5)
  • Soil temperature 3(rr_ts3)
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Atmospheric pressure(mv_pres)
  • Net radiation(mv_q)

sgp15ebbrE19.b1:
  • Right relative humidity(mv_hum_r)
  • Soil heat flow 1(mv_hft1)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Soil temperature 1(rr_ts1)
  • Soil heat flow 3(mv_hft3)
  • Atmospheric pressure(mv_pres)
  • Soil temperature 2(rr_ts2)
  • Soil heat flow 2(mv_hft2)
  • Left air temperature(tair_l)
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Soil moisture 5(r_sm5)
  • Soil temperature 4(rr_ts4)
  • Net radiation(mv_q)
  • scalar wind speed(wind_s)
  • Reference temperature(rr_tref)
  • Soil moisture 2(r_sm2)
  • Soil temperature 5(rr_ts5)
  • Left relative humidity(mv_hum_l)
  • Soil temperature 3(rr_ts3)
  • Home signal(mv_home)
  • Wind direction (relative to true north)(mv_wind_d)
  • Soil moisture 4(r_sm4)
  • Right air temperature(tair_r)
  • Soil moisture 3(r_sm3)
  • Soil heat flow 4(mv_hft4)
  • Soil heat flow 5(mv_hft5)
  • Soil moisture 1(r_sm1)

sgp5ebbrE19.a1:
  • bottom vapor pressure(vp_bot)
  • Temperature of the top humidity chamber(thum_top)
  • Bottom humidity(hum_bot)
  • Top humidity(hum_top)
  • vector wind speed(res_ws)
  • wind direction (relative to true north)(wind_d)
  • bottom air temperature(tair_bot)
  • Reference Thermistor Temperature(tref)
  • top air temperature(tair_top)
  • scalar wind speed(wind_s)
  • Retrieved pressure profile(pres)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Home signal(home)
  • top vapor pressure(vp_top)
  • net radiation(q)

sgp30ebbrE19.a1:
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • vector wind speed(res_ws)
  • 5 cm soil heat flow, site 1(shf1)
  • soil heat flow, site 2(g2)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • soil heat flow, site 1(g1)
  • 5 cm soil heat flow, site 4(shf4)
  • net radiation(q)
  • Exchange mechanism position indicator (15 to 30 min)(home_30)
  • Exchange mechanism position indicator (0 to 15 min)(home_15)
  • Reference Thermistor Temperature(tref)
  • volumetric soil moisture, site 3(sm3)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • 5 cm soil heat flow, site 5(shf5)
  • average surface soil heat flow(ave_shf)
  • top vapor pressure(vp_top)
  • volumetric soil moisture, site 4(sm4)
  • Soil heat capacity 3(cs3)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • Soil heat capacity 5(cs5)
  • Retrieved pressure profile(pres)
  • Temperature of the top humidity chamber(thum_top)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 5 cm soil heat flow, site 2(shf2)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • Soil heat capacity 2(cs2)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • soil heat flow, site 4(g4)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • volumetric soil moisture, site 2(sm2)
  • Soil heat capacity 1(cs1)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • latent heat flux(e)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • Soil heat capacity 4(cs4)
  • h(h)
  • bottom vapor pressure(vp_bot)
  • wind direction (relative to true north)(wind_d)
  • soil heat flow, site 3(g3)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Top humidity(hum_top)
  • volumetric soil moisture, site 5(sm5)
  • Bottom humidity(hum_bot)
  • soil heat flow, site 5(g5)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • 5 cm soil heat flow, site 3(shf3)
  • scalar wind speed(wind_s)
  • top air temperature(tair_top)
  • bottom air temperature(tair_bot)
  • volumetric soil moisture, site 1(sm1)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)

sgp30ebbrE19.b1:
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • Reference Thermistor Temperature(tref)
  • 5 cm soil heat flow, site 2(shf2)
  • volumetric soil moisture, site 1(sm1)
  • volumetric soil moisture, site 2(sm2)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • Exchange mechanism position indicator (15 to 30 min)(home_30)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)
  • Soil heat capacity 5(cs5)
  • Exchange mechanism position indicator (0 to 15 min)(home_15)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)
  • volumetric soil moisture, site 4(sm4)
  • Soil heat capacity 3(cs3)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • h(h)
  • Temperature of the top humidity chamber(thum_top)
  • scalar wind speed(wind_s)
  • top air temperature(tair_top)
  • Soil heat capacity 4(cs4)
  • Bottom humidity(hum_bot)
  • soil heat flow, site 4(g4)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • top vapor pressure(vp_top)
  • vector wind speed(res_ws)
  • average surface soil heat flow(ave_shf)
  • volumetric soil moisture, site 5(sm5)
  • latent heat flux(e)
  • wind direction (relative to true north)(wind_d)
  • soil heat flow, site 2(g2)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Soil heat capacity 1(cs1)
  • volumetric soil moisture, site 3(sm3)
  • soil heat flow, site 5(g5)
  • soil heat flow, site 1(g1)
  • soil heat flow, site 3(g3)
  • bottom vapor pressure(vp_bot)
  • 5 cm soil heat flow, site 1(shf1)
  • Soil heat capacity 2(cs2)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 5 cm soil heat flow, site 5(shf5)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • Top humidity(hum_top)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • bottom air temperature(tair_bot)
  • net radiation(q)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • 5 cm soil heat flow, site 3(shf3)
  • 5 cm soil heat flow, site 4(shf4)
  • Retrieved pressure profile(pres)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)


Back To Table of Contents

DQRID : D030711.11
Start DateStart TimeEnd DateEnd Time
01/09/2003163001/09/20031700
Subject:
SGP/EBBR/E19 - Soil Temperature #2 Incorrect
DataStreams:sgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1
Description:
Soil Temperature #2 was incorrect during J-panel replacements.
Measurements:sgp15ebbrE19.a1:
  • Soil temperature 2(rr_ts2)

sgp15ebbrE19.b1:
  • Soil temperature 2(rr_ts2)

sgp30ebbrE19.a1:
  • average surface soil heat flow(ave_shf)
  • soil heat flow, site 2(g2)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • latent heat flux(e)
  • h(h)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)

sgp30ebbrE19.b1:
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • h(h)
  • soil heat flow, site 2(g2)
  • average surface soil heat flow(ave_shf)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • latent heat flux(e)


Back To Table of Contents

DQRID : D030711.12
Start DateStart TimeEnd DateEnd Time
01/09/2003170001/09/20031700
02/06/2003173002/06/20031730
04/03/2003183004/03/20031830
Subject:
SGP/EBBR/E19 - All Data Zeros
DataStreams:sgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.a1
Description:
All data (5, 15, and 30 minute) was zero after replacement of sensors and restart of the 
CR10.

This is a common occurrence.
Measurements:sgp15ebbrE19.a1:
  • Right air temperature(tair_r)
  • Reference temperature(rr_tref)
  • Soil temperature 5(rr_ts5)
  • scalar wind speed(wind_s)
  • Soil temperature 4(rr_ts4)
  • Wind direction (relative to true north)(mv_wind_d)
  • Soil moisture 1(r_sm1)
  • Home signal(mv_home)
  • Soil moisture 3(r_sm3)
  • Soil moisture 2(r_sm2)
  • Soil heat flow 1(mv_hft1)
  • Soil heat flow 2(mv_hft2)
  • Left air temperature(tair_l)
  • Left relative humidity(mv_hum_l)
  • Battery(bat)
  • Soil heat flow 4(mv_hft4)
  • Right relative humidity(mv_hum_r)
  • Soil heat flow 5(mv_hft5)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Soil heat flow 3(mv_hft3)
  • Soil moisture 4(r_sm4)
  • Soil temperature 1(rr_ts1)
  • Soil temperature 2(rr_ts2)
  • Soil moisture 5(r_sm5)
  • Soil temperature 3(rr_ts3)
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Atmospheric pressure(mv_pres)
  • Net radiation(mv_q)

sgp15ebbrE19.b1:
  • Right relative humidity(mv_hum_r)
  • Soil heat flow 1(mv_hft1)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Soil temperature 1(rr_ts1)
  • Soil heat flow 3(mv_hft3)
  • Atmospheric pressure(mv_pres)
  • Soil temperature 2(rr_ts2)
  • Soil heat flow 2(mv_hft2)
  • Left air temperature(tair_l)
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Soil moisture 5(r_sm5)
  • Soil temperature 4(rr_ts4)
  • Net radiation(mv_q)
  • scalar wind speed(wind_s)
  • Reference temperature(rr_tref)
  • Soil moisture 2(r_sm2)
  • Soil temperature 5(rr_ts5)
  • Battery(bat)
  • Left relative humidity(mv_hum_l)
  • Soil temperature 3(rr_ts3)
  • Home signal(mv_home)
  • Wind direction (relative to true north)(mv_wind_d)
  • Soil moisture 4(r_sm4)
  • Right air temperature(tair_r)
  • Soil moisture 3(r_sm3)
  • Soil heat flow 4(mv_hft4)
  • Soil heat flow 5(mv_hft5)
  • Soil moisture 1(r_sm1)

sgp5ebbrE19.a1:
  • bottom vapor pressure(vp_bot)
  • Temperature of the top humidity chamber(thum_top)
  • Bottom humidity(hum_bot)
  • Top humidity(hum_top)
  • vector wind speed(res_ws)
  • wind direction (relative to true north)(wind_d)
  • bottom air temperature(tair_bot)
  • Reference Thermistor Temperature(tref)
  • top air temperature(tair_top)
  • scalar wind speed(wind_s)
  • Retrieved pressure profile(pres)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Home signal(home)
  • top vapor pressure(vp_top)
  • net radiation(q)

sgp30ebbrE19.a1:
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • vector wind speed(res_ws)
  • 5 cm soil heat flow, site 1(shf1)
  • soil heat flow, site 2(g2)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • soil heat flow, site 1(g1)
  • 5 cm soil heat flow, site 4(shf4)
  • net radiation(q)
  • Exchange mechanism position indicator (15 to 30 min)(home_30)
  • Exchange mechanism position indicator (0 to 15 min)(home_15)
  • Reference Thermistor Temperature(tref)
  • volumetric soil moisture, site 3(sm3)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • 5 cm soil heat flow, site 5(shf5)
  • average surface soil heat flow(ave_shf)
  • top vapor pressure(vp_top)
  • volumetric soil moisture, site 4(sm4)
  • Soil heat capacity 3(cs3)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • Soil heat capacity 5(cs5)
  • Retrieved pressure profile(pres)
  • Temperature of the top humidity chamber(thum_top)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 5 cm soil heat flow, site 2(shf2)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • Soil heat capacity 2(cs2)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • soil heat flow, site 4(g4)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • volumetric soil moisture, site 2(sm2)
  • Soil heat capacity 1(cs1)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • latent heat flux(e)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • Soil heat capacity 4(cs4)
  • h(h)
  • bottom vapor pressure(vp_bot)
  • wind direction (relative to true north)(wind_d)
  • soil heat flow, site 3(g3)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Top humidity(hum_top)
  • volumetric soil moisture, site 5(sm5)
  • Bottom humidity(hum_bot)
  • soil heat flow, site 5(g5)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • 5 cm soil heat flow, site 3(shf3)
  • scalar wind speed(wind_s)
  • top air temperature(tair_top)
  • bottom air temperature(tair_bot)
  • volumetric soil moisture, site 1(sm1)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)

sgp30ebbrE19.b1:
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • Reference Thermistor Temperature(tref)
  • 5 cm soil heat flow, site 2(shf2)
  • volumetric soil moisture, site 1(sm1)
  • volumetric soil moisture, site 2(sm2)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • Exchange mechanism position indicator (15 to 30 min)(home_30)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)
  • Soil heat capacity 5(cs5)
  • Exchange mechanism position indicator (0 to 15 min)(home_15)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)
  • volumetric soil moisture, site 4(sm4)
  • Soil heat capacity 3(cs3)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • h(h)
  • Temperature of the top humidity chamber(thum_top)
  • scalar wind speed(wind_s)
  • top air temperature(tair_top)
  • Soil heat capacity 4(cs4)
  • Bottom humidity(hum_bot)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • soil heat flow, site 4(g4)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • top vapor pressure(vp_top)
  • vector wind speed(res_ws)
  • average surface soil heat flow(ave_shf)
  • volumetric soil moisture, site 5(sm5)
  • latent heat flux(e)
  • wind direction (relative to true north)(wind_d)
  • soil heat flow, site 2(g2)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Soil heat capacity 1(cs1)
  • volumetric soil moisture, site 3(sm3)
  • soil heat flow, site 5(g5)
  • soil heat flow, site 1(g1)
  • soil heat flow, site 3(g3)
  • bottom vapor pressure(vp_bot)
  • 5 cm soil heat flow, site 1(shf1)
  • Soil heat capacity 2(cs2)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 5 cm soil heat flow, site 5(shf5)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • Top humidity(hum_top)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • bottom air temperature(tair_bot)
  • net radiation(q)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • 5 cm soil heat flow, site 3(shf3)
  • 5 cm soil heat flow, site 4(shf4)
  • Retrieved pressure profile(pres)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)


Back To Table of Contents

DQRID : D030711.13
Start DateStart TimeEnd DateEnd Time
01/09/2003173001/09/20031730
Subject:
SGP/EBBR/E19 - Several Soil Measurements Incorrect
DataStreams:sgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1
Description:
During replacement of a J-panel, several soil measurements were incorrect.
Measurements:sgp15ebbrE19.a1:
  • Soil temperature 3(rr_ts3)

sgp15ebbrE19.b1:
  • Soil temperature 3(rr_ts3)

sgp30ebbrE19.a1:
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • average surface soil heat flow(ave_shf)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • latent heat flux(e)
  • soil heat flow, site 5(g5)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • h(h)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • soil heat flow, site 3(g3)
  • soil heat flow, site 2(g2)
  • soil heat flow, site 1(g1)
  • soil heat flow, site 4(g4)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)

sgp30ebbrE19.b1:
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • h(h)
  • soil heat flow, site 2(g2)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • soil heat flow, site 4(g4)
  • soil heat flow, site 5(g5)
  • soil heat flow, site 1(g1)
  • soil heat flow, site 3(g3)
  • average surface soil heat flow(ave_shf)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)
  • latent heat flux(e)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)


Back To Table of Contents

DQRID : D030711.14
Start DateStart TimeEnd DateEnd Time
01/26/2003170001/26/20031700
01/26/2003180001/26/20031800
Subject:
SGP/EBBR/E19 -Several Soil Measurements Were Incorrect
DataStreams:sgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1
Description:
Several soil measurements were incorrect during replacement of soil temperature probe #2.
Measurements:sgp15ebbrE19.a1:
  • Soil temperature 2(rr_ts2)

sgp15ebbrE19.b1:
  • Soil temperature 2(rr_ts2)

sgp30ebbrE19.a1:
  • average surface soil heat flow(ave_shf)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • latent heat flux(e)
  • soil heat flow, site 5(g5)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • h(h)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • soil heat flow, site 3(g3)
  • soil heat flow, site 2(g2)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • soil heat flow, site 1(g1)
  • soil heat flow, site 4(g4)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)

sgp30ebbrE19.b1:
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • h(h)
  • soil heat flow, site 2(g2)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • soil heat flow, site 4(g4)
  • soil heat flow, site 5(g5)
  • soil heat flow, site 1(g1)
  • soil heat flow, site 3(g3)
  • average surface soil heat flow(ave_shf)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • latent heat flux(e)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)


Back To Table of Contents

DQRID : D030711.15
Start DateStart TimeEnd DateEnd Time
01/09/2003171501/09/20031715
Subject:
SGP/EBBR/E19 - Soil Temperature #3 Incorrect
DataStreams:sgp15ebbrE19.a1, sgp15ebbrE19.b1
Description:
Soil Temperature #3 was incorrect only for the 15 minute data.
Measurements:sgp15ebbrE19.a1:
  • Soil temperature 3(rr_ts3)

sgp15ebbrE19.b1:
  • Soil temperature 3(rr_ts3)


Back To Table of Contents

DQRID : D030711.17
Start DateStart TimeEnd DateEnd Time
01/09/2003173004/03/20031900
Subject:
SGP/EBBR/E19 - Soil Heat Flow #3 Incorrect
DataStreams:sgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1
Description:
Soil Heat Flow #3 was a flat -24 W/m2 for most of this period.
Measurements:sgp15ebbrE19.a1:
  • Soil heat flow 3(mv_hft3)

sgp15ebbrE19.b1:
  • Soil heat flow 3(mv_hft3)

sgp30ebbrE19.a1:
  • average surface soil heat flow(ave_shf)
  • soil heat flow, site 3(g3)
  • latent heat flux(e)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)
  • 5 cm soil heat flow, site 3(shf3)
  • h(h)

sgp30ebbrE19.b1:
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)
  • h(h)
  • 5 cm soil heat flow, site 3(shf3)
  • soil heat flow, site 3(g3)
  • average surface soil heat flow(ave_shf)
  • latent heat flux(e)


Back To Table of Contents

DQRID : D030711.18
Start DateStart TimeEnd DateEnd Time
02/06/2003170002/06/20031715
Subject:
SGP/EBBR/E19 - Soil Temperature #2 Incorrect
DataStreams:sgp15ebbrE19.a1, sgp15ebbrE19.b1
Description:
Only the 15 minute Soil Temperature #2 measurements were incorrect.
Measurements:sgp15ebbrE19.a1:
  • Soil temperature 2(rr_ts2)

sgp15ebbrE19.b1:
  • Soil temperature 2(rr_ts2)


Back To Table of Contents

DQRID : D030711.19
Start DateStart TimeEnd DateEnd Time
04/03/2003190004/03/20031900
Subject:
SGP/EBBR/E19 - Soil Heat Flow#1 Incorrect
DataStreams:sgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1
Description:
Soil Heat Flow #1 was incorrect during replacement of some other sensors.
Measurements:sgp15ebbrE19.a1:
  • Soil heat flow 1(mv_hft1)

sgp15ebbrE19.b1:
  • Soil heat flow 1(mv_hft1)

sgp30ebbrE19.a1:
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • average surface soil heat flow(ave_shf)
  • 5 cm soil heat flow, site 1(shf1)
  • soil heat flow, site 1(g1)
  • latent heat flux(e)
  • h(h)

sgp30ebbrE19.b1:
  • h(h)
  • soil heat flow, site 1(g1)
  • average surface soil heat flow(ave_shf)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • 5 cm soil heat flow, site 1(shf1)
  • latent heat flux(e)


Back To Table of Contents

DQRID : D030711.20
Start DateStart TimeEnd DateEnd Time
05/15/2003160005/15/20031700
Subject:
SGP/EBBR/E19 - Many Measurements Incorrect
DataStreams:sgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.a1
Description:
Many measurements were incorrect during replacement of a multiplexer.
Measurements:sgp15ebbrE19.a1:
  • Right air temperature(tair_r)
  • Reference temperature(rr_tref)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Soil temperature 1(rr_ts1)
  • Soil temperature 5(rr_ts5)
  • Soil temperature 2(rr_ts2)
  • Soil temperature 4(rr_ts4)
  • Soil temperature 3(rr_ts3)
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Left air temperature(tair_l)

sgp15ebbrE19.b1:
  • Soil temperature 3(rr_ts3)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Soil temperature 1(rr_ts1)
  • Soil temperature 2(rr_ts2)
  • Left air temperature(tair_l)
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Soil temperature 4(rr_ts4)
  • Right air temperature(tair_r)
  • Reference temperature(rr_tref)
  • Soil temperature 5(rr_ts5)

sgp5ebbrE19.a1:
  • Temperature of the top humidity chamber(thum_top)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • bottom air temperature(tair_bot)
  • Reference Thermistor Temperature(tref)
  • top air temperature(tair_top)

sgp30ebbrE19.a1:
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • latent heat flux(e)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • h(h)
  • soil heat flow, site 3(g3)
  • soil heat flow, site 2(g2)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • soil heat flow, site 1(g1)
  • average surface soil heat flow(ave_shf)
  • soil heat flow, site 5(g5)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • Temperature of the top humidity chamber(thum_top)
  • top air temperature(tair_top)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • soil heat flow, site 4(g4)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • bottom air temperature(tair_bot)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)

sgp30ebbrE19.b1:
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • soil heat flow, site 2(g2)
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • soil heat flow, site 5(g5)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • soil heat flow, site 1(g1)
  • soil heat flow, site 3(g3)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • h(h)
  • Temperature of the top humidity chamber(thum_top)
  • top air temperature(tair_top)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • bottom air temperature(tair_bot)
  • soil heat flow, site 4(g4)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • average surface soil heat flow(ave_shf)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • latent heat flux(e)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)


Back To Table of Contents

DQRID : D030712.20
Start DateStart TimeEnd DateEnd Time
05/01/2003144005/01/20031515
Subject:
SGP/EBBR/E19 -  6 Month Checks
DataStreams:sgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.a1
Description:
Many measurements were incorrect while the AEM was positioned with both aspirators at the 
same height for comparison checks.
Measurements:sgp15ebbrE19.a1:
  • Right relative humidity(mv_hum_r)
  • Right air temperature(tair_r)
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Left air temperature(tair_l)
  • Left relative humidity(mv_hum_l)

sgp15ebbrE19.b1:
  • Right air temperature(tair_r)
  • Right relative humidity(mv_hum_r)
  • Left relative humidity(mv_hum_l)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Left air temperature(tair_l)
  • Temperature of left humidity sensor chamber(rr_thum_l)

sgp5ebbrE19.a1:
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • bottom vapor pressure(vp_bot)
  • top vapor pressure(vp_top)
  • Temperature of the top humidity chamber(thum_top)
  • Bottom humidity(hum_bot)
  • bottom air temperature(tair_bot)
  • Top humidity(hum_top)
  • top air temperature(tair_top)

sgp30ebbrE19.a1:
  • top vapor pressure(vp_top)
  • latent heat flux(e)
  • h(h)
  • bottom vapor pressure(vp_bot)
  • Temperature of the top humidity chamber(thum_top)
  • top air temperature(tair_top)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • bottom air temperature(tair_bot)
  • Top humidity(hum_top)
  • Bottom humidity(hum_bot)

sgp30ebbrE19.b1:
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • h(h)
  • Temperature of the top humidity chamber(thum_top)
  • top air temperature(tair_top)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Top humidity(hum_top)
  • Bottom humidity(hum_bot)
  • bottom air temperature(tair_bot)
  • top vapor pressure(vp_top)
  • bottom vapor pressure(vp_bot)
  • latent heat flux(e)


Back To Table of Contents

DQRID : D030919.10
Start DateStart TimeEnd DateEnd Time
08/30/2003140009/04/20031445
Subject:
SGP/EBBR/E19  - Water in Net Radiometer; Surface Fluxes are Incorrect
DataStreams:sgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.a1
Description:
The net radiometer top dome was punctured by migrating birds.  This alone does not cause 
incorrect net radiation measurements.  However, when it subsequently rains, water collects 
in the bottom dome, causing the net radiation measurements to be too large during the 
day and too small at night.
Measurements:sgp15ebbrE19.a1:
  • Net radiation(mv_q)

sgp15ebbrE19.b1:
  • Net radiation(mv_q)

sgp5ebbrE19.a1:
  • net radiation(q)

sgp30ebbrE19.a1:
  • latent heat flux(e)
  • h(h)
  • net radiation(q)

sgp30ebbrE19.b1:
  • h(h)
  • latent heat flux(e)
  • net radiation(q)


Back To Table of Contents

DQRID : D030922.6
Start DateStart TimeEnd DateEnd Time
09/17/2003183009/17/20032000
09/17/2003223009/18/20030100
Subject:
SGP/EBBR/E19 - Tair Right Offscale
DataStreams:sgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.a1
Description:
The right air temperature data were going offscale.  This 
                  affected the calculation of sensible and latent heat 
                  fluxes.
Measurements:sgp15ebbrE19.a1:
  • Right air temperature(tair_r)

sgp15ebbrE19.b1:
  • Right air temperature(tair_r)

sgp5ebbrE19.a1:
  • bottom air temperature(tair_bot)
  • top air temperature(tair_top)

sgp30ebbrE19.a1:
  • top air temperature(tair_top)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • latent heat flux(e)
  • bottom air temperature(tair_bot)
  • h(h)

sgp30ebbrE19.b1:
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • h(h)
  • top air temperature(tair_top)
  • latent heat flux(e)
  • bottom air temperature(tair_bot)


Back To Table of Contents

DQRID : D031110.8
Start DateStart TimeEnd DateEnd Time
11/03/2003200011/13/20031615
Subject:
SGP/EBBR/E19 - Battery Voltage Too Low, Most Measurements Incorrect
DataStreams:sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1
Description:
Battery voltage dropped below levels required to sustain the 
               production of valid data.  The AC charger unit had failed.  
               This prevented charging of the battery during the night and
               prevented the solar panel regulator from properly charging 
               the battery during the day.
Measurements:sgp5ebbrE19.b1:
  • Retrieved pressure profile(pres)
  • scalar wind speed(wind_s)
  • Reference Thermistor Temperature(tref)
  • net radiation(q)
  • time(time)
  • top air temperature(tair_top)
  • Temperature of the top humidity chamber(thum_top)
  • bottom vapor pressure(vp_bot)
  • wind direction (relative to true north)(wind_d)
  • top vapor pressure(vp_top)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • Bottom humidity(hum_bot)
  • vector wind speed(res_ws)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • bottom air temperature(tair_bot)
  • Top humidity(hum_top)
  • Home signal(home)

sgp15ebbrE19.b1:
  • Right relative humidity(mv_hum_r)
  • Soil heat flow 1(mv_hft1)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Soil temperature 1(rr_ts1)
  • Soil heat flow 3(mv_hft3)
  • Atmospheric pressure(mv_pres)
  • Soil temperature 2(rr_ts2)
  • Soil heat flow 2(mv_hft2)
  • Left air temperature(tair_l)
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Soil moisture 5(r_sm5)
  • Soil temperature 4(rr_ts4)
  • Net radiation(mv_q)
  • scalar wind speed(wind_s)
  • Reference temperature(rr_tref)
  • Soil moisture 2(r_sm2)
  • Soil temperature 5(rr_ts5)
  • Left relative humidity(mv_hum_l)
  • Soil temperature 3(rr_ts3)
  • Home signal(mv_home)
  • Wind direction (relative to true north)(mv_wind_d)
  • Soil moisture 4(r_sm4)
  • Right air temperature(tair_r)
  • Soil moisture 3(r_sm3)
  • Soil heat flow 4(mv_hft4)
  • Soil heat flow 5(mv_hft5)
  • Soil moisture 1(r_sm1)

sgp30ebbrE19.b1:
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • Reference Thermistor Temperature(tref)
  • 5 cm soil heat flow, site 2(shf2)
  • volumetric soil moisture, site 1(sm1)
  • volumetric soil moisture, site 2(sm2)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • Exchange mechanism position indicator (15 to 30 min)(home_30)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)
  • Soil heat capacity 5(cs5)
  • Exchange mechanism position indicator (0 to 15 min)(home_15)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • Soil heat capacity 3(cs3)
  • volumetric soil moisture, site 4(sm4)
  • h(h)
  • scalar wind speed(wind_s)
  • Temperature of the top humidity chamber(thum_top)
  • top air temperature(tair_top)
  • Soil heat capacity 4(cs4)
  • Bottom humidity(hum_bot)
  • soil heat flow, site 4(g4)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • top vapor pressure(vp_top)
  • vector wind speed(res_ws)
  • average surface soil heat flow(ave_shf)
  • volumetric soil moisture, site 5(sm5)
  • latent heat flux(e)
  • wind direction (relative to true north)(wind_d)
  • soil heat flow, site 2(g2)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Soil heat capacity 1(cs1)
  • volumetric soil moisture, site 3(sm3)
  • soil heat flow, site 5(g5)
  • soil heat flow, site 1(g1)
  • soil heat flow, site 3(g3)
  • 5 cm soil heat flow, site 1(shf1)
  • bottom vapor pressure(vp_bot)
  • Soil heat capacity 2(cs2)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 5 cm soil heat flow, site 5(shf5)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • Top humidity(hum_top)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • bottom air temperature(tair_bot)
  • net radiation(q)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • 5 cm soil heat flow, site 3(shf3)
  • 5 cm soil heat flow, site 4(shf4)
  • Retrieved pressure profile(pres)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)


Back To Table of Contents

DQRID : D031117.2
Start DateStart TimeEnd DateEnd Time
11/14/2003211511/26/20031600
Subject:
SGP/EBBR/E19 - AEM Stuck In One Position, Surface Fluxes Incorrect
DataStreams:sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1
Description:
The AEM home 15 and 30 outputs were both 40 mv
because the AEM was not exchanging.

Sensible and Latent Heat Fluxes are therefore
incorrect.
Measurements:sgp5ebbrE19.b1:
  • bottom vapor pressure(vp_bot)
  • Bottom humidity(hum_bot)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • top air temperature(tair_top)
  • top vapor pressure(vp_top)
  • bottom air temperature(tair_bot)
  • Top humidity(hum_top)
  • Temperature of the top humidity chamber(thum_top)

sgp15ebbrE19.b1:
  • Right air temperature(tair_r)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Left air temperature(tair_l)
  • Temperature of left humidity sensor chamber(rr_thum_l)

sgp30ebbrE19.b1:
  • top vapor pressure(vp_top)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • h(h)
  • Temperature of the top humidity chamber(thum_top)
  • top air temperature(tair_top)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • bottom vapor pressure(vp_bot)
  • Top humidity(hum_top)
  • latent heat flux(e)
  • Bottom humidity(hum_bot)
  • bottom air temperature(tair_bot)


Back To Table of Contents

DQRID : D031124.11
Start DateStart TimeEnd DateEnd Time
10/10/2003150510/10/20031755
Subject:
SGP/EBBR/E19 - Data Missing During Recalibration Changeout
DataStreams:sgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30baebbrE19.c1, sgp30ebbrE19.a1, sgp30ebbrE19.b1,
sgp5ebbrE19.a1, sgpebbrE19.00, sgpebbrE19.01
Description:
Data was missing during the recalibration changeout.
Measurements:sgp15ebbrE19.a1:
  • Reference temperature(rr_tref)
  • Soil temperature 5(rr_ts5)
  • scalar wind speed(wind_s)
  • Soil temperature 4(rr_ts4)
  • Wind direction (relative to true north)(mv_wind_d)
  • Soil moisture 1(r_sm1)
  • Home signal(mv_home)
  • Soil moisture 3(r_sm3)
  • Soil moisture 2(r_sm2)
  • Left air temperature(tair_l)
  • Left relative humidity(mv_hum_l)
  • Soil heat flow 4(mv_hft4)
  • Right relative humidity(mv_hum_r)
  • Soil heat flow 5(mv_hft5)
  • Soil heat flow 3(mv_hft3)
  • Soil moisture 4(r_sm4)
  • Soil moisture 5(r_sm5)
  • Signature(signature)
  • Dummy altitude for Zeb(alt)
  • base time(base_time)
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Atmospheric pressure(mv_pres)
  • lat(lat)
  • Right air temperature(tair_r)
  • Soil heat flow 1(mv_hft1)
  • Soil heat flow 2(mv_hft2)
  • Battery(bat)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Soil temperature 1(rr_ts1)
  • Soil temperature 2(rr_ts2)
  • Time offset of tweaks from base_time(time_offset)
  • lon(lon)
  • Soil temperature 3(rr_ts3)
  • Net radiation(mv_q)

sgp15ebbrE19.b1:
  • Right relative humidity(mv_hum_r)
  • Soil heat flow 1(mv_hft1)
  • Soil temperature 1(rr_ts1)
  • Soil heat flow 3(mv_hft3)
  • Soil temperature 2(rr_ts2)
  • lon(lon)
  • Soil heat flow 2(mv_hft2)
  • Left air temperature(tair_l)
  • Reference temperature(rr_tref)
  • Battery(bat)
  • Soil temperature 3(rr_ts3)
  • Home signal(mv_home)
  • Right air temperature(tair_r)
  • Dummy altitude for Zeb(alt)
  • Soil heat flow 4(mv_hft4)
  • Soil heat flow 5(mv_hft5)
  • lat(lat)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Atmospheric pressure(mv_pres)
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Soil moisture 5(r_sm5)
  • Soil temperature 4(rr_ts4)
  • Net radiation(mv_q)
  • scalar wind speed(wind_s)
  • Time offset of tweaks from base_time(time_offset)
  • Soil moisture 2(r_sm2)
  • Soil temperature 5(rr_ts5)
  • Left relative humidity(mv_hum_l)
  • Wind direction (relative to true north)(mv_wind_d)
  • Signature(signature)
  • Soil moisture 4(r_sm4)
  • Soil moisture 3(r_sm3)
  • base time(base_time)
  • Soil moisture 1(r_sm1)

sgp30baebbrE19.c1:
  • 5 cm soil heat flow, site 3(shf3)
  • top vapor pressure(vp_top)
  • 5 cm soil heat flow, site 1(shf1)
  • Dummy altitude for Zeb(alt)
  • Exchange mechanism position indicator (15 to 30 min)(home_30)
  • Retrieved pressure profile(pres)
  • lon(lon)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • wind direction (relative to true north)(wind_d)
  • bottom vapor pressure(vp_bot)
  • Best estimate Latent heat flux(ecomb)
  • 5 cm soil heat flow, site 4(shf4)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • h(h)
  • net radiation(q)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 5 cm soil heat flow, site 5(shf5)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • 5 cm soil heat flow, site 2(shf2)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)
  • base time(base_time)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • Aerodynamic Latent heat flux(ale)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • average surface soil heat flow(ave_shf)
  • Exchange mechanism position indicator (0 to 15 min)(home_15)
  • lat(lat)
  • bottom air temperature(tair_bot)
  • soil heat flow, site 4(g4)
  • soil heat flow, site 5(g5)
  • Best estimate Sensible heat flux(hcomb)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)
  • Temperature of the top humidity chamber(thum_top)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • soil heat flow, site 1(g1)
  • Aerodynamic Sensible heat flux(ah)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • Reference Thermistor Temperature(tref)
  • top air temperature(tair_top)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • volumetric soil moisture, site 5(sm5)
  • soil heat flow, site 3(g3)
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • latent heat flux(e)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • soil heat flow, site 2(g2)
  • Bottom humidity(hum_bot)
  • Estimated height of the vegetation.(vegetation_height)
  • volumetric soil moisture, site 1(sm1)
  • Friction velocity(austar)
  • scalar wind speed(wind_s)
  • Soil heat capacity 1(cs1)
  • Soil heat capacity 5(cs5)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • Soil heat capacity 3(cs3)
  • volumetric soil moisture, site 3(sm3)
  • vector wind speed(res_ws)
  • Soil heat capacity 4(cs4)
  • Time offset of tweaks from base_time(time_offset)
  • Soil heat capacity 2(cs2)
  • Top humidity(hum_top)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • volumetric soil moisture, site 2(sm2)
  • volumetric soil moisture, site 4(sm4)

sgpebbrE19.00:
  • null(Raw data stream - documentation not supported)

sgp5ebbrE19.a1:
  • bottom vapor pressure(vp_bot)
  • Temperature of the top humidity chamber(thum_top)
  • Bottom humidity(hum_bot)
  • base time(base_time)
  • wind direction (relative to true north)(wind_d)
  • bottom air temperature(tair_bot)
  • top air temperature(tair_top)
  • scalar wind speed(wind_s)
  • top vapor pressure(vp_top)
  • lat(lat)
  • lon(lon)
  • vector wind speed(res_ws)
  • Top humidity(hum_top)
  • Dummy altitude for Zeb(alt)
  • Reference Thermistor Temperature(tref)
  • Retrieved pressure profile(pres)
  • Time offset of tweaks from base_time(time_offset)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Home signal(home)
  • net radiation(q)

sgp30ebbrE19.a1:
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • vector wind speed(res_ws)
  • soil heat flow, site 2(g2)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • soil heat flow, site 1(g1)
  • net radiation(q)
  • Exchange mechanism position indicator (0 to 15 min)(home_15)
  • volumetric soil moisture, site 3(sm3)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • average surface soil heat flow(ave_shf)
  • volumetric soil moisture, site 4(sm4)
  • Soil heat capacity 3(cs3)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • Soil heat capacity 5(cs5)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • base time(base_time)
  • Soil heat capacity 2(cs2)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)
  • volumetric soil moisture, site 2(sm2)
  • Soil heat capacity 1(cs1)
  • latent heat flux(e)
  • bottom vapor pressure(vp_bot)
  • wind direction (relative to true north)(wind_d)
  • soil heat flow, site 3(g3)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Top humidity(hum_top)
  • lat(lat)
  • soil heat flow, site 5(g5)
  • scalar wind speed(wind_s)
  • bottom air temperature(tair_bot)
  • volumetric soil moisture, site 1(sm1)
  • Dummy altitude for Zeb(alt)
  • 5 cm soil heat flow, site 1(shf1)
  • 5 cm soil heat flow, site 4(shf4)
  • Exchange mechanism position indicator (15 to 30 min)(home_30)
  • Reference Thermistor Temperature(tref)
  • 5 cm soil heat flow, site 5(shf5)
  • top vapor pressure(vp_top)
  • lon(lon)
  • Retrieved pressure profile(pres)
  • Temperature of the top humidity chamber(thum_top)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 5 cm soil heat flow, site 2(shf2)
  • soil heat flow, site 4(g4)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • Time offset of tweaks from base_time(time_offset)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • Soil heat capacity 4(cs4)
  • h(h)
  • volumetric soil moisture, site 5(sm5)
  • Bottom humidity(hum_bot)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • 5 cm soil heat flow, site 3(shf3)
  • top air temperature(tair_top)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)

sgpebbrE19.01:
  • null(Raw data stream - documentation not supported)

sgp30ebbrE19.b1:
  • Dummy altitude for Zeb(alt)
  • Reference Thermistor Temperature(tref)
  • 5 cm soil heat flow, site 2(shf2)
  • volumetric soil moisture, site 1(sm1)
  • volumetric soil moisture, site 2(sm2)
  • Exchange mechanism position indicator (15 to 30 min)(home_30)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)
  • Exchange mechanism position indicator (0 to 15 min)(home_15)
  • volumetric soil moisture, site 4(sm4)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • h(h)
  • scalar wind speed(wind_s)
  • top air temperature(tair_top)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • top vapor pressure(vp_top)
  • average surface soil heat flow(ave_shf)
  • volumetric soil moisture, site 5(sm5)
  • latent heat flux(e)
  • lat(lat)
  • wind direction (relative to true north)(wind_d)
  • soil heat flow, site 2(g2)
  • Soil heat capacity 1(cs1)
  • base time(base_time)
  • soil heat flow, site 5(g5)
  • soil heat flow, site 3(g3)
  • Soil heat capacity 2(cs2)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 5 cm soil heat flow, site 5(shf5)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • bottom air temperature(tair_bot)
  • net radiation(q)
  • 5 cm soil heat flow, site 4(shf4)
  • Retrieved pressure profile(pres)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • Soil heat capacity 5(cs5)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)
  • Soil heat capacity 3(cs3)
  • Temperature of the top humidity chamber(thum_top)
  • Soil heat capacity 4(cs4)
  • Bottom humidity(hum_bot)
  • soil heat flow, site 4(g4)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • lon(lon)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • vector wind speed(res_ws)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Time offset of tweaks from base_time(time_offset)
  • volumetric soil moisture, site 3(sm3)
  • soil heat flow, site 1(g1)
  • 5 cm soil heat flow, site 1(shf1)
  • bottom vapor pressure(vp_bot)
  • Top humidity(hum_top)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • 5 cm soil heat flow, site 3(shf3)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)


Back To Table of Contents

DQRID : D031204.17
Start DateStart TimeEnd DateEnd Time
10/10/2003153010/10/20031600
Subject:
SGP/EBBR/E19 - 6 Month Checks
DataStreams:sgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.a1
Description:
Many measurements were incorrect while the AEM was positioned with both aspirators at the 
same height for comparison checks.
Measurements:sgp15ebbrE19.a1:
  • Right relative humidity(mv_hum_r)
  • Right air temperature(tair_r)
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Left air temperature(tair_l)
  • Left relative humidity(mv_hum_l)

sgp15ebbrE19.b1:
  • Right air temperature(tair_r)
  • Right relative humidity(mv_hum_r)
  • Left relative humidity(mv_hum_l)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Left air temperature(tair_l)
  • Temperature of left humidity sensor chamber(rr_thum_l)

sgp5ebbrE19.a1:
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • bottom vapor pressure(vp_bot)
  • top vapor pressure(vp_top)
  • Temperature of the top humidity chamber(thum_top)
  • Bottom humidity(hum_bot)
  • bottom air temperature(tair_bot)
  • Top humidity(hum_top)
  • top air temperature(tair_top)

sgp30ebbrE19.a1:
  • top vapor pressure(vp_top)
  • latent heat flux(e)
  • h(h)
  • bottom vapor pressure(vp_bot)
  • Temperature of the top humidity chamber(thum_top)
  • top air temperature(tair_top)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • bottom air temperature(tair_bot)
  • Top humidity(hum_top)
  • Bottom humidity(hum_bot)

sgp30ebbrE19.b1:
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • h(h)
  • Temperature of the top humidity chamber(thum_top)
  • top air temperature(tair_top)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Top humidity(hum_top)
  • Bottom humidity(hum_bot)
  • bottom air temperature(tair_bot)
  • top vapor pressure(vp_top)
  • bottom vapor pressure(vp_bot)
  • latent heat flux(e)


Back To Table of Contents

DQRID : D031204.18
Start DateStart TimeEnd DateEnd Time
10/16/2003150010/16/20031530
Subject:
SGP/EBBR/E19 - 6 Month Checks
DataStreams:sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1
Description:
Many measurements were incorrect while the AEM was positioned with both aspirators at the 
same height for comparison checks.
Measurements:sgp5ebbrE19.b1:
  • bottom vapor pressure(vp_bot)
  • Bottom humidity(hum_bot)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • top vapor pressure(vp_top)
  • top air temperature(tair_top)
  • bottom air temperature(tair_bot)
  • Top humidity(hum_top)
  • Temperature of the top humidity chamber(thum_top)

sgp15ebbrE19.b1:
  • Right air temperature(tair_r)
  • Right relative humidity(mv_hum_r)
  • Left relative humidity(mv_hum_l)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Left air temperature(tair_l)
  • Temperature of left humidity sensor chamber(rr_thum_l)

sgp30ebbrE19.b1:
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • h(h)
  • Temperature of the top humidity chamber(thum_top)
  • top air temperature(tair_top)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Top humidity(hum_top)
  • Bottom humidity(hum_bot)
  • bottom air temperature(tair_bot)
  • top vapor pressure(vp_top)
  • bottom vapor pressure(vp_bot)
  • latent heat flux(e)


Back To Table of Contents

DQRID : D031204.2
Start DateStart TimeEnd DateEnd Time
10/10/2003180010/10/20031800
Subject:
SGP/EBBR/E19 - Data All Zeros
DataStreams:sgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.a1
Description:
The time period after restart of the datalogger produced all
zero data.
Measurements:sgp15ebbrE19.a1:
  • Reference temperature(rr_tref)
  • Soil temperature 5(rr_ts5)
  • scalar wind speed(wind_s)
  • Soil temperature 4(rr_ts4)
  • Wind direction (relative to true north)(mv_wind_d)
  • Soil moisture 1(r_sm1)
  • Home signal(mv_home)
  • Soil moisture 3(r_sm3)
  • Soil moisture 2(r_sm2)
  • Left air temperature(tair_l)
  • Left relative humidity(mv_hum_l)
  • Soil heat flow 4(mv_hft4)
  • Soil heat flow 5(mv_hft5)
  • Right relative humidity(mv_hum_r)
  • Soil moisture 4(r_sm4)
  • Soil heat flow 3(mv_hft3)
  • Soil moisture 5(r_sm5)
  • Signature(signature)
  • Dummy altitude for Zeb(alt)
  • base time(base_time)
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Atmospheric pressure(mv_pres)
  • lat(lat)
  • Right air temperature(tair_r)
  • Soil heat flow 1(mv_hft1)
  • Soil heat flow 2(mv_hft2)
  • Battery(bat)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Soil temperature 1(rr_ts1)
  • Soil temperature 2(rr_ts2)
  • Time offset of tweaks from base_time(time_offset)
  • lon(lon)
  • Soil temperature 3(rr_ts3)
  • Net radiation(mv_q)

sgp15ebbrE19.b1:
  • Right relative humidity(mv_hum_r)
  • Soil heat flow 1(mv_hft1)
  • Soil temperature 1(rr_ts1)
  • Soil heat flow 3(mv_hft3)
  • Soil temperature 2(rr_ts2)
  • Soil heat flow 2(mv_hft2)
  • Left air temperature(tair_l)
  • Reference temperature(rr_tref)
  • Battery(bat)
  • Soil temperature 3(rr_ts3)
  • Home signal(mv_home)
  • Right air temperature(tair_r)
  • Soil heat flow 4(mv_hft4)
  • Soil heat flow 5(mv_hft5)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Atmospheric pressure(mv_pres)
  • Soil moisture 5(r_sm5)
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Soil temperature 4(rr_ts4)
  • Net radiation(mv_q)
  • scalar wind speed(wind_s)
  • Soil moisture 2(r_sm2)
  • Soil temperature 5(rr_ts5)
  • Left relative humidity(mv_hum_l)
  • Wind direction (relative to true north)(mv_wind_d)
  • Soil moisture 4(r_sm4)
  • Soil moisture 3(r_sm3)
  • Soil moisture 1(r_sm1)

sgp5ebbrE19.a1:
  • bottom vapor pressure(vp_bot)
  • Temperature of the top humidity chamber(thum_top)
  • Bottom humidity(hum_bot)
  • wind direction (relative to true north)(wind_d)
  • bottom air temperature(tair_bot)
  • top air temperature(tair_top)
  • scalar wind speed(wind_s)
  • top vapor pressure(vp_top)
  • vector wind speed(res_ws)
  • Top humidity(hum_top)
  • Reference Thermistor Temperature(tref)
  • Retrieved pressure profile(pres)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Home signal(home)
  • net radiation(q)

sgp30ebbrE19.a1:
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • vector wind speed(res_ws)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • soil heat flow, site 2(g2)
  • soil heat flow, site 1(g1)
  • net radiation(q)
  • Exchange mechanism position indicator (0 to 15 min)(home_15)
  • volumetric soil moisture, site 3(sm3)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • average surface soil heat flow(ave_shf)
  • Soil heat capacity 3(cs3)
  • volumetric soil moisture, site 4(sm4)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • Soil heat capacity 5(cs5)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • base time(base_time)
  • Soil heat capacity 2(cs2)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)
  • volumetric soil moisture, site 2(sm2)
  • Soil heat capacity 1(cs1)
  • latent heat flux(e)
  • bottom vapor pressure(vp_bot)
  • wind direction (relative to true north)(wind_d)
  • soil heat flow, site 3(g3)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Top humidity(hum_top)
  • lat(lat)
  • soil heat flow, site 5(g5)
  • scalar wind speed(wind_s)
  • bottom air temperature(tair_bot)
  • volumetric soil moisture, site 1(sm1)
  • Dummy altitude for Zeb(alt)
  • 5 cm soil heat flow, site 1(shf1)
  • 5 cm soil heat flow, site 4(shf4)
  • Exchange mechanism position indicator (15 to 30 min)(home_30)
  • Reference Thermistor Temperature(tref)
  • 5 cm soil heat flow, site 5(shf5)
  • top vapor pressure(vp_top)
  • lon(lon)
  • Retrieved pressure profile(pres)
  • Temperature of the top humidity chamber(thum_top)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 5 cm soil heat flow, site 2(shf2)
  • soil heat flow, site 4(g4)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • Time offset of tweaks from base_time(time_offset)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • Soil heat capacity 4(cs4)
  • h(h)
  • volumetric soil moisture, site 5(sm5)
  • Bottom humidity(hum_bot)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • 5 cm soil heat flow, site 3(shf3)
  • top air temperature(tair_top)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)

sgp30ebbrE19.b1:
  • Reference Thermistor Temperature(tref)
  • 5 cm soil heat flow, site 2(shf2)
  • volumetric soil moisture, site 1(sm1)
  • volumetric soil moisture, site 2(sm2)
  • Exchange mechanism position indicator (15 to 30 min)(home_30)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)
  • Exchange mechanism position indicator (0 to 15 min)(home_15)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • volumetric soil moisture, site 4(sm4)
  • h(h)
  • scalar wind speed(wind_s)
  • top air temperature(tair_top)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • top vapor pressure(vp_top)
  • average surface soil heat flow(ave_shf)
  • volumetric soil moisture, site 5(sm5)
  • latent heat flux(e)
  • wind direction (relative to true north)(wind_d)
  • soil heat flow, site 2(g2)
  • Soil heat capacity 1(cs1)
  • soil heat flow, site 5(g5)
  • soil heat flow, site 3(g3)
  • Soil heat capacity 2(cs2)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 5 cm soil heat flow, site 5(shf5)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • bottom air temperature(tair_bot)
  • net radiation(q)
  • 5 cm soil heat flow, site 4(shf4)
  • Retrieved pressure profile(pres)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • Soil heat capacity 5(cs5)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)
  • Soil heat capacity 3(cs3)
  • Temperature of the top humidity chamber(thum_top)
  • Soil heat capacity 4(cs4)
  • Bottom humidity(hum_bot)
  • soil heat flow, site 4(g4)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • vector wind speed(res_ws)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • volumetric soil moisture, site 3(sm3)
  • soil heat flow, site 1(g1)
  • 5 cm soil heat flow, site 1(shf1)
  • bottom vapor pressure(vp_bot)
  • Top humidity(hum_top)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • 5 cm soil heat flow, site 3(shf3)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)


Back To Table of Contents

DQRID : D040106.1
Start DateStart TimeEnd DateEnd Time
12/12/2003182512/12/20032000
Subject:
SGP/EBBR/E19 - Wind Speed Sensor Frozen Stopped by Ice Storm
DataStreams:sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1
Description:
An ice storm froze the wind speed sensor to a stopped condition.
It is not known how long this condition existed, as data is missing
for more than 10 days after the end of the stated period.

The wind speed and residual wind speed measurements are incorrect.

This sensor failure does not affect the sensible and latent heat fluxes.
Measurements:sgp5ebbrE19.b1:
  • scalar wind speed(wind_s)
  • vector wind speed(res_ws)

sgp15ebbrE19.b1:
  • scalar wind speed(wind_s)

sgp30ebbrE19.b1:
  • vector wind speed(res_ws)
  • scalar wind speed(wind_s)


Back To Table of Contents

DQRID : D040206.4
Start DateStart TimeEnd DateEnd Time
01/08/2004160001/22/20041530
Subject:
SGP/EBBR/E19 - Low Battery Voltage Caused Incorrect Data
DataStreams:sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1
Description:
Battery voltage less than 10.5 V caused much data to be questionable or incorrect.
Measurements:sgp5ebbrE19.b1:
  • Retrieved pressure profile(pres)
  • scalar wind speed(wind_s)
  • Reference Thermistor Temperature(tref)
  • net radiation(q)
  • time(time)
  • top air temperature(tair_top)
  • Temperature of the top humidity chamber(thum_top)
  • bottom vapor pressure(vp_bot)
  • wind direction (relative to true north)(wind_d)
  • top vapor pressure(vp_top)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • Bottom humidity(hum_bot)
  • vector wind speed(res_ws)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • bottom air temperature(tair_bot)
  • Top humidity(hum_top)
  • Home signal(home)

sgp15ebbrE19.b1:
  • Right relative humidity(mv_hum_r)
  • Soil heat flow 1(mv_hft1)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Soil temperature 1(rr_ts1)
  • Soil heat flow 3(mv_hft3)
  • Atmospheric pressure(mv_pres)
  • Soil temperature 2(rr_ts2)
  • Soil heat flow 2(mv_hft2)
  • Left air temperature(tair_l)
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Soil moisture 5(r_sm5)
  • Soil temperature 4(rr_ts4)
  • Net radiation(mv_q)
  • scalar wind speed(wind_s)
  • Reference temperature(rr_tref)
  • Soil moisture 2(r_sm2)
  • Soil temperature 5(rr_ts5)
  • Left relative humidity(mv_hum_l)
  • Soil temperature 3(rr_ts3)
  • Home signal(mv_home)
  • Wind direction (relative to true north)(mv_wind_d)
  • Soil moisture 4(r_sm4)
  • Right air temperature(tair_r)
  • Soil moisture 3(r_sm3)
  • Soil heat flow 4(mv_hft4)
  • Soil heat flow 5(mv_hft5)
  • Soil moisture 1(r_sm1)

sgp30ebbrE19.b1:
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • Reference Thermistor Temperature(tref)
  • 5 cm soil heat flow, site 2(shf2)
  • volumetric soil moisture, site 1(sm1)
  • volumetric soil moisture, site 2(sm2)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • Exchange mechanism position indicator (15 to 30 min)(home_30)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)
  • Soil heat capacity 5(cs5)
  • Exchange mechanism position indicator (0 to 15 min)(home_15)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • Soil heat capacity 3(cs3)
  • volumetric soil moisture, site 4(sm4)
  • h(h)
  • scalar wind speed(wind_s)
  • Temperature of the top humidity chamber(thum_top)
  • top air temperature(tair_top)
  • Soil heat capacity 4(cs4)
  • Bottom humidity(hum_bot)
  • soil heat flow, site 4(g4)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • top vapor pressure(vp_top)
  • vector wind speed(res_ws)
  • average surface soil heat flow(ave_shf)
  • volumetric soil moisture, site 5(sm5)
  • latent heat flux(e)
  • wind direction (relative to true north)(wind_d)
  • soil heat flow, site 2(g2)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Soil heat capacity 1(cs1)
  • volumetric soil moisture, site 3(sm3)
  • soil heat flow, site 5(g5)
  • soil heat flow, site 1(g1)
  • soil heat flow, site 3(g3)
  • 5 cm soil heat flow, site 1(shf1)
  • bottom vapor pressure(vp_bot)
  • Soil heat capacity 2(cs2)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 5 cm soil heat flow, site 5(shf5)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • Top humidity(hum_top)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • bottom air temperature(tair_bot)
  • net radiation(q)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • 5 cm soil heat flow, site 3(shf3)
  • 5 cm soil heat flow, site 4(shf4)
  • Retrieved pressure profile(pres)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)


Back To Table of Contents

DQRID : D040304.26
Start DateStart TimeEnd DateEnd Time
02/01/2004212002/02/20041910
Subject:
SGP/EBBR/E19  - Wind Speed Frozen Stopped By Ice
DataStreams:sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1
Description:
An ice storm froze the wind speed sensor to a stopped condition.

The wind speed and residual wind speed measurements are incorrect.

This sensor failure does not affect the sensible and latent heat fluxes.
Measurements:sgp5ebbrE19.b1:
  • scalar wind speed(wind_s)
  • vector wind speed(res_ws)

sgp15ebbrE19.b1:
  • scalar wind speed(wind_s)

sgp30ebbrE19.b1:
  • vector wind speed(res_ws)
  • scalar wind speed(wind_s)


Back To Table of Contents

DQRID : D040525.4
Start DateStart TimeEnd DateEnd Time
04/01/2001000005/06/20032359
Subject:
SGP/EBBR/E19 - Metadata errors
DataStreams:sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.a1
Description:
The latitude, longitude and altitude of the El Reno (E19) site were
incorrectly entered into the ARM database.  The correct location
of the EBBR.E19 instrument is:
    Lat: 35.557N
    Lon: 98.017W
    Alt:    421m
Measurements:sgp15ebbrE19.b1:
  • lat(lat)
  • Dummy altitude for Zeb(alt)
  • lon(lon)

sgp5ebbrE19.a1:
  • Dummy altitude for Zeb(alt)
  • lat(lat)
  • lon(lon)

sgp30ebbrE19.b1:
  • Dummy altitude for Zeb(alt)
  • lon(lon)
  • lat(lat)


Back To Table of Contents

DQRID : D040608.7
Start DateStart TimeEnd DateEnd Time
05/29/2004221506/10/20041510
Subject:
SGP/EBBR/E19 - AEM Hung Inbetween Home Positions
DataStreams:sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1
Description:
The AEM hung inbetween home positions, producing a home signal of zero.

Sensible and latent heat fluxes during this period are incorrect.
Measurements:sgp5ebbrE19.b1:
  • bottom vapor pressure(vp_bot)
  • Bottom humidity(hum_bot)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • top air temperature(tair_top)
  • top vapor pressure(vp_top)
  • bottom air temperature(tair_bot)
  • Top humidity(hum_top)
  • Temperature of the top humidity chamber(thum_top)

sgp15ebbrE19.b1:
  • Right air temperature(tair_r)
  • Left air temperature(tair_l)

sgp30ebbrE19.b1:
  • top vapor pressure(vp_top)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • h(h)
  • Temperature of the top humidity chamber(thum_top)
  • top air temperature(tair_top)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • bottom vapor pressure(vp_bot)
  • Top humidity(hum_top)
  • latent heat flux(e)
  • Bottom humidity(hum_bot)
  • bottom air temperature(tair_bot)


Back To Table of Contents

DQRID : D041222.15
Start DateStart TimeEnd DateEnd Time
10/28/2004143010/28/20041500
Subject:
SGP/EBBR/E19 - 6 Month Checks
DataStreams:sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1
Description:
Many measurements were incorrect while the AEM was positioned with both aspirators at the 
same height for comparison checks.
Measurements:sgp5ebbrE19.b1:
  • bottom vapor pressure(vp_bot)
  • Bottom humidity(hum_bot)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • top vapor pressure(vp_top)
  • top air temperature(tair_top)
  • bottom air temperature(tair_bot)
  • Top humidity(hum_top)
  • Temperature of the top humidity chamber(thum_top)

sgp15ebbrE19.b1:
  • Right air temperature(tair_r)
  • Right relative humidity(mv_hum_r)
  • Left relative humidity(mv_hum_l)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Left air temperature(tair_l)
  • Temperature of left humidity sensor chamber(rr_thum_l)

sgp30ebbrE19.b1:
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • h(h)
  • Temperature of the top humidity chamber(thum_top)
  • top air temperature(tair_top)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Top humidity(hum_top)
  • Bottom humidity(hum_bot)
  • bottom air temperature(tair_bot)
  • top vapor pressure(vp_top)
  • bottom vapor pressure(vp_bot)
  • latent heat flux(e)


Back To Table of Contents

DQRID : D050206.13
Start DateStart TimeEnd DateEnd Time
01/05/2005065501/07/20051545
Subject:
SGP/EBBR/E19  - Wind Speed Frozen Stopped By Ice Storm
DataStreams:sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1
Description:
An ice storm froze the wind speed sensor to a stopped condition.

The wind speed and residual wind speed measurements are incorrect.

This sensor failure does not affect the sensible and latent heat fluxes.
Measurements:sgp5ebbrE19.b1:
  • scalar wind speed(wind_s)
  • vector wind speed(res_ws)

sgp15ebbrE19.b1:
  • scalar wind speed(wind_s)

sgp30ebbrE19.b1:
  • vector wind speed(res_ws)
  • scalar wind speed(wind_s)


Back To Table of Contents

DQRID : D050207.9
Start DateStart TimeEnd DateEnd Time
01/05/2005111501/06/20052055
Subject:
SGP/EBBR/E19  - Wind Direction Frozen Stopped By Ice Storm
DataStreams:sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1
Description:
The wind direction sensor was frozen stopped by ice from a storm.  
The wind direction, sigma_wd, and residual wind speed measurements 
are incorrect.
Measurements:sgp5ebbrE19.b1:
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • wind direction (relative to true north)(wind_d)
  • vector wind speed(res_ws)

sgp15ebbrE19.b1:
  • Wind direction (relative to true north)(mv_wind_d)

sgp30ebbrE19.b1:
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • vector wind speed(res_ws)
  • wind direction (relative to true north)(wind_d)


Back To Table of Contents

DQRID : D050215.20
Start DateStart TimeEnd DateEnd Time
01/29/2005093001/29/20051300
Subject:
SGP/EBBR/E19  - Wind Speed Constant
DataStreams:sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1
Description:
Wind Speed was constant at 5 m/s during this period; the reason is unknown.

This sensor failure does not affect the sensible and latent heat fluxes.
Measurements:sgp5ebbrE19.b1:
  • scalar wind speed(wind_s)
  • vector wind speed(res_ws)

sgp15ebbrE19.b1:
  • scalar wind speed(wind_s)

sgp30ebbrE19.b1:
  • vector wind speed(res_ws)
  • scalar wind speed(wind_s)


Back To Table of Contents

DQRID : D050624.3
Start DateStart TimeEnd DateEnd Time
05/07/2005230005/12/20051445
Subject:
SGP/EBBR/E19 - Net Radiation Incorrect - Cable Chewed by Rodents
DataStreams:sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1
Description:
The net radiation measurement was incorrect after the net radiometer cable was chewed by 
rodents.  Sensible and latent heat fluxes were therefore incorrect.
Measurements:sgp5ebbrE19.b1:
  • net radiation(q)

sgp15ebbrE19.b1:
  • Net radiation(mv_q)

sgp30ebbrE19.b1:
  • h(h)
  • latent heat flux(e)
  • net radiation(q)


Back To Table of Contents

DQRID : D051112.1
Start DateStart TimeEnd DateEnd Time
10/13/2003000007/22/20052359
Subject:
SGP/EBBR/E19 - metadata corrections
DataStreams:sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1
Description:
On 20050723, a series of metadata corrections and additions were completed.  These 
metadata changes apply to all EBBR.E19 data collected by ARM back to the beginning of the data 
set in May 1997. Please see the current metadata for correct information.  These changes 
do not affect data values or quality.	

The changes are summarized below:
1) Update of "sensor location" information
2) Addition of installation dates for the systems
3) Correction of soil type
4) Correction of soil moisture units (from "by volume" to "gravimetric")
Measurements:sgp5ebbrE19.b1:
  • base time(base_time)

sgp15ebbrE19.b1:
  • base time(base_time)

sgp30ebbrE19.b1:
  • base time(base_time)


Back To Table of Contents

DQRID : D051214.2
Start DateStart TimeEnd DateEnd Time
10/24/2005060010/25/20050100
10/31/2005103012/20/20051940
Subject:
SGP/EBBR/E19 - Net Radiation, Sensible and Latent Heat Fluxes Incorrect
DataStreams:sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1
Description:
The net radiometer cable was chewed by an animal.  Net Radiation, Sensible Heat Flux, and 
Latent Heat Flux were incorrect during the stated periods.
Measurements:sgp5ebbrE19.b1:
  • net radiation(q)

sgp15ebbrE19.b1:
  • Net radiation(mv_q)

sgp30ebbrE19.b1:
  • h(h)
  • latent heat flux(e)
  • net radiation(q)


Back To Table of Contents

DQRID : D051214.3
Start DateStart TimeEnd DateEnd Time
11/13/2005000001/19/20061630
Subject:
SGP/EBBR/E19 - Incorrect Sensible and Latent Heat Fluxes
DataStreams:sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1
Description:
The gradient measurements and latent and sensible heat fluxes were incorrect because the 
AEM was not exchanging. 

The problem was fixed by replacing the AEM.
Measurements:sgp5ebbrE19.b1:
  • bottom vapor pressure(vp_bot)
  • Bottom humidity(hum_bot)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • top air temperature(tair_top)
  • top vapor pressure(vp_top)
  • bottom air temperature(tair_bot)
  • Top humidity(hum_top)
  • Temperature of the top humidity chamber(thum_top)

sgp15ebbrE19.b1:
  • Right air temperature(tair_r)
  • Left air temperature(tair_l)

sgp30ebbrE19.b1:
  • top vapor pressure(vp_top)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • h(h)
  • Temperature of the top humidity chamber(thum_top)
  • top air temperature(tair_top)
  • soil heat flow, site 1(g1)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Top humidity(hum_top)
  • bottom vapor pressure(vp_bot)
  • latent heat flux(e)
  • Bottom humidity(hum_bot)
  • bottom air temperature(tair_bot)


Back To Table of Contents

DQRID : D080527.2
Start DateStart TimeEnd DateEnd Time
05/07/2008060005/07/20080900
05/07/2008153005/07/20082130
05/09/2008133005/09/20081330
Subject:
SGP/EBBR/E19 - Sensible and latent heat flux and net radiation incorrect
DataStreams:sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1
Description:
Net radiation was too large, resulting in sensible and latent heat fluxes being too large. 
 The cause of the problem is unknown.
Measurements:sgp5ebbrE19.b1:
  • net radiation(q)

sgp15ebbrE19.b1:
  • Net radiation(mv_q)

sgp30ebbrE19.b1:
  • h(h)
  • latent heat flux(e)
  • net radiation(q)


Back To Table of Contents



END OF DATA