Start DateStart TimeEnd DateEnd Time
07/05/2006180007/18/20061647
Subject:
SGP/THWAPS/B4 - Temp and RH data incorrect
DataStreams:sgpthwapsB4.b1
Description:
After sensor replacement the temperature values are too low.  The range of the sensor was 
incorrectly set at -40 to 80 instead of -30 to 70.  RH values are also suspect.
Measurements:sgpthwapsB4.b1:
  • Standard Deviation of Air Temperature(sd_temp)
  • Beam 0 Temperature(temp)
  • Relative humidity inside the instrument enclosure(rh)
  • Standard Deviation of Vapor Pressure(sd_vap_pres)
  • Vapor Pressure (kiloPascals)(vap_pres)
  • Standard Deviation of Relative Humidity(sd_rh)


Back To Table of Contents

DQRID : D060830.1
Start DateStart TimeEnd DateEnd Time
07/05/2006182509/26/20060000
Subject:
SGP/THWAPS/B4 - Reprocess: RH% values 9% High
DataStreams:sgpthwapsB4.b1
Description:
After replacement of the THWAPS T/RH sensor the span was incorrectly set to 0-100 instead 
of 0-110.  This caused the RH values to be 9% higher than actual values.
Measurements:sgpthwapsB4.b1:
  • Relative humidity inside the instrument enclosure(rh)
  • Vapor Pressure (kiloPascals)(vap_pres)


Back To Table of Contents