Data Quality Reports for Session: 115560 User: sherman Completed: 11/07/2008


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D040831.1SGP/SIRS/E1 - Reprocess: Incorrect shortwave radiometer calibration coefficientssgpsirsE1.00, sgpsirsE1.b1
D050205.1SGP/SIRS/E1 - broken cable on DSDH (aka DD) measurment instrument - Eppley 8-48.sgpsirs20sE1.a0, sgpsirsE1.00, sgpsirsE1.b1
D060331.2SGP/SIRS/E1 - Intermittantly missing downwelling diffusesgpsirs20sE1.a0, sgpsirsE1.00, sgpsirsE1.b1
D060630.9SGP/SIRS/E1 - Reprocessed: Longwave Calibration errorsgpsirsE1.b1
D061121.2SGP/SIRS/C1 - Upwelling SW cable problemsgpsirs20sC1.a0, sgpsirsC1.00, sgpsirsC1.b1
D070306.1SGP/SIRS/C1 - Reprocess: Incorrect calibration coefficientssgpsirs20sC1.a0, sgpsirsC1.00, sgpsirsC1.b1
D070424.1SGP/SIRS/C1 - Upwelling LW cable problemssgpsirs20sC1.a0, sgpsirsC1.00, sgpsirsC1.b1
D071022.4SGP/SIRS/C1 - Sensor Cable Problemsgpsirs20sC1.a0, sgpsirsC1.b1


DQRID : D040831.1
Start DateStart TimeEnd DateEnd Time
02/10/2004205004/20/20042015
Subject:
SGP/SIRS/E1 - Reprocess: Incorrect shortwave radiometer calibration coefficients
DataStreams:sgpsirsE1.00, sgpsirsE1.b1
Description:
During the pyrgeometer instrument (PIR) changout a datalogger program was loaded with 
incorrect calibration coefficients for the shortwave instruments.  The affected measurements 
(instruments) were : downwelling diffuse (8-48), upwelling shortwave(PSP), direct normal 
(NIP) and downwelling shortwave (PSP)

Recalculate irradiances by determining original mV signal and reapplying the correct 
calibration coefficients.  For each one minute irradiance value perform the following 
calculations by instrument:
DD(corrected) = (DD/107.24)*109.01
US(corrected) = (US/113.19)*111.47
NIP(corrected)= (NIP/120.17)*114.82
DS(corrected) = (DS/118.46)*116.51
Measurements:sgpsirsE1.b1:
  • Shortwave Direct Normal Irradiance, Pyrgeometer(short_direct_normal)
  • Downwelling Shortwave Diffuse Hemispheric Irradiance, Ventilated Pyranometer,
    Maxima(down_short_diffuse_hemisp_max)
  • Downwelling Shortwave Hemispheric Irradiance, Ventilated Pyranometer, Maxima(down_short_hemisp_max)
  • Downwelling Shortwave Hemispheric Irradiance, Ventilated Pyranometer, Standard
    Deviation(down_short_hemisp_std)
  • Shortwave Direct Normal Irradiance, Pyrgeometer, Maxima(short_direct_normal_max)
  • Shortwave Direct Normal Irradiance, Pyrgeometer, Minima(short_direct_normal_min)
  • Downwelling Shortwave Hemispheric Irradiance, Ventilated Pyranometer, Minima(down_short_hemisp_min)
  • Upwelling (10 meter) Shortwave Hemispheric Irradiance, Pyranometer, Minima(up_short_hemisp_min)
  • Shortwave Direct Normal Irradiance, Pyrgeometer, Standard Deviation(short_direct_normal_std)
  • Upwelling (10 meter) Shortwave Hemispheric Irradiance, Pyranometer, Standard
    Deviation(up_short_hemisp_std)
  • Upwelling (10 meter) Shortwave Hemispheric Irradiance, Pyranometer(up_short_hemisp)
  • Downwelling Shortwave Hemispheric Irradiance, Pyranometer, Standard Deviation(down_short_diffuse_hemisp_std)
  • Downwelling Shortwave Diffuse Hemispheric Irradiance, Ventilated Pyranometer,
    Minima(down_short_diffuse_hemisp_min)
  • Upwelling (10 meter) Shortwave Hemispheric Irradiance, Pyranometer, Maxima(up_short_hemisp_max)
  • Downwelling Shortwave Diffuse Hemispheric Irradiance, Ventilated Pyranometer(down_short_diffuse_hemisp)
  • Down-welling unshaded pyranometer voltage(down_short_hemisp)

sgpsirsE1.00:
  • null(Raw data stream - documentation not supported)


Back To Table of Contents

DQRID : D050205.1
Start DateStart TimeEnd DateEnd Time
12/30/2004210002/08/20052055
Subject:
SGP/SIRS/E1 - broken cable on DSDH (aka DD) measurment instrument - Eppley 8-48.
DataStreams:sgpsirs20sE1.a0, sgpsirsE1.00, sgpsirsE1.b1
Description:
Occasionally, the downwelling shortwave diffuse hemispheric (DSDH) readings spike to 
unreasonably low levels, causing derived downwelling shortwave hemispheric (DSH) readings to 
spike well below the actual DSH readings.  Technicians repaired the DSDH instument (an 
Eppley 8-48) cable on 02/08/2005 2055GMT.  There have been no dropouts since that time.
Measurements:sgpsirs20sE1.a0:
  • Instantaneous Uncorrected Downwelling Shortwave Diffuse, Shaded Pyranometer
    Thermopile Voltage(inst_diffuse)

sgpsirsE1.b1:
  • Downwelling Shortwave Hemispheric Irradiance, Pyranometer, Standard Deviation(down_short_diffuse_hemisp_std)
  • Downwelling Shortwave Diffuse Hemispheric Irradiance, Ventilated Pyranometer,
    Minima(down_short_diffuse_hemisp_min)
  • Downwelling Shortwave Diffuse Hemispheric Irradiance, Ventilated Pyranometer,
    Maxima(down_short_diffuse_hemisp_max)
  • Downwelling Shortwave Diffuse Hemispheric Irradiance, Ventilated Pyranometer(down_short_diffuse_hemisp)

sgpsirsE1.00:
  • null(Raw data stream - documentation not supported)


Back To Table of Contents

DQRID : D060331.2
Start DateStart TimeEnd DateEnd Time
10/26/2005163004/04/20062005
Subject:
SGP/SIRS/E1 - Intermittantly missing downwelling diffuse
DataStreams:sgpsirs20sE1.a0, sgpsirsE1.00, sgpsirsE1.b1
Description:
Starting on or about 10/26/2005 the downwelling diffuse hemispheric irradiance (Eppley 
8-48) instrument had intermittant missing values. Technicians found a loose wire on the 
instrument connector and repaired it on 4/4/2006. 

Use global shortwave hemispheric (PSP) component - normal incident component (CosZA*NIP) 
to calulate diffuse component.  Comparison of absolute difference between actual and 
computed diffuse component values and diffuse standard deviations > 1-2 W/m2 can help filter 
the noisy diffuse data.  There is no standard method.
Measurements:sgpsirs20sE1.a0:
  • Instantaneous Uncorrected Downwelling Shortwave Diffuse, Shaded Pyranometer
    Thermopile Voltage(inst_diffuse)

sgpsirsE1.b1:
  • Downwelling Shortwave Hemispheric Irradiance, Pyranometer, Standard Deviation(down_short_diffuse_hemisp_std)
  • Downwelling Shortwave Diffuse Hemispheric Irradiance, Ventilated Pyranometer,
    Minima(down_short_diffuse_hemisp_min)
  • Downwelling Shortwave Diffuse Hemispheric Irradiance, Ventilated Pyranometer,
    Maxima(down_short_diffuse_hemisp_max)
  • Downwelling Shortwave Diffuse Hemispheric Irradiance, Ventilated Pyranometer(down_short_diffuse_hemisp)

sgpsirsE1.00:
  • null(Raw data stream - documentation not supported)


Back To Table of Contents

DQRID : D060630.9
Start DateStart TimeEnd DateEnd Time
02/10/2004205502/21/20062007
Subject:
SGP/SIRS/E1 - Reprocessed: Longwave Calibration error
DataStreams:sgpsirsE1.b1
Description:
Modified pyrgeometer calibration procedures were implemented beginning in February 2004. 
These modified procedures introduced a calibration bias in the longwave data. The previous 
procedures were re-implemented at all sites between December 2005 and February 2006 to 
restore proper calibrations.	

The data collected while the incorrect procedures were in place have been reprocessed to 
remove the calibration bias.  The reprocessed 60 second averaged data are based on 3 
instantaneous 20 second data records rather than on 60 1 second instantaneous data records.  
Still, these data are considered far superior to the originally processed data.  The 
reprocessed data were archived in October 2006.
Measurements:sgpsirsE1.b1:
  • Upwelling (10 meter) Longwave Hemispheric Irradiance, Pyrgeometer(up_long_hemisp)
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer(down_long_hemisp_shaded)
  • Upwelling Longwave Hemispheric Net Infrared(up_long_netir)
  • Downwelling Longwave Hemispheric Net Infrared(down_long_netir)


Back To Table of Contents

DQRID : D061121.2
Start DateStart TimeEnd DateEnd Time
11/13/2006090001/23/20071930
Subject:
SGP/SIRS/C1 - Upwelling SW cable problem
DataStreams:sgpsirs20sC1.a0, sgpsirsC1.00, sgpsirsC1.b1
Description:
Starting 11/13/2006, the C1 SIRS Upwelling Shortwave (up_short_hemisp) started to show 
large differences in comparison to co-located E13 SIRS and C1 IRT25m.  After much 
troubleshooting and gnashing of teeth it was discovered that rodents had chewed through the 
instrument cables at the bottom of the tilt tower. Cables were replaced 01/23/2007 1930GMT.
Measurements:sgpsirsC1.b1:
  • Upwelling (10 meter) Shortwave Hemispheric Irradiance, Pyranometer, Standard
    Deviation(up_short_hemisp_std)
  • Upwelling (10 meter) Shortwave Hemispheric Irradiance, Pyranometer, Minima(up_short_hemisp_min)
  • Upwelling (10 meter) Shortwave Hemispheric Irradiance, Pyranometer, Maxima(up_short_hemisp_max)
  • Upwelling (10 meter) Shortwave Hemispheric Irradiance, Pyranometer(up_short_hemisp)

sgpsirsC1.00:
  • null(Raw data stream - documentation not supported)

sgpsirs20sC1.a0:
  • Instantaneous Upwelling Shortwave Hemispheric Irradiance, Pyranometer(inst_up_short_hemisp)


Back To Table of Contents

DQRID : D070306.1
Start DateStart TimeEnd DateEnd Time
07/03/2006150903/06/20071610
Subject:
SGP/SIRS/C1 - Reprocess: Incorrect calibration coefficients
DataStreams:sgpsirs20sC1.a0, sgpsirsC1.00, sgpsirsC1.b1
Description:
Incorrection calibration coefficients were used in the C1/SIRS datalogger program. The 
PSP-US (upwelling shortwave global hemispheric), 30943F3 should be 126.93 not 129.93 
[W/m2/mV].
Measurements:sgpsirsC1.b1:
  • Upwelling (10 meter) Shortwave Hemispheric Irradiance, Pyranometer, Standard
    Deviation(up_short_hemisp_std)
  • Upwelling (10 meter) Shortwave Hemispheric Irradiance, Pyranometer, Minima(up_short_hemisp_min)
  • Upwelling (10 meter) Shortwave Hemispheric Irradiance, Pyranometer, Maxima(up_short_hemisp_max)
  • Upwelling (10 meter) Shortwave Hemispheric Irradiance, Pyranometer(up_short_hemisp)

sgpsirsC1.00:
  • null(Raw data stream - documentation not supported)

sgpsirs20sC1.a0:
  • Instantaneous Upwelling Shortwave Hemispheric Irradiance, Pyranometer(inst_up_short_hemisp)


Back To Table of Contents

DQRID : D070424.1
Start DateStart TimeEnd DateEnd Time
01/03/2007200501/23/20071930
Subject:
SGP/SIRS/C1 - Upwelling LW cable problems
DataStreams:sgpsirs20sC1.a0, sgpsirsC1.00, sgpsirsC1.b1
Description:
Troubleshooting of the upwelling shortwave instrument on Jan 3rd 2007 caused problems with 
the co-located upwelling longwave hemispheric irradiance Eppley PIR instrument (aka 
UIR).  Data dropouts persisted from this date to the final resolution of both shortwave and 
longwave instrument problems.  Rodents had chewed through the instrument cables.  The 
cables were replaced 01/23/2007 1930GMT.
Measurements:sgpsirsC1.b1:
  • Instantaneous Case Temperature(inst_up_long_case_temp)
  • Upwelling (10 meter) Longwave Hemispheric Irradiance, Pyrgeometer, Minima(up_long_hemisp_min)
  • Upwelling (10 meter) Longwave Hemispheric Irradiance, Pyrgeometer, Maxima(up_long_hemisp_max)
  • Instantaneous Dome Temperature(inst_up_long_dome_temp)
  • Upwelling (10 meter) Longwave Hemispheric Irradiance, Pyrgeometer(up_long_hemisp)
  • Upwelling (10 meter) Longwave Hemispheric Irradiance, Pyrgeometer, Standard
    Deviation(up_long_hemisp_std)
  • Upwelling Longwave Hemispheric Net Infrared(up_long_netir)

sgpsirsC1.00:
  • null(Raw data stream - documentation not supported)

sgpsirs20sC1.a0:
  • Instantaneous Upwelling Pyrgeometer Dome Thermistor Resistance, Pyrgeometer(inst_up_long_dome_resist)
  • Instantaneous Upwelling Pyrgeometer Thermopile(inst_up_long_hemisp_tp)
  • Instantaneous Upwelling Pyrgeometer Case Thermistor Resistance, Pyrgeometer(inst_up_long_case_resist)


Back To Table of Contents

DQRID : D071022.4
Start DateStart TimeEnd DateEnd Time
10/03/2007140004/22/20082052
Subject:
SGP/SIRS/C1 - Sensor Cable Problem
DataStreams:sgpsirs20sC1.a0, sgpsirsC1.b1
Description:
Beginning October 3, 2007, the upwelling longwave irradiance (up_long_hemisp, aka "UIR") 
started to spike to unreasonable values positively offset from the co-located E13 by over 
100W/m2.  On January 17 2008 the UIR values become consistently bad due to bad UIR dome 
thermistor wiring.  Problem appears as a resistive offset in the dome thermistor but due 
to nature of failure (unknown cable problem) data correction would not yield any better 
data than the co-located E13 UIR instrument that is known to be operating correctly.  
Instrument cable replacement on 4/22 fixed the problem.

Thermopile data (netIR) from this instrument is good.  However, Incoming irradiance data 
is not correctable due to bad UIR dome thermistor wiring within the instrument cable.
Measurements:sgpsirsC1.b1:
  • Upwelling (10 meter) Longwave Hemispheric Irradiance, Pyrgeometer, Minima(up_long_hemisp_min)
  • Upwelling (10 meter) Longwave Hemispheric Irradiance, Pyrgeometer, Maxima(up_long_hemisp_max)
  • Upwelling (10 meter) Longwave Hemispheric Irradiance, Pyrgeometer(up_long_hemisp)
  • Upwelling (10 meter) Longwave Hemispheric Irradiance, Pyrgeometer, Standard
    Deviation(up_long_hemisp_std)
  • Upwelling Longwave Hemispheric Net Infrared(up_long_netir)

sgpsirs20sC1.a0:
  • Instantaneous Upwelling Pyrgeometer Dome Thermistor Resistance, Pyrgeometer(inst_up_long_dome_resist)
  • Instantaneous Upwelling Pyrgeometer Thermopile(inst_up_long_hemisp_tp)
  • Instantaneous Upwelling Pyrgeometer Case Thermistor Resistance, Pyrgeometer(inst_up_long_case_resist)


Back To Table of Contents



END OF DATA