Data Quality Reports for Session: 114921 User: jimchet197 Completed: 10/16/2008


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D050421.5SGP/SONDE/C1 - Surface pressure incorrectsgpsondewnpnC1.b1
D051121.1SGP/SONDE/C1 - Sonde RH values incorrectsgpsondewnpnC1.b1
D060928.3SGP/SONDE/C1 - Increased precision for alt, dp, pres, rh, tdrysgpsondewnpnC1.b1


DQRID : D050421.5
Start DateStart TimeEnd DateEnd Time
04/11/2005174904/11/20051920
Subject:
SGP/SONDE/C1 - Surface pressure incorrect
DataStreams:sgpsondewnpnC1.b1
Description:
I think the operator inadvdertently entered an incorrect surface pressure value.  The 
value entered is 988.8 hPa but the value shown in the raw data file at launch is 968.8.  As a 
result, the absolute altitudes in this sounding will be slightly incorrect as will be 
the ascent rate during the first 30 seconds of flight.
Measurements:sgpsondewnpnC1.b1:
  • Ascent Rate(asc)
  • Dummy altitude for Zeb(alt)


Back To Table of Contents

DQRID : D051121.1
Start DateStart TimeEnd DateEnd Time
04/13/2005234604/15/20051732
Subject:
SGP/SONDE/C1 - Sonde RH values incorrect
DataStreams:sgpsondewnpnC1.b1
Description:
RH values for all sondes launched during this time were incorrect due to a faulty Ground 
Check set which damaged or conditioned the RH sensor improperly.
Measurements:sgpsondewnpnC1.b1:
  • Surface dew point temperature(dp)
  • Relative humidity inside the instrument enclosure(rh)


Back To Table of Contents

DQRID : D060928.3
Start DateStart TimeEnd DateEnd Time
09/28/2006172812/31/99992359
Subject:
SGP/SONDE/C1 - Increased precision for alt, dp, pres, rh, tdry
DataStreams:sgpsondewnpnC1.b1
Description:
The output precision of the following variables was increased 20060928

alt  changed from 1m to 0.1m
dp   changed from 0.1degC to 0.01degC
pres changed from 0.1 hPa to 0.01 hPa
RH   changed from 1%RH to 0.01%RH
tdry changed from .1degC to 0.01degC

Note that the increase in precision does not necessarily imply an increase in accuracy.  
Users are cautioned to continue to use their best scientific judgement when using these or 
any other data.
Measurements:sgpsondewnpnC1.b1:
  • Surface dew point temperature(dp)
  • Dry bulb temperature(tdry)
  • Relative humidity inside the instrument enclosure(rh)
  • Retrieved pressure profile(pres)
  • Dummy altitude for Zeb(alt)


Back To Table of Contents