Data Quality Reports for Session: 114843 User: mcfarlan Completed: 10/13/2008


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D030317.1TWP/SMET/C1 - Reprocess: Lower Wind Speed incorrecttwpsmet60sC1.a1, twpsmet60sC1.b1
D040220.1TWP/MWR/C1 - wrong azimuthtwpmwrlosC1.b1
D040428.1TWP/SMET/C1 - Poor Performing Sensortwpsmet60sC1.b1
D040824.3TWP/SMET/C1 - Incorrect rainrate equation usedtwpsmet60sC1.b1
D050421.6TWP/SMET/C1 - Negative rainfall ratestwpsmet60sC1.b1
D050725.9TWP/MWR/C1 - Reprocessed: Revised Retrieval Coefficientstwp5mwravgC1.c1, twpmwrlosC1.b1, twpmwrtipC1.a1
D050928.1TWP/MWR/C1 - New software version (4.15) installedtwpmwrlosC1.b1, twpmwrtipC1.a1
D051119.1TWP/SKYRAD/C1 - IRT failuretwpskyrad60sC1.b1
D060630.34TWP/GNDRAD/C1 - Reprocess: Longwave Calibration errortwpgndrad60sC1.b1
D060817.3TWP/SMET/C1 - Reprocess: Barometric pressure units changetwpsmet60sC1.b1
D061011.2TWP/SMET/C1 - Reprocess: Tipping bucket rain gauge addedtwpsmet60sC1.b1


DQRID : D030317.1
Start DateStart TimeEnd DateEnd Time
02/27/2003175003/03/20031900
Subject:
TWP/SMET/C1 - Reprocess: Lower Wind Speed incorrect
DataStreams:twpsmet60sC1.a1, twpsmet60sC1.b1
Description:
The lower wind speed sensor (sensor #2) was reporting values 10 times above the actual 
value.  During an update to the logger configuration file the calibration multiplier was 
incorrectly entered.  The multiplier inserted was .97630 when it should have been .09763.  
This caused the readings from this sensor to be incorrect.
Measurements:twpsmet60sC1.b1:
  • Lower sensor, wind speed vector average(lo_spd_vec_avg)
  • Lower sensor, wind speed arithmetic average(lo_spd_arith_avg)
  • Lower sensor, wind speed minimum(lo_spd_min)
  • Lower sensor, wind speed standard deviation(lo_spd_sd)
  • Lower sensor, wind speed maximum(lo_spd_max)

twpsmet60sC1.a1:
  • Wind #2 speed minimum(wind2_spd_min)
  • Wind #2 speed arithmetic average(wind2_spd_arith_avg)
  • Wind #2 speed vector average(wind2_spd_vec_avg)
  • Wind #2 speed standard deviation(wind2_spd_sd)
  • Wind #2 speed maximum(wind2_spd_max)
  • Wind #2 speed maximum gust(wind2_spd_max_gust)


Back To Table of Contents

DQRID : D040220.1
Start DateStart TimeEnd DateEnd Time
10/11/1996000002/18/20042250
Subject:
TWP/MWR/C1 - wrong azimuth
DataStreams:twpmwrlosC1.b1
Description:
The MWR was initially installed at an azimuth angle defined as 180 degrees but the value 
in the configuration file was not changed from the default of 0 degrees. In examining 
photos taken during the installation of the AWS tower, I noticed that the MWR was rotated 
opposite the normal orientation. The value in the configuration file was changed to reflect 
the actual azimuth of the instrument.
Measurements:twpmwrlosC1.b1:
  • Actual Azimuth(actaz)


Back To Table of Contents

DQRID : D040428.1
Start DateStart TimeEnd DateEnd Time
04/23/2004000001/19/20051530
Subject:
TWP/SMET/C1 - Poor Performing Sensor
DataStreams:twpsmet60sC1.b1
Description:
On 4/23/2004 condensation was found inside the target area of the ORG lens. A spare was 
sent and installed but did not solve the problem of high background voltage values causing 
rainrate values to be recorded during no rainfall.  Values range anywhere from 0.06 to 
0.12 mm/hr.  

A new Optical Rain Gage was installed on 01/19/2005 @ 13:30 GMT alleviating the high 
background voltage values that lead to consistent rainfall rates of around 0.11 to 0.15 mm/hr 
on a consistent basis.  

The formula used to calculate rainfall rates from the new Optical Rain Gage is:  RR 
(mm/hr) = 25(V^1.87) - 0.15.

Older model formula = 20(V^2) - 0.05

The two formulas result in comparable data.
Measurements:twpsmet60sC1.b1:
  • Precipitation mean(precip_mean)
  • Precipitation maximum(precip_max)
  • Precipitation standard deviation(precip_sd)
  • precipitation minimum(precip_min)


Back To Table of Contents

DQRID : D040824.3
Start DateStart TimeEnd DateEnd Time
08/03/2004030708/17/20042138
Subject:
TWP/SMET/C1 - Incorrect rainrate equation used
DataStreams:twpsmet60sC1.b1
Description:
There are two different Optical Raingages in use at the TWP sites.  The incorrect equation 
was used at this site.  The incorrect equation is (25*(V^1.87)) - 0.15.
The correct equation is (20*(V^2)) - 0.05.  The correct equation was installed.
Measurements:twpsmet60sC1.b1:
  • Precipitation mean(precip_mean)
  • Precipitation maximum(precip_max)
  • Precipitation standard deviation(precip_sd)
  • precipitation minimum(precip_min)


Back To Table of Contents

DQRID : D050421.6
Start DateStart TimeEnd DateEnd Time
10/09/1996000003/09/20042243
Subject:
TWP/SMET/C1 - Negative rainfall rates
DataStreams:twpsmet60sC1.b1
Description:
Due to limitations of the Zeno dataloggers all voltage measurements
from the optical rain gauge (ORG) were converted to rainfall rates. 
This resulted in negative values for rain rates.  Any value below 0.10 mm/hr should be 
considered to be 0 mm/hr.  New dataloggers 
alleviated this problem.  New loggers were installed 03/09/2004 @ 2243 GMT.
Measurements:twpsmet60sC1.b1:
  • Precipitation mean(precip_mean)
  • Precipitation maximum(precip_max)
  • Precipitation standard deviation(precip_sd)
  • precipitation minimum(precip_min)


Back To Table of Contents

DQRID : D050725.9
Start DateStart TimeEnd DateEnd Time
05/04/2002020006/30/20052100
Subject:
TWP/MWR/C1 - Reprocessed: Revised Retrieval Coefficients
DataStreams:twp5mwravgC1.c1, twpmwrlosC1.b1, twpmwrtipC1.a1
Description:
IN THE BEGINNING (June 1992), the retrieval coefficients used to derive 
the precipitable water vapor (PWV) and liquid water path (LWP) from the 
MWR brightness temperatures were based on the Liebe and Layton (1987) 
water vapor and oxygen absorption model and the Grant (1957) liquid 
water absorption model.

Following the SHEBA experience, revised retrievals based on the more 
recent Rosenkranz (1998) water vapor and oxygen absorption models and 
the Liebe (1991) liquid waer absorption model were developed.  The 
Rosenkranz water vapor absorption model resulted a 2 percent increase 
in PWV relative to the earlier Liebe and Layton model.  The Liebe 
liquid water absorption model decreased the LWP by 10% relative to the 
Grant model.  However, the increased oxygen absorption caused a 
0.02-0.03 mm (20-30 g/m2) reduction in LWP, which was particularly 
significant for low LWP conditions (i.e. thin clouds encountered at 
SHEBA).

Recently, it has been shown (Liljegren, Boukabara, Cady-Pereira, and 
Clough, TGARS v. 43, pp 1102-1108, 2005) that the half-width of the 
22 GHz water vapor line from the HITRAN compilation, which is 5 percent 
smaller than the Liebe and Dillon (1969) half-width used in Rosenkranz 
(1998), provided a better fit to the microwave brightness temperature 
measurements at 5 frequencies in the range 22-30 GHz, and yielded more 
accurate retrievals. Accordingly, revised MWR retrieval coefficients 
have been developed using MONORTM, which utilizes the HITRAN compilation 
for its spectroscopic parameters.  These new retrievals provide 3 
percent less PWV and 2.6 percent greater LWP than the previous 
retrievals based on Rosenkranz (1998).

The Rosenkranz-based retrieval coefficients became active at TWP.C1 
20020504.0200.  The MONORTM-based retrieval coefficients became active 
at TWP.C1 20050630.2100.

Note: The TWP.C1 data for 19961011-20050630 have been reprocessed to apply the
Measurements:twpmwrlosC1.b1:
  • MWR column precipitable water vapor(vap)
  • Averaged total liquid water along LOS path(liq)

twp5mwravgC1.c1:
  • MWR column precipitable water vapor(vap)
  • Averaged total liquid water along LOS path(liq)

twpmwrtipC1.a1:
  • Total water vapor along zenith path using tip-derived brightness temperatures(vaptip)
  • Total liquid water along zenith path using tip-derived brightness temperatures(liqtip)


Back To Table of Contents

DQRID : D050928.1
Start DateStart TimeEnd DateEnd Time
10/31/2002220009/13/20051854
Subject:
TWP/MWR/C1 - New software version (4.15) installed
DataStreams:twpmwrlosC1.b1, twpmwrtipC1.a1
Description:
A problem began with the installation of MWR.EXE version 4.12 in October 2002. The 
software had been upgraded from a "DOS" to a "Windows"-compiled program to address an earlier 
problem.  The software upgrade corrected the earlier problem but introduced a new one that 
caused line-of-sight observing cycles to be skipped, a 15% reduction in the number of tip 
curves, and saturation of CPU usage.  Software versions 4.13 and 4.14 also produced 
these problems.

The new MWR software version (4.15) was installed on 9/13/2005. As a consequence of this 
upgrade, the tip curve frequency increased. The tip cycle time decreased from ~60s to ~50s.
Measurements:twpmwrlosC1.b1:
  • Mixer kinetic (physical) temperature(tkxc)
  • Temperature correction coefficient at 23.8 GHz(tc23)
  • IR Brightness Temperature(ir_temp)
  • 31.4 GHz blac2body+noise injection signal(bbn31)
  • Sky brightness temperature at 23.8 GHz(tbsky23)
  • 23.8 GHz Blackbody signal(bb23)
  • Blackbody kinetic temperature(tkbb)
  • (tknd)
  • 31.4 GHz blackbody(bb31)
  • 31.4 GHz sky signal(sky31)
  • Sky brightness temperature at 31.4 GHz(tbsky31)
  • Temperature correction coefficient at 31.4 GHz(tc31)
  • MWR column precipitable water vapor(vap)
  • 23.8 GHz blackbody+noise injection signal(bbn23)
  • Ambient temperature(tkair)
  • Noise injection temp at nominal temperature at 31.4 GHz(tnd_nom31)
  • Averaged total liquid water along LOS path(liq)
  • Noise injection temp at nominal temperature at 23.8 GHz(tnd_nom23)
  • Sky Infra-Red Temperature(sky_ir_temp)
  • 23.8 GHz sky signal(sky23)
  • Noise injection temp at 31.4 GHz adjusted to tkbb(tnd31)
  • Noise injection temp at 23.8 GHz adjusted to tkbb(tnd23)

twpmwrtipC1.a1:
  • Noise injection temp at 31.4 GHz adjusted to tkbb(tnd31)
  • Noise injection temp at nominal temperature at 23.8 GHz(tnd_nom23)
  • 23.8 GHz goodness-of-fit coefficient(r23)
  • Noise injection temp at 31.4 GHz derived from this tip(tnd31I)
  • 31.8 GHz sky brightness temperature derived from tip curve(tbskytip31)
  • Total liquid water along zenith path using tip-derived brightness temperatures(liqtip)
  • Tip configuration number(tipn)
  • Mixer kinetic (physical) temperature(tkxc)
  • IR Brightness Temperature(ir_temp)
  • Total water vapor along zenith path using tip-derived brightness temperatures(vaptip)
  • 31.4 GHz blac2body+noise injection signal(bbn31)
  • 31.4 GHz sky signal(tipsky31)
  • 23.8 GHz Blackbody signal(bb23)
  • Ambient temperature(tkair)
  • (tknd)
  • 31.4 GHz goodness-of-fit coefficient(r31)
  • Blackbody kinetic temperature(tkbb)
  • 23.8 GHz blackbody+noise injection signal(bbn23)
  • 23.8 GHz sky brightness temperature derived from tip curve(tbskytip23)
  • Noise injection temp at 23.8 GHz adjusted to tkbb(tnd23)
  • 23.8 GHz sky signal(tipsky23)
  • Noise injection temp at nominal temperature at 31.4 GHz(tnd_nom31)
  • Temperature correction coefficient at 23.8 GHz(tc23)
  • 31.4 GHz blackbody(bb31)
  • Blackbody temperature 1(tkbb1)
  • Noise injection temp at 23.8 GHz derived from this tip(tnd23I)
  • Blackbody temperature 2(tkbb2)
  • Temperature correction coefficient at 31.4 GHz(tc31)


Back To Table of Contents

DQRID : D051119.1
Start DateStart TimeEnd DateEnd Time
11/13/2005010001/10/20060145
Subject:
TWP/SKYRAD/C1  - IRT failure
DataStreams:twpskyrad60sC1.b1
Description:
A moisture related fault was found in the IRT cable after a day of very heavy rain. The 
cable was replaced but the internal power supply was apparently damaged. The instrument was 
replaced with a spare IRT.
Measurements:twpskyrad60sC1.b1:
  • Sky Infra Red Temperature Minima(sky_ir_temp_min)
  • Sky Infra Red Temperature Maxima(sky_ir_temp_max)
  • Sky Infra-Red Temperature(sky_ir_temp)
  • Standard Deviation of Sky Infra Red Temperature(sky_ir_temp_std)


Back To Table of Contents

DQRID : D060630.34
Start DateStart TimeEnd DateEnd Time
08/02/2004070012/21/20052204
Subject:
TWP/GNDRAD/C1 - Reprocess: Longwave Calibration error
DataStreams:twpgndrad60sC1.b1
Description:
Modified pyrgeometer calibration procedures were implemented beginning in December 2003.  
These modified procedures introduced a calibration bias in the longwave data.  The 
previous procedures were re-implemented at all sites between December 2005 and February 2006 to 
restore proper calibrations.                                                             
                The data collected while the incorrect procedures were in place are 
being reprocessing to remove the calibration bias.  Note, this also affected standard 
deviation, maximum and minimum data fields.
Measurements:twpgndrad60sC1.b1:
  • Upwelling (10 meter) Longwave Hemispheric Irradiance, Pyrgeometer(up_long_hemisp)


Back To Table of Contents

DQRID : D060817.3
Start DateStart TimeEnd DateEnd Time
10/09/1996000008/02/20041911
Subject:
TWP/SMET/C1 - Reprocess: Barometric pressure units change
DataStreams:twpsmet60sC1.b1
Description:
Barometric pressure data during this time period is reported in hPa.  After this time, the 
data have been converted to kPa to standardize the measurement units among ARM sites and 
to conform to accepted standard units determined by the scientific community.
Measurements:twpsmet60sC1.b1:
  • Barometric Pressure(atmos_pressure)


Back To Table of Contents

DQRID : D061011.2
Start DateStart TimeEnd DateEnd Time
10/09/1996000010/17/20060000
Subject:
TWP/SMET/C1 - Reprocess: Tipping bucket rain gauge added
DataStreams:twpsmet60sC1.b1
Description:
A tipping bucket rain gauge was added to the TWP.C1 SMET suite of instruments on 20061017. 
 Effective 20060926, two new variables (count and rain amount) were added to the end of 
the twpsmet60sC1.b1 datastream.  Between 9/26 and 10/17, the data for these two variables 
are filled with zeros.  Users should not use the data during this period as the zeros 
are not representative of the rain fall at the site.  When reprocessing of this datastream 
occurs these variables will be added for the 19961009-20060926 and data from 
19961009-20061017 will be filled with -9999 (missing).
Measurements:twpsmet60sC1.b1:
  • base time(base_time)


Back To Table of Contents



END OF DATA