Data Quality Reports for Session: 113102 User: sbenson Completed: 07/11/2008


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D000622.3TWP/C1 - Incorrect elevation reported in Manus DatatwpsondewnpnC1.a1
D020620.66TWP/C1 - Incorrect elevation reported in Manus DatatwpsondewnpnC1.a1
D021229.1TWP/SONDE/C1 - Broken Temperature SensortwpsondewnpnC1.a1, twpsondewnpnC1.b1
D021231.2TWP/Sonde/C1 - Bad RH ValuestwpsondewnpnC1.a1, twpsondewnpnC1.b1
D030427.2TWP/SONDE/C1 - Questionable RHtwpsondewnpnC1.a1, twpsondewnpnC1.b1
D030503.1TWP/SONDE/C1 - Bad RHtwpsondewnpnC1.a1, twpsondewnpnC1.b1
D030603.3TWP/SONDE/C1 - Broken temperature sensortwpsondewnpnC1.a1, twpsondewnpnC1.b1
D040530.1TWP/SONDE/C1 - Bad RH sensortwpsondewnpnC1.a1, twpsondewnpnC1.b1


DQRID : D000622.3
Start DateStart TimeEnd DateEnd Time
08/28/1997045506/22/20002359
Subject:
TWP/C1 - Incorrect elevation reported in Manus Data
DataStreams:twpsondewnpnC1.a1
Description:
(From Bill Clements) We have been using 6m MSL as the station
elevation of Momote (Manus).  This was based on information
posted near the station barometer in the Momote Weather Service
Office.  It turns out that the official WMO station elevation
for Momote is 4m MSL and the upper air elevation is 5m MSL.

All Manus instruments except the sonde station are at 4m MSL.
The sonde launching station is at 5m MSL.
Measurements:twpsondewnpnC1.a1:
  • Dummy altitude for Zeb(alt)


Back To Table of Contents

DQRID : D020620.66
Start DateStart TimeEnd DateEnd Time
11/03/1997000006/22/20002359
Subject:
TWP/C1 - Incorrect elevation reported in Manus Data
DataStreams:twpsondewnpnC1.a1
Description:
(From Bill Clements) We have been using 6m MSL as the station
elevation of Momote (Manus).  This was based on information
posted near the station barometer in the Momote Weather Service
Office.  It turns out that the official WMO station elevation
for Momote is 4m MSL and the upper air elevation is 5m MSL.

All Manus instruments except the sonde station are at 4m MSL.
The sonde launching station is at 5m MSL.
Measurements:twpsondewnpnC1.a1:
  • Dummy altitude for Zeb(alt)


Back To Table of Contents

DQRID : D021229.1
Start DateStart TimeEnd DateEnd Time
09/18/2002233409/19/20020111
09/21/2002233009/22/20020059
10/01/2002234910/02/20020117
10/13/2002113910/13/20021303
10/18/2002235810/19/20020113
05/23/2003233005/24/20030135
06/14/2003233006/15/20030147
06/22/2003115906/22/20031332
06/22/2003234906/23/20030110
06/23/2003114606/23/20031323
06/25/2003234206/26/20030118
06/28/2003233006/29/20030056
06/29/2003113406/29/20031312
Subject:
TWP/SONDE/C1  - Broken Temperature Sensor
DataStreams:twpsondewnpnC1.a1, twpsondewnpnC1.b1
Description:
The temperature sensor on these radiosondes broke after launch.
All the temperature related data are incorrect.
Measurements:twpsondewnpnC1.b1:
  • Surface dew point temperature(dp)
  • Relative humidity inside the instrument enclosure(rh)
  • Dry bulb temperature(tdry)

twpsondewnpnC1.a1:
  • Relative humidity inside the instrument enclosure(rh)
  • Dry bulb temperature(tdry)
  • Surface dew point temperature(dp)


Back To Table of Contents

DQRID : D021231.2
Start DateStart TimeEnd DateEnd Time
12/21/2002233612/22/20020119
Subject:
TWP/Sonde/C1 - Bad RH Values
DataStreams:twpsondewnpnC1.a1, twpsondewnpnC1.b1
Description:
The RH values from the surface to approximately 11.5 km oscillate between reasonable 
values and zero.  This is apparently caused by a failure in the RH sensor heating process.  
Although the oscillation stops when the heating stops (at temperatures below ~ -45degC) I 
think the values must be viewed as suspect.
Measurements:twpsondewnpnC1.b1:
  • Surface dew point temperature(dp)
  • Relative humidity inside the instrument enclosure(rh)

twpsondewnpnC1.a1:
  • Relative humidity inside the instrument enclosure(rh)
  • Surface dew point temperature(dp)


Back To Table of Contents

DQRID : D030427.2
Start DateStart TimeEnd DateEnd Time
04/08/2003234704/09/20030003
Subject:
TWP/SONDE/C1 - Questionable RH
DataStreams:twpsondewnpnC1.a1, twpsondewnpnC1.b1
Description:
It appears that the RH sensor heating circuit failed during this portion of the sounding 
(approximately between 5km and 10km).  The RH oscillates rapidly between low (~5 %RH) and 
not so low (~50 %RH) and the oscillation seems to stop when the temperature is below ~35 
degC.  According to Vaisala these are symptoms of failure in the ASIC (application 
specific integrated circuit) that controls the heating.
Measurements:twpsondewnpnC1.b1:
  • Surface dew point temperature(dp)
  • Relative humidity inside the instrument enclosure(rh)

twpsondewnpnC1.a1:
  • Relative humidity inside the instrument enclosure(rh)
  • Surface dew point temperature(dp)


Back To Table of Contents

DQRID : D030503.1
Start DateStart TimeEnd DateEnd Time
04/29/2003112304/29/20031205
Subject:
TWP/SONDE/C1 - Bad RH
DataStreams:twpsondewnpnC1.a1, twpsondewnpnC1.b1
Description:
The RH sensor heating circuit failed on this radiosonde.  The RH-related values obtained 
between 0 and 10000 meters oscillate between near-zero and reasonable ambient values.  
Above 10000 the RH values are likely correct.
Measurements:twpsondewnpnC1.b1:
  • Surface dew point temperature(dp)
  • Relative humidity inside the instrument enclosure(rh)

twpsondewnpnC1.a1:
  • Relative humidity inside the instrument enclosure(rh)
  • Surface dew point temperature(dp)


Back To Table of Contents

DQRID : D030603.3
Start DateStart TimeEnd DateEnd Time
05/23/2003233005/24/20030135
Subject:
TWP/SONDE/C1 - Broken temperature sensor
DataStreams:twpsondewnpnC1.a1, twpsondewnpnC1.b1
Description:
The temperature sensor on this radiosonde broke after launch.  All the temperature-related 
data are incorrect.
Measurements:twpsondewnpnC1.b1:
  • Surface dew point temperature(dp)
  • Relative humidity inside the instrument enclosure(rh)
  • Dry bulb temperature(tdry)

twpsondewnpnC1.a1:
  • Relative humidity inside the instrument enclosure(rh)
  • Dry bulb temperature(tdry)
  • Surface dew point temperature(dp)


Back To Table of Contents

DQRID : D040530.1
Start DateStart TimeEnd DateEnd Time
05/28/2003233005/29/20030010
Subject:
TWP/SONDE/C1 - Bad RH sensor
DataStreams:twpsondewnpnC1.a1, twpsondewnpnC1.b1
Description:
It appears that the RH sensor heating circuit failed during this portion of the sounding 
(approximately between the surface and 11km).  The RH oscillates rapidly between low (~0 
%RH) and high (~80 %RH) and the oscillation seems to stop when the temperature is below 
~40 degC.  According to Vaisala these are symptoms of failure in the ASIC (application 
specific integrated circuit) that controls the heating.
Measurements:twpsondewnpnC1.b1:
  • Surface dew point temperature(dp)
  • Relative humidity inside the instrument enclosure(rh)

twpsondewnpnC1.a1:
  • Relative humidity inside the instrument enclosure(rh)
  • Surface dew point temperature(dp)


Back To Table of Contents