Data Quality Reports for Session: 112901 User: gaustad Completed: 07/03/2008


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D021229.1TWP/SONDE/C1 - Broken Temperature SensortwpsondewnpnC1.a1, twpsondewnpnC1.b1
D021231.2TWP/Sonde/C1 - Bad RH ValuestwpsondewnpnC1.a1, twpsondewnpnC1.b1
D030427.2TWP/SONDE/C1 - Questionable RHtwpsondewnpnC1.a1, twpsondewnpnC1.b1
D030503.1TWP/SONDE/C1 - Bad RHtwpsondewnpnC1.a1, twpsondewnpnC1.b1
D030603.3TWP/SONDE/C1 - Broken temperature sensortwpsondewnpnC1.a1, twpsondewnpnC1.b1
D040530.1TWP/SONDE/C1 - Bad RH sensortwpsondewnpnC1.a1, twpsondewnpnC1.b1


DQRID : D021229.1
Start DateStart TimeEnd DateEnd Time
09/18/2002233409/19/20020111
09/21/2002233009/22/20020059
10/01/2002234910/02/20020117
10/13/2002113910/13/20021303
10/18/2002235810/19/20020113
05/23/2003233005/24/20030135
06/14/2003233006/15/20030147
06/22/2003115906/22/20031332
06/22/2003234906/23/20030110
06/23/2003114606/23/20031323
06/25/2003234206/26/20030118
06/28/2003233006/29/20030056
06/29/2003113406/29/20031312
Subject:
TWP/SONDE/C1  - Broken Temperature Sensor
DataStreams:twpsondewnpnC1.a1, twpsondewnpnC1.b1
Description:
The temperature sensor on these radiosondes broke after launch.
All the temperature related data are incorrect.
Measurements:twpsondewnpnC1.b1:
  • Surface dew point temperature(dp)
  • Relative humidity inside the instrument enclosure(rh)
  • Dry bulb temperature(tdry)

twpsondewnpnC1.a1:
  • Relative humidity inside the instrument enclosure(rh)
  • Dry bulb temperature(tdry)
  • Surface dew point temperature(dp)


Back To Table of Contents

DQRID : D021231.2
Start DateStart TimeEnd DateEnd Time
12/21/2002233612/22/20020119
Subject:
TWP/Sonde/C1 - Bad RH Values
DataStreams:twpsondewnpnC1.a1, twpsondewnpnC1.b1
Description:
The RH values from the surface to approximately 11.5 km oscillate between reasonable 
values and zero.  This is apparently caused by a failure in the RH sensor heating process.  
Although the oscillation stops when the heating stops (at temperatures below ~ -45degC) I 
think the values must be viewed as suspect.
Measurements:twpsondewnpnC1.b1:
  • Surface dew point temperature(dp)
  • Relative humidity inside the instrument enclosure(rh)

twpsondewnpnC1.a1:
  • Relative humidity inside the instrument enclosure(rh)
  • Surface dew point temperature(dp)


Back To Table of Contents

DQRID : D030427.2
Start DateStart TimeEnd DateEnd Time
04/08/2003234704/09/20030003
Subject:
TWP/SONDE/C1 - Questionable RH
DataStreams:twpsondewnpnC1.a1, twpsondewnpnC1.b1
Description:
It appears that the RH sensor heating circuit failed during this portion of the sounding 
(approximately between 5km and 10km).  The RH oscillates rapidly between low (~5 %RH) and 
not so low (~50 %RH) and the oscillation seems to stop when the temperature is below ~35 
degC.  According to Vaisala these are symptoms of failure in the ASIC (application 
specific integrated circuit) that controls the heating.
Measurements:twpsondewnpnC1.b1:
  • Surface dew point temperature(dp)
  • Relative humidity inside the instrument enclosure(rh)

twpsondewnpnC1.a1:
  • Relative humidity inside the instrument enclosure(rh)
  • Surface dew point temperature(dp)


Back To Table of Contents

DQRID : D030503.1
Start DateStart TimeEnd DateEnd Time
04/29/2003112304/29/20031205
Subject:
TWP/SONDE/C1 - Bad RH
DataStreams:twpsondewnpnC1.a1, twpsondewnpnC1.b1
Description:
The RH sensor heating circuit failed on this radiosonde.  The RH-related values obtained 
between 0 and 10000 meters oscillate between near-zero and reasonable ambient values.  
Above 10000 the RH values are likely correct.
Measurements:twpsondewnpnC1.b1:
  • Surface dew point temperature(dp)
  • Relative humidity inside the instrument enclosure(rh)

twpsondewnpnC1.a1:
  • Relative humidity inside the instrument enclosure(rh)
  • Surface dew point temperature(dp)


Back To Table of Contents

DQRID : D030603.3
Start DateStart TimeEnd DateEnd Time
05/23/2003233005/24/20030135
Subject:
TWP/SONDE/C1 - Broken temperature sensor
DataStreams:twpsondewnpnC1.a1, twpsondewnpnC1.b1
Description:
The temperature sensor on this radiosonde broke after launch.  All the temperature-related 
data are incorrect.
Measurements:twpsondewnpnC1.b1:
  • Surface dew point temperature(dp)
  • Relative humidity inside the instrument enclosure(rh)
  • Dry bulb temperature(tdry)

twpsondewnpnC1.a1:
  • Relative humidity inside the instrument enclosure(rh)
  • Dry bulb temperature(tdry)
  • Surface dew point temperature(dp)


Back To Table of Contents

DQRID : D040530.1
Start DateStart TimeEnd DateEnd Time
05/28/2003233005/29/20030010
Subject:
TWP/SONDE/C1 - Bad RH sensor
DataStreams:twpsondewnpnC1.a1, twpsondewnpnC1.b1
Description:
It appears that the RH sensor heating circuit failed during this portion of the sounding 
(approximately between the surface and 11km).  The RH oscillates rapidly between low (~0 
%RH) and high (~80 %RH) and the oscillation seems to stop when the temperature is below 
~40 degC.  According to Vaisala these are symptoms of failure in the ASIC (application 
specific integrated circuit) that controls the heating.
Measurements:twpsondewnpnC1.b1:
  • Surface dew point temperature(dp)
  • Relative humidity inside the instrument enclosure(rh)

twpsondewnpnC1.a1:
  • Relative humidity inside the instrument enclosure(rh)
  • Surface dew point temperature(dp)


Back To Table of Contents



END OF DATA