Data Quality Reports for Session: 111942 User: ql_data Completed: 05/13/2008


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D030730.1TWP/GNDRAD/C2 - IRT problemtwpgndrad60sC2.b1
D041015.3TWP/GNDRAD/C2 - incorrect IRT calibrationtwpgndrad60sC2.b1
D050504.2TWP/GNDRAD/C2 - IRT connectortwpgndrad60sC2.b1
D060630.35TWP/GNDRAD/C2 - Reprocess: Longwave Calibration errortwpgndrad60sC2.b1
D070110.1TWP/GNDRAD/C2 - IRT signaltwpgndrad60sC2.b1
D070406.6TWP/GNDRAD/C2 - Reprocess: IRT calibrationtwpgndrad60sC2.b1
D071204.1TWP/GNDRAD/C2 - Reprocessed: Incorrect calibrationtwpgndrad20sC2.a0, twpgndrad60sC2.b1


DQRID : D030730.1
Start DateStart TimeEnd DateEnd Time
07/12/2003210009/14/20031100
Subject:
TWP/GNDRAD/C2 - IRT problem
DataStreams:twpgndrad60sC2.b1
Description:
The GNDRAD IRT exhibited a positive bias compared to the PIR from 2100 to 0700 UTC each 
day. The problem was corrected by removing the signal cable, cleaning the contacts, and 
reconnecting the cable.
Measurements:twpgndrad60sC2.b1:
  • Surface Infra-Red Temperature Minima(sfc_ir_temp_min)
  • Surface Infra-Red Temperature, Standard Deviation(sfc_ir_temp_std)
  • Surface Infra-Red Temperature Maxima(sfc_ir_temp_max)
  • Effective Surface (Skin) Temperature(sfc_ir_temp)


Back To Table of Contents

DQRID : D041015.3
Start DateStart TimeEnd DateEnd Time
07/05/2004050107/19/20040206
Subject:
TWP/GNDRAD/C2 - incorrect IRT calibration
DataStreams:twpgndrad60sC2.b1
Description:
When the GNDRAD data logger was upgraded, the calibration offset for the IRT was initially 
incorrect.
Measurements:twpgndrad60sC2.b1:
  • Surface Infra-Red Temperature Minima(sfc_ir_temp_min)
  • Surface Infra-Red Temperature Maxima(sfc_ir_temp_max)
  • Effective Surface (Skin) Temperature(sfc_ir_temp)


Back To Table of Contents

DQRID : D050504.2
Start DateStart TimeEnd DateEnd Time
03/27/2005233004/03/20050240
Subject:
TWP/GNDRAD/C2 - IRT connector
DataStreams:twpgndrad60sC2.b1
Description:
The IRT had a problem with the connector of either the signal or power cable which 
produced an output of 0 volts equivalent to a measurement of 223K. The problem was corrected 
when the connectors were cleaned and tightened.
Measurements:twpgndrad60sC2.b1:
  • Surface Infra-Red Temperature Minima(sfc_ir_temp_min)
  • Surface Infra-Red Temperature, Standard Deviation(sfc_ir_temp_std)
  • Surface Infra-Red Temperature Maxima(sfc_ir_temp_max)
  • Effective Surface (Skin) Temperature(sfc_ir_temp)


Back To Table of Contents

DQRID : D060630.35
Start DateStart TimeEnd DateEnd Time
07/05/2004220212/21/20052149
Subject:
TWP/GNDRAD/C2 - Reprocess: Longwave Calibration error
DataStreams:twpgndrad60sC2.b1
Description:
Modified pyrgeometer calibration procedures were implemented beginning in December 2003.  
These modified procedures introduced a calibration bias in the longwave data.  The 
previous procedures were re-implemented at all sites between December 2005 and February 2006 to 
restore proper calibrations.                                                             
                The data collected while the incorrect procedures were in place are 
being reprocessing to remove the calibration bias.  Note, this also affected standard 
deviation, maximum and minimum data fields.
Measurements:twpgndrad60sC2.b1:
  • Upwelling (10 meter) Longwave Hemispheric Irradiance, Pyrgeometer(up_long_hemisp)


Back To Table of Contents

DQRID : D070110.1
Start DateStart TimeEnd DateEnd Time
12/03/2006040002/04/20072316
Subject:
TWP/GNDRAD/C2 - IRT signal
DataStreams:twpgndrad60sC2.b1
Description:
When the other GNDRAD radiometers were replaced on 12/3/2006, the quality of the IRT 
signal began to degrade until 12/26 when it became unusable. The instrument was replaced on 
2/4/2007.
Measurements:twpgndrad60sC2.b1:
  • Surface Infra-Red Temperature Minima(sfc_ir_temp_min)
  • Surface Infra-Red Temperature, Standard Deviation(sfc_ir_temp_std)
  • Surface Infra-Red Temperature Maxima(sfc_ir_temp_max)
  • Effective Surface (Skin) Temperature(sfc_ir_temp)


Back To Table of Contents

DQRID : D070406.6
Start DateStart TimeEnd DateEnd Time
02/04/2007231702/17/20070400
Subject:
TWP/GNDRAD/C2 - Reprocess: IRT calibration
DataStreams:twpgndrad60sC2.b1
Description:
The configuration of the IRT was set to that used at TWP/C3. This caused the calibration 
of the data to be wrong. It was corrected by reconfiguring the IRT to match the datalogger 
program.
   
These data can be reprocessed and corrected by adding 10K, i.e. sfc_ir_temp(new) = 
sfc_ir_temp(old) + 10K.
Measurements:twpgndrad60sC2.b1:
  • Surface Infra-Red Temperature Minima(sfc_ir_temp_min)
  • Surface Infra-Red Temperature Maxima(sfc_ir_temp_max)
  • Effective Surface (Skin) Temperature(sfc_ir_temp)


Back To Table of Contents

DQRID : D071204.1
Start DateStart TimeEnd DateEnd Time
11/25/2007220211/30/20071724
Subject:
TWP/GNDRAD/C2 - Reprocessed: Incorrect calibration
DataStreams:twpgndrad20sC2.a0, twpgndrad60sC2.b1
Description:
The Gndrad PSP was changed out 11/25 but the datalogger program was not updated until 
11/30.

PSP #31293 installed 20071125.2202.  Correct calibration coefficient = 123.46.

Reprocessed to correct upwelling shortwave hemispheric irradiance (US) by:
US(new) = (US(old) / 123.56 ) * 123.46

Reprocessing completed March 2008.
Measurements:twpgndrad60sC2.b1:
  • Upwelling (10 meter) Shortwave Hemispheric Irradiance, Pyranometer(up_short_hemisp)
  • Upwelling (10 meter) Shortwave Hemispheric Irradiance, Pyranometer, Maxima(up_short_hemisp_max)
  • Upwelling (10 meter) Shortwave Hemispheric Irradiance, Pyranometer, Standard
    Deviation(up_short_hemisp_std)
  • Upwelling (10 meter) Shortwave Hemispheric Irradiance, Pyranometer, Minima(up_short_hemisp_min)

twpgndrad20sC2.a0:
  • Instantaneous Upwelling Shortwave Hemispheric Irradiance, Pyranometer(inst_up_short_hemisp)


Back To Table of Contents



END OF DATA