D041026.3SGP/AOS/C1 - PCASP instrument problemsgpaosC1.a0, sgpaosauxC1.a0
D050120.2SGP/AOS/C1 - Failed RH sensorsgpaosC1.a0
D980528.1SGP/AOS/C1 - Optical Particle Counter (PCASP-X) downsgpaosC1.a0, sgpaosauxC1.a0


DQRID : D041026.3
Start DateStart TimeEnd DateEnd Time
01/10/2004000010/26/20041200
Subject:
SGP/AOS/C1 - PCASP instrument problem
DataStreams:sgpaosC1.a0, sgpaosauxC1.a0
Description:
PCASP (optical particle counter) gave unusually low aerosol signal counts despite numerous 
attempts to fix the instrument by the mentor and manufacturer.
Measurements:sgpaosauxC1.a0:
  • OPC sheath(OPCSheath)
  • OPC sample flow(OPCFlow)
  • Analog signal from PMS PCASP(OPCLaserVolt)

sgpaosC1.a0:
  • PMS PCASP Chan. 7 (0.23 micrometers < Dp < 0.26)(pa23_p26conc)
  • PMS PCASP Chan. 26 (3.00 micrometers < Dp < 3.50)(pa300_p350conc)
  • PMS PCASP Chan. 28 (4.00 micrometers < Dp < 5.00)(pa400_p500conc)
  • PMS PCASP Chan. 15 (0.70 micrometers < Dp < 0.80)(pa70_p80conc)
  • PMS PCASP Chan. 25 (2.60 micrometers < Dp < 3.00)(pa260_p300conc)
  • PMS PCASP Chan. 0 (Dp > 10 micrometers)(Pa1000Conc)
  • PMS PCASP Chan. 2 (0.12 micrometers < Dp < 0.14)(pa12_p14conc)
  • PMS PCASP Chan. 30 (6.50 micrometers < Dp < 8.00)(pa650_p800conc)
  • PMS PCASP Chan. 6 (0.20 micrometers < Dp < 0.23)(pa20_p23conc)
  • PMS PCASP Chan. 19 (1.30 micrometers < Dp < 1.40)(pa130_p140conc)
  • PMS PCASP Chan. 17 (0.90 micrometers < Dp < 1.00)(pa90_p100conc)
  • PMS PCASP Chan. 18 (1.00 micrometers < Dp < 1.30)(pa100_p130conc)
  • PMS PCASP Chan. 3 (0.14 micrometers < Dp < 0.16)(pa14_p16conc)
  • PMS PCASP Chan. 31 (8.00 micrometers < Dp < 10.0)(pa800_p1000conc)
  • PMS PCASP Chan. 1 1/cm^3 (0.10 micrometers < Dp < 0.12)(pa10_p12conc)
  • PMS PCASP Chan. 23 1/cm^3 (2.00 micrometers < Dp < 2.30)(pa200_p230conc)
  • PMS PCASP Chan. 24 (2.30 micrometers < Dp < 2.60)(pa230_p260conc)
  • PMS PCASP Chan. 21 (1.60 micrometers < Dp < 1.80)(pa160_p180conc)
  • PMS PCASP Chan. 22 (1.80 micrometers < Dp < 2.00)(pa180_p200conc)
  • PMS PCASP Chan. 9 (0.30 micrometers < Dp < 0.35)(pa30_p35conc)
  • PMS PCASP Chan. 27 (3.50 micrometers < Dp < 4.00)(pa350_p400conc)
  • PMS PCASP Chan. 13 (0.50 micrometers < Dp < 0.60)(pa50_p60conc)
  • PMS PCASP Chan. 4 (0.16 micrometers < Dp < 0.18)(pa16_p18conc)
  • PMS PCASP Chan. 20 (1.40 micrometers < Dp < 1.60)(pa140_p160conc)
  • PMS PCASP Chan. 8 (0.26 micrometers < Dp < 0.30)(pa26_p30conc)
  • PMS PCASP Chan. 5 (0.18 micrometers < Dp < 0.20)(pa18_p20conc)
  • PMS PCASP Chan. 14 (0.60 micrometers < Dp < 0.70)(pa60_p70conc)
  • PMS PCASP Chan. 10 (0.35 micrometers < Dp < 0.40)(pa35_p40conc)
  • PMS PCASP Chan. 11 (0.35 micrometers < Dp < 0.40)(pa40_p45conc)
  • PMS PCASP Chan. 16 (0.80 micrometers < Dp < 0.90)(pa80_p90conc)
  • PMS PCASP Chan. 12 (0.45 micrometers < Dp < 0.50)(pa45_p50conc)
  • PMS PCASP Chan. 29 (5.00 micrometers < Dp < 6.50)(pa500_p650conc)


Back To Table of Contents

DQRID : D050120.2
Start DateStart TimeEnd DateEnd Time
01/01/2004000005/01/20052359
Subject:
SGP/AOS/C1 - Failed RH sensor
DataStreams:sgpaosC1.a0
Description:
The RH sensor inside the humidified nephelometer began to slowly decline in value in the 
upper RH range beginning in early 2004 and becoming more noticable poor on 20041117. The 
sensor currently reaches a high RH value of about 60% when the actual RH is closer to 80%.

The sensor was replaced in the Spring of 2005, but the problem remained.  Effective 
1/1/2004, this sensor is not used to calculate scientific data and is therefore essentially 
abandoned in the field.  An end-date of 20050501 is being used as a placeholder.
Measurements:sgpaosC1.a0:
  • TSI High RH Neph. relative humidity(TSINephRHSamp_HRH)


Back To Table of Contents

DQRID : D980528.1
Start DateStart TimeEnd DateEnd Time
07/01/1997093607/03/19971556
08/01/1997132908/08/19971348
05/05/1998092606/16/19981505
06/09/1999082407/26/19992205
09/08/2000042610/13/20001630
05/17/2002060205/20/20051847
05/23/2002201105/31/20021258
09/09/2002000010/07/20021905
Subject:
SGP/AOS/C1 - Optical Particle Counter (PCASP-X) down
DataStreams:sgpaosC1.a0, sgpaosauxC1.a0
Description:
The AOS Optical Particle Spectrometer (PCASP-X sizing probe) was down during these 
periods.  This lowers the laser reference voltage and results in inaccurate sizing.
Measurements:sgpaosauxC1.a0:
  • OPC sheath(OPCSheath)
  • OPC sample flow(OPCFlow)
  • Analog signal from PMS PCASP(OPCLaserVolt)

sgpaosC1.a0:
  • PMS PCASP Chan. 7 (0.23 micrometers < Dp < 0.26)(pa23_p26conc)
  • PMS PCASP Chan. 26 (3.00 micrometers < Dp < 3.50)(pa300_p350conc)
  • PMS PCASP Chan. 28 (4.00 micrometers < Dp < 5.00)(pa400_p500conc)
  • PMS PCASP Chan. 15 (0.70 micrometers < Dp < 0.80)(pa70_p80conc)
  • PMS PCASP Chan. 25 (2.60 micrometers < Dp < 3.00)(pa260_p300conc)
  • PMS PCASP Chan. 0 (Dp > 10 micrometers)(Pa1000Conc)
  • PMS PCASP Chan. 2 (0.12 micrometers < Dp < 0.14)(pa12_p14conc)
  • PMS PCASP Chan. 30 (6.50 micrometers < Dp < 8.00)(pa650_p800conc)
  • PMS PCASP Chan. 6 (0.20 micrometers < Dp < 0.23)(pa20_p23conc)
  • PMS PCASP Chan. 19 (1.30 micrometers < Dp < 1.40)(pa130_p140conc)
  • PMS PCASP Chan. 17 (0.90 micrometers < Dp < 1.00)(pa90_p100conc)
  • PMS PCASP Chan. 18 (1.00 micrometers < Dp < 1.30)(pa100_p130conc)
  • PMS PCASP Chan. 3 (0.14 micrometers < Dp < 0.16)(pa14_p16conc)
  • PMS PCASP Chan. 31 (8.00 micrometers < Dp < 10.0)(pa800_p1000conc)
  • PMS PCASP Chan. 1 1/cm^3 (0.10 micrometers < Dp < 0.12)(pa10_p12conc)
  • PMS PCASP Chan. 23 1/cm^3 (2.00 micrometers < Dp < 2.30)(pa200_p230conc)
  • PMS PCASP Chan. 24 (2.30 micrometers < Dp < 2.60)(pa230_p260conc)
  • PMS PCASP Chan. 21 (1.60 micrometers < Dp < 1.80)(pa160_p180conc)
  • PMS PCASP Chan. 22 (1.80 micrometers < Dp < 2.00)(pa180_p200conc)
  • PMS PCASP Chan. 9 (0.30 micrometers < Dp < 0.35)(pa30_p35conc)
  • PMS PCASP Chan. 27 (3.50 micrometers < Dp < 4.00)(pa350_p400conc)
  • PMS PCASP Chan. 13 (0.50 micrometers < Dp < 0.60)(pa50_p60conc)
  • PMS PCASP Chan. 4 (0.16 micrometers < Dp < 0.18)(pa16_p18conc)
  • PMS PCASP Chan. 20 (1.40 micrometers < Dp < 1.60)(pa140_p160conc)
  • PMS PCASP Chan. 8 (0.26 micrometers < Dp < 0.30)(pa26_p30conc)
  • PMS PCASP Chan. 5 (0.18 micrometers < Dp < 0.20)(pa18_p20conc)
  • PMS PCASP Chan. 14 (0.60 micrometers < Dp < 0.70)(pa60_p70conc)
  • PMS PCASP Chan. 10 (0.35 micrometers < Dp < 0.40)(pa35_p40conc)
  • PMS PCASP Chan. 11 (0.35 micrometers < Dp < 0.40)(pa40_p45conc)
  • PMS PCASP Chan. 16 (0.80 micrometers < Dp < 0.90)(pa80_p90conc)
  • PMS PCASP Chan. 12 (0.45 micrometers < Dp < 0.50)(pa45_p50conc)
  • PMS PCASP Chan. 29 (5.00 micrometers < Dp < 6.50)(pa500_p650conc)


Back To Table of Contents