Data Quality Reports for Session: 110911 User: hacker Completed: 03/07/2008


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D030306.2SGP/IRT25m/C1 - Odd Instantaneous Surface IRT datasgpirt25mC1.a1
D030730.3SGP/IRT25m/C1 - error in data logger programsgpirt25m20sC1.a0, sgpirt25mC1.a1


DQRID : D030306.2
Start DateStart TimeEnd DateEnd Time
11/21/2002000006/05/20032001
Subject:
SGP/IRT25m/C1 - Odd Instantaneous Surface IRT data
DataStreams:sgpirt25mC1.a1
Description:
The sgpirt25mC1.a1 data for variable
   inst_sfc_ir_temp
look odd.  Each day, the first two records and the last 58 records have reasonable data 
but the remaining records are all zero. This behavior is seen all the way back to 20021121 
when the instantaneous variables were added to the sgpirt25mC1.a1 data stream.

The other instantaneous data variables do not show a similar pattern and the 60 sec 
sfc_ir_temp data look reasonable.
Measurements:sgpirt25mC1.a1:
  • Instantaneous Surface Infra-red Temperature(inst_sfc_ir_temp)


Back To Table of Contents

DQRID : D030730.3
Start DateStart TimeEnd DateEnd Time
12/16/2002170005/21/20031807
Subject:
SGP/IRT25m/C1 - error in data logger program
DataStreams:sgpirt25m20sC1.a0, sgpirt25mC1.a1
Description:
On 12/16/2002, the logger software for the IRT25m was
inappropriately edited. This resulted in incorrect instantaneous
values at 20 and 40 seconds after each minute and incorrect 
calculation of 60-second averaged instantaneous values (because
the 20 and 40 second values are two of three instantaneous
values used to calculate this average). On 5/21/2003, the data
logger program was corrected.
Measurements:sgpirt25mC1.a1:
  • Instantaneous Surface Infra-red Temperature(inst_sfc_ir_temp)
  • Instantaneous Upwelling Longwave Hemispheric Irradiance, Pyrgeometer(inst_up_long_hemisp)
  • Instantaneous Dome Temperature(inst_dome_temp)
  • Instantaneous Dome Temperature(inst_up_long_dome_temp)
  • Instantaneous Case Temperature(inst_up_long_case_temp)
  • Instantaneous Upwelling Shortwave Hemispheric Irradiance, Pyranometer(inst_up_short_hemisp)
  • Instantaneous Case Temperature(inst_case_temp)

sgpirt25m20sC1.a0:
  • Instantaneous Upwelling Pygeometer Thermopile Voltage(inst_up_long_hemisp_tp)
  • Instantaneous Upwelling Pyrgeometer Dome Thermistor Resistance, Pyrgeometer(inst_up_long_dome_resist)
  • Instantaneous Surface Infra-red Temperature(inst_sfc_ir_temp)
  • Instantaneous Upwelling Pyrgeometer Case Thermistor Resistance, Pyrgeometer(inst_up_long_case_resist)
  • Instantaneous Upwelling Pyrgeometer Thermopile Voltage, Pyrgeometer(inst_up_short_hemisp_tp)


Back To Table of Contents



END OF DATA