Data Quality Reports for Session: 110877 User: smayaki Completed: 03/05/2008


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D060630.42NIM/GNDRAD/M1 - Reprocessed: Longwave calibration errornimgndrad60sM1.b1
D061003.4NIM/AERI/M1 - IRTs do not agree with AERInimaerisummaryM1.b1


DQRID : D060630.42
Start DateStart TimeEnd DateEnd Time
11/25/2005170012/22/20051100
Subject:
NIM/GNDRAD/M1 - Reprocessed: Longwave calibration error
DataStreams:nimgndrad60sM1.b1
Description:
Modified pyrgeometer calibration procedures were implemented beginning in December 2003. 
These modified procedures introduced a calibration bias in the longwave data.  The 
previous procedures were re-implemented at all sites between December 2005 and February 2006 to 
restore proper calibrations.									     
The data collected while the incorrect procedures were in place have been reprocessed to 
remove the calibration bias.  The reprocessed 60 second averaged data are based on 3 
instantaneous 20 second data records rather than on 60 1 second instantaneous data records.  
Still, these data are considered far superior to the originally processed data. The 
reprocessed data were archived in February 2007.
Measurements:nimgndrad60sM1.b1:
  • Upwelling (10 meter) Longwave Hemispheric Irradiance, Pyrgeometer(up_long_hemisp)


Back To Table of Contents

DQRID : D061003.4
Start DateStart TimeEnd DateEnd Time
01/07/2006000906/07/20061900
Subject:
NIM/AERI/M1 - IRTs do not agree with AERI
DataStreams:nimaerisummaryM1.b1
Description:
Since deployment at PYE, and then at NIM, the AMF SKYRAD IRT measured about 10K higher sky 
temperatures than the AERI and the MWRP IRT measured about 20K higher than the AERI.  
Several actions were taken to diagnose the problem including confirming the correct 
configuration of the IRTs and data logger, cleaning the mirror and lens, and replacing the mirror.

After several days of rain beginning 6/2/2006, the three instruments came into agreement. 
It is unknown whether this was a problem with the AERI, MWRP-IRT or SKYRAD-IRT.
Measurements:nimaerisummaryM1.b1:
  • AERI SW Scene Brightness Temp Spectral Averages (Ch2)(SkyBrightnessTempSpectralAveragesCh2)
  • AERI LW Scene Brightness Temp Spectral Averages (Ch1)(SkyBrightnessTempSpectralAveragesCh1)


Back To Table of Contents



END OF DATA