Data Quality Reports for Session: 110016 User: perez Completed: 01/18/2008


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D020827.1SGP/NIMFR/E13 - NIMFR Centrovision filter leakagesgpnimfrE13.a0, sgpnimfrE13.b1
D020827.2SGP/NIMFR/E13 - Centrovision filter leakagesgpnimfrE13.a1


DQRID : D020827.1
Start DateStart TimeEnd DateEnd Time
04/01/2001000011/29/20012359
Subject:
SGP/NIMFR/E13 - NIMFR Centrovision filter leakage
DataStreams:sgpnimfrE13.a0, sgpnimfrE13.b1
Description:
Testing of the batch of centrovision filters being used in the NIMFR EF13
instrument indicate that the first four filters leak 'out of band' light. The
leakage for each filter is (1)415nm 3%, (2)500nm 1-2%, (3)615nm 1-2%, (4)673nm
3%. Acceptable leakage is less than 0.5%.

This leakage causes an error in the irradiance values (too high) at these four
wavelengths and the optical depths calculated using this data will be too high.
Measurements:sgpnimfrE13.b1:
  • Direct Normal Narrowband Irradiance, Filter 1(direct_norm_narrowband_filter1)
  • Direct Normal Narrowband Irradiance, Filter 3(direct_norm_narrowband_filter3)
  • Direct Normal Narrowband Irradiance, Filter 2(direct_norm_narrowband_filter2)
  • Direct Normal Narrowband Irradiance, Filter 4(direct_norm_narrowband_filter4)

sgpnimfrE13.a0:
  • Direct Normal Narrowband Irradiance, Filter 4(direct_norm_narrowband_filter4)
  • Direct Normal Narrowband Irradiance, Filter 1(direct_norm_narrowband_filter1)
  • Direct Normal Narrowband Irradiance, Filter 2(direct_norm_narrowband_filter2)
  • Direct Normal Narrowband Irradiance, Filter 3(direct_norm_narrowband_filter3)


Back To Table of Contents

DQRID : D020827.2
Start DateStart TimeEnd DateEnd Time
04/17/2000010111/29/20012359
Subject:
SGP/NIMFR/E13 - Centrovision filter leakage
DataStreams:sgpnimfrE13.a1
Description:
Testing of the batch of centrovision filters being used in the NIMFR EF13
instrument indicate that the first four filters leak 'out of band' light. The
leakage for each filter is (1)415nm 3%, (2)500nm 1-2%, (3)615nm 1-2%, (4)673nm
3%. Acceptable leakage is less than 0.5%.

This leakage causes an error in the irradiance values (too high) at these four
wavelengths and the optical depths calculated using this data will be too high
Measurements:sgpnimfrE13.a1:
  • Direct Normal Narrowband Irradiance, Filter 2(direct_norm_narrowband_filter2)
  • Direct Normal Narrowband Irradiance, Filter 4(direct_norm_narrowband_filter4)
  • Direct Normal Narrowband Irradiance, Filter 1(direct_norm_narrowband_filter1)
  • Direct Normal Narrowband Irradiance, Filter 3(direct_norm_narrowband_filter3)


Back To Table of Contents



END OF DATA