Data Quality Reports for Session: 109970 User: kirstie Completed: 01/17/2008


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D040213.4NSA/SKYRAD/C1 - Downwelling IRT biasnsaskyrad60sC1.b1
D040421.2NSA/SKYRAD/C1 - Tracker Inoperativensaskyrad60sC1.b1
D040421.3NSA/SKYRAD/C1 - itermittant IRT biasnsaskyrad60sC1.b1
D040919.1NSA/GNDRAD/C1 - IRT power failurensagndrad60sC1.b1
D060630.39NSA/GNDRAD/C1 - Reprocess: Longwave Calibration errornsagndrad60sC1.b1


DQRID : D040213.4
Start DateStart TimeEnd DateEnd Time
12/11/2003120001/09/20042120
Subject:
NSA/SKYRAD/C1 - Downwelling IRT bias
DataStreams:nsaskyrad60sC1.b1
Description:
The downwelling IRT was measuring higher sky/cloud temperatures than the AERI. The problem 
was corrected when the mirror was replaced.
Measurements:nsaskyrad60sC1.b1:
  • Sky Infra Red Temperature Minima(sky_ir_temp_min)
  • Sky Infra Red Temperature Maxima(sky_ir_temp_max)
  • Standard Deviation of Sky Infra Red Temperature(sky_ir_temp_std)
  • Sky Infra-Red Temperature(sky_ir_temp)


Back To Table of Contents

DQRID : D040421.2
Start DateStart TimeEnd DateEnd Time
11/10/2003000002/16/20040230
Subject:
NSA/SKYRAD/C1 - Tracker Inoperative
DataStreams:nsaskyrad60sC1.b1
Description:
A blown fuse resulted in the tracker not shading the PIR and B&W
and the NIP not following the sun.
Measurements:nsaskyrad60sC1.b1:
  • Downwelling Shortwave Diffuse Hemispheric Irradiance, Ventilated Pyranometer,
    Minima(down_short_diffuse_hemisp_min)
  • Downwelling Shortwave Diffuse Hemispheric Irradiance, Ventilated Pyranometer(down_short_diffuse_hemisp)
  • Shortwave Direct Normal Irradiance, Pyrgeometer, Minima(short_direct_normal_min)
  • Shortwave Direct Normal Irradiance, Pyrgeometer, Standard Deviation(short_direct_normal_std)
  • Shortwave Direct Normal Irradiance, Pyrgeometer, Maxima(short_direct_normal_max)
  • Downwelling Shortwave Hemispheric Irradiance, Pyranometer, Standard Deviation(down_short_diffuse_hemisp_std)
  • Downwelling Shortwave Diffuse Hemispheric Irradiance, Ventilated Pyranometer,
    Maxima(down_short_diffuse_hemisp_max)
  • Shortwave Direct Normal Irradiance, Pyrgeometer(short_direct_normal)


Back To Table of Contents

DQRID : D040421.3
Start DateStart TimeEnd DateEnd Time
03/28/2003020006/21/20060215
Subject:
NSA/SKYRAD/C1 - itermittant IRT bias
DataStreams:nsaskyrad60sC1.b1
Description:
A problem with the SKYRAD IRT that itermittantly produced a gradual increase in the sky 
temperature measurement then a sudden correction, as compared to AERI, appears to have been 
caused by a combination of a failing IR detector, a poor signal cable, a contaminated 
(with condensation or frost) mirror or lens, and low ambient temperatures. The positive 
bias was apparently not consistant in either frequency or duration but on days when the 
unusual behavior is observed, it usually occurred from about 00:00 to 06:00 local time. 

The instrument was repaired and recalibrated and the signal cable replaced.
Measurements:nsaskyrad60sC1.b1:
  • Sky Infra Red Temperature Minima(sky_ir_temp_min)
  • Sky Infra Red Temperature Maxima(sky_ir_temp_max)
  • Standard Deviation of Sky Infra Red Temperature(sky_ir_temp_std)
  • Sky Infra-Red Temperature(sky_ir_temp)


Back To Table of Contents

DQRID : D040919.1
Start DateStart TimeEnd DateEnd Time
09/15/2004000009/22/20040552
Subject:
NSA/GNDRAD/C1 - IRT power failure
DataStreams:nsagndrad60sC1.b1
Description:
After performing corrective maintenance which involved turning off power to the GNDRAD 
instruments and data logger, power to the IRT was not restored. The problem was corrected by 
replacing the data logger.
Measurements:nsagndrad60sC1.b1:
  • Surface Infra-Red Temperature Maxima(sfc_ir_temp_max)
  • Surface Infra-Red Temperature Minima(sfc_ir_temp_min)
  • Effective Surface (Skin) Temperature(sfc_ir_temp)
  • Surface Infra-Red Temperature, Standard Deviation(sfc_ir_temp_std)


Back To Table of Contents

DQRID : D060630.39
Start DateStart TimeEnd DateEnd Time
11/04/2004160012/21/20051855
Subject:
NSA/GNDRAD/C1 - Reprocess: Longwave Calibration error
DataStreams:nsagndrad60sC1.b1
Description:
Modified pyrgeometer calibration procedures were implemented beginning in December 2003.  
These modified procedures introduced a calibration bias in the longwave data.  The 
previous procedures were re-implemented at all sites between December 2005 and February 2006 to 
restore proper calibrations.                                                             
                The data collected while the incorrect procedures were in place are 
being reprocessing to remove the calibration bias.  Note, this also affected standard 
deviation, maximum and minimum data fields.
Measurements:nsagndrad60sC1.b1:
  • Upwelling (10 meter) Longwave Hemispheric Irradiance, Pyrgeometer(up_long_hemisp)


Back To Table of Contents



END OF DATA