Data Quality Reports for Session: 109969 User: kirstie Completed: 01/17/2008


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D040421.3NSA/SKYRAD/C1 - itermittant IRT biasnsaskyrad60sC1.b1
D051101.2NSA/SKYRAD/C1 - IRT failurensaskyrad60sC1.b1
D060630.39NSA/GNDRAD/C1 - Reprocess: Longwave Calibration errornsagndrad60sC1.b1


DQRID : D040421.3
Start DateStart TimeEnd DateEnd Time
03/28/2003020006/21/20060215
Subject:
NSA/SKYRAD/C1 - itermittant IRT bias
DataStreams:nsaskyrad60sC1.b1
Description:
A problem with the SKYRAD IRT that itermittantly produced a gradual increase in the sky 
temperature measurement then a sudden correction, as compared to AERI, appears to have been 
caused by a combination of a failing IR detector, a poor signal cable, a contaminated 
(with condensation or frost) mirror or lens, and low ambient temperatures. The positive 
bias was apparently not consistant in either frequency or duration but on days when the 
unusual behavior is observed, it usually occurred from about 00:00 to 06:00 local time. 

The instrument was repaired and recalibrated and the signal cable replaced.
Measurements:nsaskyrad60sC1.b1:
  • Sky Infra Red Temperature Minima(sky_ir_temp_min)
  • Sky Infra Red Temperature Maxima(sky_ir_temp_max)
  • Standard Deviation of Sky Infra Red Temperature(sky_ir_temp_std)
  • Sky Infra-Red Temperature(sky_ir_temp)


Back To Table of Contents

DQRID : D051101.2
Start DateStart TimeEnd DateEnd Time
10/31/2005060001/18/20060032
Subject:
NSA/SKYRAD/C1 - IRT failure
DataStreams:nsaskyrad60sC1.b1
Description:
The IRT failed and was replaced with a spare.
Measurements:nsaskyrad60sC1.b1:
  • Sky Infra Red Temperature Minima(sky_ir_temp_min)
  • Sky Infra Red Temperature Maxima(sky_ir_temp_max)
  • Standard Deviation of Sky Infra Red Temperature(sky_ir_temp_std)
  • Sky Infra-Red Temperature(sky_ir_temp)


Back To Table of Contents

DQRID : D060630.39
Start DateStart TimeEnd DateEnd Time
11/04/2004160012/21/20051855
Subject:
NSA/GNDRAD/C1 - Reprocess: Longwave Calibration error
DataStreams:nsagndrad60sC1.b1
Description:
Modified pyrgeometer calibration procedures were implemented beginning in December 2003.  
These modified procedures introduced a calibration bias in the longwave data.  The 
previous procedures were re-implemented at all sites between December 2005 and February 2006 to 
restore proper calibrations.                                                             
                The data collected while the incorrect procedures were in place are 
being reprocessing to remove the calibration bias.  Note, this also affected standard 
deviation, maximum and minimum data fields.
Measurements:nsagndrad60sC1.b1:
  • Upwelling (10 meter) Longwave Hemispheric Irradiance, Pyrgeometer(up_long_hemisp)


Back To Table of Contents



END OF DATA