Data Quality Reports for Session: 109678 User: jean1charl Completed: 01/04/2008


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D021206.1TWP/SONDE/C2 - Broken Temperature SensortwpsondewnpnC2.a1, twpsondewnpnC2.b1
D030419.1TWP/SONDE/C2 - Bad RH Data (AIRS Validation Sonde)twpsondewnpnC2.a1, twpsondewnpnC2.b1
D030608.4TWP/SONDE/C2 - Failure of RH sensortwpsondewnpnC2.a1, twpsondewnpnC2.b1
D031008.1TWP/SONDE/C2 - Strange problem with data -twpsondeC2.00, twpsondewnpnC2.00, twpsondewnpnC2.b1
D031231.1TWP/SONDE/C2 - RH sensor badtwpsondewnpnC2.b1
D040127.3TWP/SONDE/C2 - Broken temperature sensorstwpsondewnpnC2.b1
D070920.2TWP/SONDE/C2 - Failure of RH sensortwpsondewnpnC2.b1


DQRID : D021206.1
Start DateStart TimeEnd DateEnd Time
09/18/2002113709/18/20021301
09/24/2002012909/24/20020204
09/24/2002022009/24/20020352
11/23/2002233011/24/20020108
11/25/2002233711/26/20020001
12/16/2002113812/16/20021248
12/22/2002013612/22/20020204
12/24/2002233112/25/20020112
12/27/2002234412/28/20020131
03/28/2003021403/28/20030339
04/22/2003234104/23/20030122
04/23/2003112904/23/20031308
04/24/2003235804/25/20030126
06/02/2003113406/02/20031321
06/06/2003233506/07/20030057
06/09/2003234006/10/20030116
06/11/2003234506/12/20030112
06/14/2003113306/14/20031300
06/26/2003233506/27/20030114
09/27/2003232709/28/20030132
Subject:
TWP/SONDE/C2  - Broken Temperature Sensor
DataStreams:twpsondewnpnC2.a1, twpsondewnpnC2.b1
Description:
The temperature sensors on these radiosondes broke at launch.  All
the temperature-related data are incorrect.
Measurements:twpsondewnpnC2.b1:
  • Relative humidity inside the instrument enclosure(rh)
  • Dry bulb temperature(tdry)
  • Surface dew point temperature(dp)

twpsondewnpnC2.a1:
  • Surface dew point temperature(dp)
  • Dry bulb temperature(tdry)
  • Relative humidity inside the instrument enclosure(rh)


Back To Table of Contents

DQRID : D030419.1
Start DateStart TimeEnd DateEnd Time
04/13/2003012404/13/20030200
Subject:
TWP/SONDE/C2 - Bad RH Data (AIRS Validation Sonde)
DataStreams:twpsondewnpnC2.a1, twpsondewnpnC2.b1
Description:
The RH sensor heating circuit failed on this sonde.  This results in
oscillation of reported RH between reasonable values and values near zero.
This was an AIRS validation sounding.
Measurements:twpsondewnpnC2.b1:
  • Relative humidity inside the instrument enclosure(rh)
  • Surface dew point temperature(dp)

twpsondewnpnC2.a1:
  • Surface dew point temperature(dp)
  • Relative humidity inside the instrument enclosure(rh)


Back To Table of Contents

DQRID : D030608.4
Start DateStart TimeEnd DateEnd Time
06/03/2003233906/04/20030016
Subject:
TWP/SONDE/C2  - Failure of RH sensor
DataStreams:twpsondewnpnC2.a1, twpsondewnpnC2.b1
Description:
It appears that the RH sensor heating circuit failed during this sounding.  The RH 
oscillates rapidly between low (~0 %RH) and not so low (ambient %RH) and the oscillation seems 
to stop when the temperature is below ~35 degC.  According to Vaisala these are symptoms 
of failure in the ASIC (application specific integrated circuit) that controls the heating.
Measurements:twpsondewnpnC2.b1:
  • Relative humidity inside the instrument enclosure(rh)
  • Surface dew point temperature(dp)

twpsondewnpnC2.a1:
  • Surface dew point temperature(dp)
  • Relative humidity inside the instrument enclosure(rh)


Back To Table of Contents

DQRID : D031008.1
Start DateStart TimeEnd DateEnd Time
10/02/2003023210/02/20030300
Subject:
TWP/SONDE/C2 - Strange problem with data -
DataStreams:twpsondeC2.00, twpsondewnpnC2.00, twpsondewnpnC2.b1
Description:
This is very strange.  The data has what appears to be a pressure jump at launch (from 
1007.2 to 956.5 hPa) but the sounding temperature and RH data clearly look to be 
interpolated between the surface and tropopause.  By examining the absolutely raw data, I found that 
the system assigned an elapsed time of 408 seconds (since system start up) to the launch 
point, but there is no raw data between 410 and 532 seconds.  The raw pressure value at 
410 seconds is 1007.6 hPa and that at 532 seconds is 40.9 hPa!  Unless this sounding was 
done by using a rocket-powered balloon (maybe even then), it would be impossible for the 
sonde to get to 40.9 hPa in 2 minutes.  Because the apparent gap is only two minutes, the 
system was able to interpolate but is not sophisticated enough to understand that the 
limits of the interpolation are unreasonable.

I can't imagine what went wrong - I've not seen anything like this before.
Measurements:twpsondeC2.00:
  • Raw data stream - documentation not supported(Raw data stream - documentation not supported)

twpsondewnpnC2.b1:
  • Dry bulb temperature(tdry)
  • Ascent Rate(asc)
  • lon(lon)
  • lat(lat)
  • Wind Status(wstat)
  • Mean Wind Speed(wspd)
  • V-component(v_wind)
  • Retrieved pressure profile(pres)
  • Time offset of tweaks from base_time(time_offset)
  • Relative humidity inside the instrument enclosure(rh)
  • Surface dew point temperature(dp)
  • base time(base_time)
  • U-component(u_wind)
  • Wind Direction(deg)
  • Dummy altitude for Zeb(alt)

twpsondewnpnC2.00:
  • Raw data stream - documentation not supported(Raw data stream - documentation not supported)


Back To Table of Contents

DQRID : D031231.1
Start DateStart TimeEnd DateEnd Time
10/22/2003002510/22/20030155
10/22/2003234010/23/20030121
Subject:
TWP/SONDE/C2 - RH sensor bad
DataStreams:twpsondewnpnC2.b1
Description:
It appears that the RH sensor heating circuits failed during these
soundings.  The RH oscillates rapidly between low (~0 %RH) and not
so low (ambient %RH) and the oscillation seems to stop when the
temperature is below ~35 degC.  According to Vaisala these are
symptoms of failure in the ASIC (application specific integrated
circuit) that controls the heating.
Measurements:twpsondewnpnC2.b1:
  • Relative humidity inside the instrument enclosure(rh)
  • Surface dew point temperature(dp)


Back To Table of Contents

DQRID : D040127.3
Start DateStart TimeEnd DateEnd Time
10/13/2003004010/13/20030203
11/08/2003233511/09/20030215
11/18/2003120511/18/20031340
11/19/2003115611/19/20031346
11/21/2003004411/21/20030113
12/20/2003122512/20/20031408
12/21/2003005612/21/20030126
12/21/2003115512/21/20031344
12/23/2003004212/23/20030212
12/23/2003113712/23/20031259
12/23/2003234212/24/20030132
01/08/2004113501/08/20041309
01/09/2004115701/09/20041342
01/09/2004135901/09/20041537
01/12/2004234201/13/20040214
01/16/2004140901/16/20041548
01/16/2004233701/17/20040109
01/18/2004113501/18/20041256
01/19/2004022301/19/20040346
01/19/2004232901/20/20040157
01/21/2004125201/21/20041400
01/22/2004005201/22/20040227
01/23/2004113701/23/20041319
01/26/2004120201/26/20040146
02/17/2004140602/17/20041536
02/26/2004122102/26/20041350
03/09/2004114503/09/20041315
03/14/2004113103/14/20041300
03/22/2004113403/22/20041300
03/29/2004233203/30/20040047
03/30/2004233403/31/20040059
04/11/2004234804/12/20040120
05/01/2004233505/02/20040110
05/02/2004233605/03/20040109
05/13/2004233405/14/20040112
05/14/2004140605/14/20041411
06/07/2004022406/07/20040351
06/25/2004114606/25/20041328
07/02/2004005907/02/20040217
07/20/2004022007/20/20040358
07/24/2004123507/24/20041355
08/22/2004114408/22/20041207
09/08/2004142509/08/20041605
09/08/2004233809/08/20042345
10/18/2004234610/19/20040128
11/29/2004232111/30/20040105
12/08/2004114312/08/20041306
12/30/2004113012/30/20041239
Subject:
TWP/SONDE/C2 - Broken temperature sensors
DataStreams:twpsondewnpnC2.b1
Description:
The temperature sensors on these radiosondes broke after launch.  All
temperature-related data are incorrect.
Measurements:twpsondewnpnC2.b1:
  • Relative humidity inside the instrument enclosure(rh)
  • Dry bulb temperature(tdry)
  • Surface dew point temperature(dp)


Back To Table of Contents

DQRID : D070920.2
Start DateStart TimeEnd DateEnd Time
03/19/2003122103/19/20031253
08/20/2003234008/21/20030002
Subject:
TWP/SONDE/C2 - Failure of RH sensor
DataStreams:twpsondewnpnC2.b1
Description:
The RH sensor ASIC that controls the sensor heating failed.  As a result the RH at the 
beginning of the sounding oscillates between near-zero and ambient values.  The RH heating 
cycle stops when the ambient temperature reaches below -30 degC; above this level RH 
values should be valid.
Measurements:twpsondewnpnC2.b1:
  • Relative humidity inside the instrument enclosure(rh)
  • Surface dew point temperature(dp)


Back To Table of Contents