Data Quality Reports for Session: 109599 User: morris Completed: 01/02/2008


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D071106.1FKB/GNDRAD/M1 - Incorrect calibration coefficient applied to up_short_hemispheric
instrument
fkbgndrad20sM1.a0, fkbgndrad60sM1.b1, fkbgndradM1.00


DQRID : D071106.1
Start DateStart TimeEnd DateEnd Time
03/16/2007000011/05/20071344
Subject:
FKB/GNDRAD/M1 - Incorrect calibration coefficient applied to up_short_hemispheric 
instrument
DataStreams:fkbgndrad20sM1.a0, fkbgndrad60sM1.b1, fkbgndradM1.00
Description:
An incorrect calibration coefficient was applied to the up_short_hemispheric instrument 
(PSP) for the specified time period.  Changes in the ARM PSP calibration procedure now 
generates two differnt calibration coefficients for the same instrument, Cf(non "IR 
corrected")and Cfc("IR corrected").  CFc should have been applied to this instrument when CF was 
used.
The affected PSP is S/N 34292F3 CF=117.61, CFc(now applied)=116.07W/m2/mV
Measurements:fkbgndrad60sM1.b1:
  • Upwelling (10 meter) Shortwave Hemispheric Irradiance, Pyranometer, Minima(up_short_hemisp_min)
  • Upwelling (10 meter) Shortwave Hemispheric Irradiance, Pyranometer, Standard
    Deviation(up_short_hemisp_std)
  • Upwelling (10 meter) Shortwave Hemispheric Irradiance, Pyranometer, Maxima(up_short_hemisp_max)
  • Upwelling (10 meter) Shortwave Hemispheric Irradiance, Pyranometer(up_short_hemisp)

fkbgndrad20sM1.a0:
  • Instantaneous Upwelling Shortwave Hemispheric Irradiance, Pyranometer(inst_up_short_hemisp)

fkbgndradM1.00:
  • Raw data stream - documentation not supported(Raw data stream - documentation not supported)


Back To Table of Contents



END OF DATA