Data Quality Reports for Session: 109487 User: srrb Completed: 12/19/2007


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D051206.5TWP/GNDRAD/C3 - IRT malfunctiontwpgndrad60sC3.b1
D060111.1TWP/GNDRAD/C3 - IRT powertwpgndrad60sC3.b1
D060426.8TWP/SMET/C3 - ORG data noisy due to degraded groundtwpsmet60sC3.b1
D060502.3TWP/SMET/C3 - Reprocess: Windspeed2 calibration errortwpsmet60sC3.b1
D060509.1TWP/GNDRAD/C3 - IRT failedtwpgndrad60sC3.b1
D060627.1TWP/SMET/C3 - Anemometer Readings Incorrecttwpsmet60sC3.b1
D060630.36TWP/GNDRAD/C3 - Reprocess: Longwave Calibration errortwpgndrad60sC3.b1
D060818.3TWP/SMET/C3 - Reprocess: Barometric Data Changed from hPa to kPatwpsmet60sC3.b1
D060823.1TWP/GNDRAD/C3 - IRT biastwpgndrad60sC3.b1
D061011.4TWP/SMET/C3 - Reprocess: Tipping bucket rain gauge addedtwpsmet60sC3.b1
D070306.2TWP/SMET/C3 - Barometer Data Incorrecttwpsmet60sC3.b1
D070306.3TWP/SMET/C3 - Data Incorrect Due to Water Intrusiontwpsmet60sC3.b1
D070803.1TWP/SMET/C3 - Upper wind speed artificially slowtwpsmet60sC3.b1


DQRID : D051206.5
Start DateStart TimeEnd DateEnd Time
11/21/2005150012/05/20050100
Subject:
TWP/GNDRAD/C3 - IRT malfunction
DataStreams:twpgndrad60sC3.b1
Description:
The internal IRT power supply failed (probably due to a lightning strike on 11/14) and was 
drawing excessive current on the data logger's DC system. The problem was corrected by 
replacing the instrument with a spare.
Measurements:twpgndrad60sC3.b1:
  • Surface Infra-Red Temperature, Standard Deviation(sfc_ir_temp_std)
  • Surface Infra-Red Temperature Minima(sfc_ir_temp_min)
  • Effective Surface (Skin) Temperature(sfc_ir_temp)
  • Surface Infra-Red Temperature Maxima(sfc_ir_temp_max)


Back To Table of Contents

DQRID : D060111.1
Start DateStart TimeEnd DateEnd Time
12/20/2005233001/02/20062300
Subject:
TWP/GNDRAD/C3 - IRT power
DataStreams:twpgndrad60sC3.b1
Description:
The IRT failed to report after a planned power outage to add lightning protection modules 
to the AC power wiring. The problem was corrected when a loose power connection to the 
logger was found, rectified, and power restored to the logger.
Measurements:twpgndrad60sC3.b1:
  • Surface Infra-Red Temperature, Standard Deviation(sfc_ir_temp_std)
  • Surface Infra-Red Temperature Minima(sfc_ir_temp_min)
  • Effective Surface (Skin) Temperature(sfc_ir_temp)
  • Surface Infra-Red Temperature Maxima(sfc_ir_temp_max)


Back To Table of Contents

DQRID : D060426.8
Start DateStart TimeEnd DateEnd Time
12/01/2005000008/24/20062359
Subject:
TWP/SMET/C3 - ORG data noisy due to degraded ground
DataStreams:twpsmet60sC3.b1
Description:
SMET Rain Data began to be noisy during December 2005 and the problem slowly became worse 
over time so that in February 2006 the ORG was reporting with a constant background value 
of around 0.10 mm/hr rain rate.  ORG voltage showed a near constant value above the 85mV 
threshold.  The ground for the ORG had become degraded and was lost due to corrosion.
Measurements:twpsmet60sC3.b1:
  • Precipitation mean(precip_mean)
  • Precipitation maximum(precip_max)
  • Precipitation standard deviation(precip_sd)
  • precipitation minimum(precip_min)


Back To Table of Contents

DQRID : D060502.3
Start DateStart TimeEnd DateEnd Time
04/27/2006233005/02/20060201
Subject:
TWP/SMET/C3 - Reprocess: Windspeed2 calibration error
DataStreams:twpsmet60sC3.b1
Description:
The lower anemometer was replaced afer being removed in anticipation of cyclone Monica. 
The calibration coefficient was not changed during this period so the wind speeds are a 
little low (0.2%).
Measurements:twpsmet60sC3.b1:
  • Wind #2 speed vector average(wind2_spd_vec_avg)
  • Wind #2 speed arithmetic average(wind2_spd_arith_avg)
  • Wind #2 speed maximum gust(wind2_spd_max_gust)
  • Wind #2 speed minimum(wind2_spd_min)
  • Wind #2 speed maximum(wind2_spd_max)
  • Wind #2 speed standard deviation(wind2_spd_sd)


Back To Table of Contents

DQRID : D060509.1
Start DateStart TimeEnd DateEnd Time
04/07/2006000005/18/20062359
Subject:
TWP/GNDRAD/C3 - IRT failed
DataStreams:twpgndrad60sC3.b1
Description:
The IRT began to fail on 4/7/2006 and failed completely (apparently due to the internal 
power supply) on 4/27. It was replaced on 5/18.
Measurements:twpgndrad60sC3.b1:
  • Surface Infra-Red Temperature, Standard Deviation(sfc_ir_temp_std)
  • Surface Infra-Red Temperature Minima(sfc_ir_temp_min)
  • Effective Surface (Skin) Temperature(sfc_ir_temp)
  • Surface Infra-Red Temperature Maxima(sfc_ir_temp_max)


Back To Table of Contents

DQRID : D060627.1
Start DateStart TimeEnd DateEnd Time
06/27/2006011506/27/20060344
Subject:
TWP/SMET/C3 - Anemometer Readings Incorrect
DataStreams:twpsmet60sC3.b1
Description:
The Anemometer readings were incorrect due to the tower being lowered for maintenance.
Measurements:twpsmet60sC3.b1:
  • Wind #2 speed arithmetic average(wind2_spd_arith_avg)
  • Wind #1 speed minimum(wind1_spd_min)
  • Wind #1 direction vector average(wind1_dir_vec_avg)
  • Wind #1 speed standard deviation(wind1_spd_sd)
  • Wind #2 speed maximum(wind2_spd_max)
  • Wind #1 speed vector average(wind1_spd_vec_avg)
  • Wind #2 direction vector average(wind2_dir_vec_avg)
  • Wind #1 direction standard deviation(wind1_dir_sd)
  • Wind #2 speed vector average(wind2_spd_vec_avg)
  • Wind #2 speed maximum gust(wind2_spd_max_gust)
  • Wind #1 speed maximum(wind1_spd_max)
  • Wind #2 speed minimum(wind2_spd_min)
  • Wind #1 speed maximum gust(wind1_spd_max_gust)
  • Wind #1 speed arithmetic average(wind1_spd_arith_avg)
  • Wind #2 direction standard deviation(wind2_dir_sd)
  • Wind #2 speed standard deviation(wind2_spd_sd)


Back To Table of Contents

DQRID : D060630.36
Start DateStart TimeEnd DateEnd Time
05/20/2004010012/21/20052137
Subject:
TWP/GNDRAD/C3 - Reprocess: Longwave Calibration error
DataStreams:twpgndrad60sC3.b1
Description:
Modified pyrgeometer calibration procedures were implemented beginning in December 2003.  
These modified procedures introduced a calibration bias in the longwave data.  The 
previous procedures were re-implemented at all sites between December 2005 and February 2006 to 
restore proper calibrations.                                                             
                The data collected while the incorrect procedures were in place are 
being reprocessing to remove the calibration bias.  Note, this also affected standard 
deviation, maximum and minimum data fields.
Measurements:twpgndrad60sC3.b1:
  • Upwelling (10 meter) Longwave Hemispheric Irradiance, Pyrgeometer(up_long_hemisp)


Back To Table of Contents

DQRID : D060818.3
Start DateStart TimeEnd DateEnd Time
12/02/2003000007/13/20062219
Subject:
TWP/SMET/C3 - Reprocess: Barometric Data Changed from hPa to kPa
DataStreams:twpsmet60sC3.b1
Description:
Barometric pressure data was converted from hPa to kPa in order to standardize the 
measurement units among ARM sites and to conform to accepted standard units determined by the 
scientific community.
Measurements:twpsmet60sC3.b1:
  • Barometric Pressure(atmos_pressure)


Back To Table of Contents

DQRID : D060823.1
Start DateStart TimeEnd DateEnd Time
05/19/2006000006/27/20060344
Subject:
TWP/GNDRAD/C3 - IRT bias
DataStreams:twpgndrad60sC3.b1
Description:
The IRT reported higher surface temperatures than the PIR due to corrosion on the 
terminals of the lightning protection module and a ground wire error. The corrosion was cleaned 
and the wiring corrected.
Measurements:twpgndrad60sC3.b1:
  • Surface Infra-Red Temperature, Standard Deviation(sfc_ir_temp_std)
  • Surface Infra-Red Temperature Minima(sfc_ir_temp_min)
  • Effective Surface (Skin) Temperature(sfc_ir_temp)
  • Surface Infra-Red Temperature Maxima(sfc_ir_temp_max)


Back To Table of Contents

DQRID : D061011.4
Start DateStart TimeEnd DateEnd Time
03/12/2002000009/26/20061435
Subject:
TWP/SMET/C3 - Reprocess: Tipping bucket rain gauge added
DataStreams:twpsmet60sC3.b1
Description:
A tipping bucket rain gauge was added to the SMET suite of instruments effective 20060926. 
 Two new variables (count and rain amount) were added at the end of the twpsmet60sC1.b1 
datastream.  When reprocessing of this datastream occurs these variables will be added 
with -9999 (missing) as the value for dates prior to gauge installation.
Measurements:twpsmet60sC3.b1:
  • base time(base_time)


Back To Table of Contents

DQRID : D070306.2
Start DateStart TimeEnd DateEnd Time
03/02/2007172303/03/20071228
Subject:
TWP/SMET/C3 - Barometer Data Incorrect
DataStreams:twpsmet60sC3.b1
Description:
The barometer data was incorrect or missing due to water intrusion.
Measurements:twpsmet60sC3.b1:
  • Barometric Pressure(atmos_pressure)


Back To Table of Contents

DQRID : D070306.3
Start DateStart TimeEnd DateEnd Time
03/03/2007005603/03/20070123
Subject:
TWP/SMET/C3 - Data Incorrect Due to Water Intrusion
DataStreams:twpsmet60sC3.b1
Description:
Data became erratic and unreasonable after water intrusion into the enclosure.
Measurements:twpsmet60sC3.b1:
  • Vapor pressure standard deviation(vappress_sd)
  • Precipitation maximum(precip_max)
  • Relative humidity maximum(relh_max)
  • Mean Relative Humidity(relh_mean)
  • Std Dev of Relative Humidity or Hardware Error Code(relh_sd)
  • Relative humidity minimum(relh_min)
  • precipitation minimum(precip_min)
  • Precipitation mean(precip_mean)
  • Temperature minimum(temp_min)
  • Mean Air Temperature or Hardware Error(temp_mean)
  • Temperature maximum(temp_max)
  • Precipitation standard deviation(precip_sd)
  • Vapor pressure mean(vappress_mean)
  • Std Dev of Air Temperature(temp_sd)


Back To Table of Contents

DQRID : D070803.1
Start DateStart TimeEnd DateEnd Time
07/01/2007000009/18/20072315
Subject:
TWP/SMET/C3 - Upper wind speed artificially slow
DataStreams:twpsmet60sC3.b1
Description:
The upper wind measurement was consistently slower by around 0.5 - 1 m/s than the lower 
wind measurement.  The bearings on the upper wind measurement were binding.
Measurements:twpsmet60sC3.b1:
  • Wind #1 speed minimum(wind1_spd_min)
  • Wind #1 speed maximum(wind1_spd_max)
  • Wind #1 speed standard deviation(wind1_spd_sd)
  • Wind #1 speed maximum gust(wind1_spd_max_gust)
  • Wind #1 speed vector average(wind1_spd_vec_avg)


Back To Table of Contents