Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/13/2006 | 0135 | 12/04/2006 | 2120 |
Subject: | TWP/SKYRAD/C2 - IRT signal |
DataStreams: | twpskyrad60sC2.b1 |
Description: | The IRT exhibited a bias of the analog voltage output that caused problems with the signals from the other SKYRAD radiometers. The bias was due to corrosion between the power and signal connections in the IRT cable. The IRT was disconnected on 10/18/2006 at 01:20 and replaced on 11/6/2006 at 23:00 but it too appeared to have grounding problems due to internal corrosion and was disconnected on 11/11/2006 at 11:42 and then intermittently reconnected for testing. A second spare was installed on 12/4/2006 which did not cause problems with the signals from the other instruments. |
Measurements: | twpskyrad60sC2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/31/2007 | 0223 | 02/10/2007 | 0055 |
Subject: | TWP/SKYRAD/C2 - Short direct normal |
DataStreams: | twpskyrad20sC2.a0, twpskyrad60sC2.b1 |
Description: | The short_direct_normal (NIP) data are missing or incorrect due to instrument cable failure. The cable was replaced. The short_direct_normal can be calculated using the following relationship for shortwave radiometers: global hemispheric = diffuse hemispheric + direct normal *cos (zenith angle) |
Measurements: | twpskyrad60sC2.b1:
twpskyrad20sC2.a0:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/04/2006 | 2120 | 02/22/2007 | 0125 |
Subject: | TWP/SKYRAD/C2 - IRT calibration |
DataStreams: | twpskyrad60sC2.b1 |
Description: | The IRT was not properly calibrated and was not able to measure sky temperatures below 273.2 K. (This instrument was the same one damaged by lightning in November 2005 in Darwin. It was returned for repair and received a new lens but was calibrated only down to 0 deg C, so the problem was neither detected nor repaired.) The IRT was replaced on 2/22/2007. |
Measurements: | twpskyrad60sC2.b1:
|