Data Quality Reports for Session: 105418 User: clayton Completed: 04/26/2007


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D050725.1SGP/SONDE/C1 - Bad RH DatasgpsondewnpnC1.b1
D050811.5SGP/SONDE/C1 - Bad T/RH and surface P.sgpsondewnpnC1.b1
D060928.3SGP/SONDE/C1 - Increased precision for alt, dp, pres, rh, tdrysgpsondewnpnC1.b1


DQRID : D050725.1
Start DateStart TimeEnd DateEnd Time
07/17/2005234707/18/20050141
Subject:
SGP/SONDE/C1 - Bad RH Data
DataStreams:sgpsondewnpnC1.b1
Description:
The RH values below 6 km are incorrect.  The problem relates to a failure of the GC25 
ground check set used before this particular sounding.  The failure seems to have affected 
the sensor calibration with the effect that one of the two sensors was reading artificially 
high.  Because the internal processing software uses the highest of the two sensor 
readings, the bad sensor value is chosen whenever it is not heated.
Measurements:sgpsondewnpnC1.b1:
  • Surface dew point temperature(dp)
  • Relative humidity scaled, by total column amount from MWR(rh)


Back To Table of Contents

DQRID : D050811.5
Start DateStart TimeEnd DateEnd Time
08/01/2005113208/01/20050116
Subject:
SGP/SONDE/C1 - Bad T/RH and surface P.
DataStreams:sgpsondewnpnC1.b1
Description:
Apparently multiple failures.  The radiosonde temperature sensor broke on launch causing 
an approximate 35 degC negative offset and as a result, the RH is incorrect.  Furthermore, 
it appears that the pressure sensor was off by approximately 4 hPa, though this would 
not affect the altitude calculation, the absolute pressure is wrong.
Measurements:sgpsondewnpnC1.b1:
  • Dry-Bulb Temperature(tdry)
  • Relative humidity scaled, by total column amount from MWR(rh)
  • pressure at constant pressure surface(pres)


Back To Table of Contents

DQRID : D060928.3
Start DateStart TimeEnd DateEnd Time
09/28/2006172812/31/99992359
Subject:
SGP/SONDE/C1 - Increased precision for alt, dp, pres, rh, tdry
DataStreams:sgpsondewnpnC1.b1
Description:
The output precision of the following variables was increased 20060928

alt  changed from 1m to 0.1m
dp   changed from 0.1degC to 0.01degC
pres changed from 0.1 hPa to 0.01 hPa
RH   changed from 1%RH to 0.01%RH
tdry changed from .1degC to 0.01degC

Note that the increase in precision does not necessarily imply an increase in accuracy.  
Users are cautioned to continue to use their best scientific judgement when using these or 
any other data.
Measurements:sgpsondewnpnC1.b1:
  • Surface dew point temperature(dp)
  • Dry-Bulb Temperature(tdry)
  • Relative humidity scaled, by total column amount from MWR(rh)
  • pressure at constant pressure surface(pres)
  • altitude above sea levelaltunits(alt)


Back To Table of Contents



END OF DATA