DQRID : D051104.3
Start DateStart TimeEnd DateEnd Time
08/17/2005223010/05/20051600
Subject:
SGP/EBBR/E13 - Loose connections
DataStreams:sgp5ebbrE13.b1, sgp15ebbrE13.b1, sgp30ebbrE13.b1
Description:
The tair_bot spiked around midday, as if it is temperature related, and may have indicated 
a loose connection.

The sensible and latyent heat fluxes and Bowen ratio were incorrect when this condition 
occurred.
Measurements:sgp15ebbrE13.b1:
  • Left air temperature(tair_l)
  • Right air temperature(tair_r)

sgp5ebbrE13.b1:
  • bottom air temperature(tair_bot)

sgp30ebbrE13.b1:
  • bottom air temperature(tair_bot)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • latent heat flux(e)
  • corrected sensible heat flux(h)


Back To Table of Contents

DQRID : D060630.8
Start DateStart TimeEnd DateEnd Time
12/11/2003195002/16/20061901
Subject:
SGP/SIRS/C1 - Reprocessed: Longwave Calibration error
DataStreams:sgpsirsC1.b1
Description:
Modified pyrgeometer calibration procedures were implemented beginning in December 2003. 
These modified procedures introduced a calibration bias in the longwave data. The previous 
procedures were re-implemented at all sites between December 2005 and February 2006 to 
restore proper calibrations. 

The data collected while the incorrect procedures were in place have been reprocessed to 
remove the calibration bias.  The reprocessed 60 second averaged data are based on 3 
instantaneous 20 second data records rather than on 60 1 second instantaneous data records. 
Still, these data are considered far superior to the originally processed data. The 
reprocessed data were archived in October 2006.
Measurements:sgpsirsC1.b1:
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer(down_long_hemisp_shaded)
  • Downwelling Longwave Hemispheric Net Infrared(down_long_netir)
  • Upwelling (10 meter) Longwave Hemispheric Irradiance, Ventilated Pyrgeometer(up_long_hemisp)
  • Upwelling Longwave Hemispheric Net Infrared(up_long_netir)


Back To Table of Contents

DQRID : D060928.3
Start DateStart TimeEnd DateEnd Time
09/28/2006172812/31/99992359
Subject:
SGP/SONDE/C1 - Increased precision for alt, dp, pres, rh, tdry
DataStreams:sgpsondewnpnC1.b1
Description:
The output precision of the following variables was increased 20060928

alt  changed from 1m to 0.1m
dp   changed from 0.1degC to 0.01degC
pres changed from 0.1 hPa to 0.01 hPa
RH   changed from 1%RH to 0.01%RH
tdry changed from .1degC to 0.01degC

Note that the increase in precision does not necessarily imply an increase in accuracy.  
Users are cautioned to continue to use their best scientific judgement when using these or 
any other data.
Measurements:sgpsondewnpnC1.b1:
  • Surface dew point temperature(dp)
  • Dry-Bulb Temperature(tdry)
  • Relative humidity scaled, by total column amount from MWR(rh)
  • pressure at constant pressure surface(pres)
  • altitude above sea levelaltunits(alt)