MAO/CCN/M1 - Instrument undercounting of lowest super saturation values
DataStreams:maoaosccn100M1.a1
Description:
The slow temperature decay of the OPC induced droplet evaporation and caused an under 
counting of the droplets at the lowest %ss, 0.15.
Suggestions: 
Use the next highest setting of 0.25 %ss.
Measurements:maoaosccn100M1.a1:
  • AOS CCN sample saturation setpoint value reported by instrument(CCN_ss_set)