Subject:
NSA/MET/C1 - 2m RH values incorrect
DataStreams:nsatwrC1.b1, nsametC1.b1
Description:
The new t/rh probe installed during annual calibrations and checks began to exhibit 
readings that were biased high compared to the Chilled Mirror Hygrometer.  In high RH 
conditions the sensor would consistently report RH values in excess of 105%.  Comparisons to the 
other tower levels also suggested that the 2m level RH data is suspect.
Suggestions: 
Use the co-located Chilled Mirror Hygrometer data.
Measurements:nsametC1.b1:
  • Standard deviation in logger enclosure relative humidity(rh_std)
  • Vapor pressure standard deviation(vapor_pressure_std)
  • Quality check results on field: Dew point mean, calculated(qc_dew_point_mean)
  • Dew point mean, calculated(dew_point_mean)
  • Quality check results on field: Relative humidity mean(qc_rh_mean)
  • Dew point standard deviation(dew_point_std)
  • Vapor pressure mean, calculated(vapor_pressure_mean)
  • Relative humidity observed by the Raman lidar(rh_mean)
more
nsatwrC1.b1:
  • Relative humidity observed by the Raman lidar(rh_mean)
  • Quality check results on field: Dew point mean, calculated(qc_dew_point_mean)
  • Quality check results on field: Vapor pressure mean, calculated(qc_vapor_pressure_mean)
  • Dew point standard deviation(dew_point_std)
  • Vapor pressure standard deviation(vapor_pressure_std)
  • Dew point mean, calculated(dew_point_mean)
  • Standard deviation in logger enclosure relative humidity(rh_std)
  • Vapor pressure mean, calculated(vapor_pressure_mean)
  • Quality check results on field: Relative humidity mean(qc_rh_mean)
more

Close this window