TWP/VCEIL/C1 - Reprocess: Incorrect units on installed ceilometer
DataStreams:twpvceil25kC1.a1, twpvceil25kC1.b1
Description:
The units setting on the ceilometer installed on 18 Jan. 2002 were incorrectly set to feet 
rather than meters.
Suggestions: 
The units are being reset. All heights should be multiplied by 0.3048 during this period.
Measurements:twpvceil25kC1.a1:
  • Second lowest cloud base height(second_cbh)
  • Lowest cloud base height detected(first_cbh)
  • Third cloud base height(third_cbh)
  • Aerosol volume backscattering coefficient at 355 nm(backscatter)
more
twpvceil25kC1.b1:
  • Lowest cloud base height detected(first_cbh)
  • Third cloud base height(third_cbh)
  • Aerosol volume backscattering coefficient at 355 nm(backscatter)
  • Second lowest cloud base height(second_cbh)