Subject: | SGP/MWR/B1 out of calibration |
DataStreams: | sgpmwrlosB1.a1, sgpmwrlosB1.a0
|
Description: | A shift in the calibration of MWR/B1 occurred on 10 March 1998,
apparently due to a power failure (see P980513.1). The calibration
was corrected on 19 May. |
Suggestions: | Reprocess the data with the correct calibration coefficients. |
Measurements: | sgpmwrlosB1.a1: - lon
- actel
- wet_window
- actaz
- ir_temp
- alt
- 31tbsky
- 23tbsky
- vap
- time_offset
- lat
- liq
- base_time
more
sgpmwrlosB1.a0: - 31skyn
- base_time
- 23skyn
- time_offset
- 31unoise
- wet_window
- 23bb
- lon
- tkxc
- 31sky
- 23unoise
- 31bbn
- actel
- 31tbsky
- 23bbn
- 23tbsky
- ir_temp
- lat
- tkbb
- liq
- losn
- 31bb
- 23sky
- vap
- actaz
- alt
more
|