Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
10/21/1996 | 1803 | 08/27/1997 | 2359 | Incorrect |
Subject: | TWP/MFRSR/C1 - Calibration Drift |
DataStreams: | twpmfrsrC1.a0, twpmfrsrC1.b1 |
Description: | Through this period we see significant drift in all of the MFRSR channels except the 862 nm channel. The most serious drift occurred in the 610 and 665 nm channels which suffered a loss in sensitivity of a factor of 5 or more over the affected period. The MFRSR head was replaced on August 27, 1997 during a RESET visit. |
Suggestions: | |
Measurements: | twpmfrsrC1.a0: more
twpmfrsrC1.b1: more |