Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
09/12/1997 | 0000 | 09/26/1997 | 2047 | Incorrect |
Subject: | SGP/TWR/C1 - SE 25m RH Incorrect |
DataStreams: | sgp1twr10xC1.a1, sgp30twr10xC1.b1, sgp30twr10xC1.a1, sgp1twr10xC1.b1 |
Description: | The RH and therefore vapor pressure measurements on the 25 meter level, southeast side of the 60 m tower (data from CR10X) were incorrect because the relative humidity portion of the T/RH probe malfunctioned. It is unknown at what day and time the failure occurred (it should be obvious from the data), although it was in the middle of the Fall combined IOP; I was not receiving the data at the time. The T/RH probe was replaced on 26 September. |
Suggestions: | |
Measurements: | sgp30twr10xC1.a1: more
sgp30twr10xC1.b1: more sgp1twr10xC1.a1: more sgp1twr10xC1.b1: more |