Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
03/17/1994 | 1600 | 03/17/1994 | 1630 | Incorrect |
Subject: | SGP/EBBR/E20 - Pressure Sensor Malfunction |
DataStreams: | sgp5ebbrE20.b1, sgp30ebbrE20.b1, sgp15ebbrE20.b1 |
Description: | Pressure sensor malfunction during attempt to monitor it during Preventative Maintenance visit. 30 minute values of pressure for 1600 GMT and 1630 GMT, 15 minute values for 1600, 1615, and 1630, and 5 minute values for 1600, 1605, 1610, 1615, 1620, 1625, and 1630 GMT are incorrect. However, the range of the incorrect values was such as to not adversely affect flux measurements. Some noise was apparently induced in the pressure output voltage as a result of the monitoring, although it is not known how. |