Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
04/14/2017 | 1557 | 04/14/2017 | 1559 | Incorrect |
Subject: | OLI/CPC/M1 - Instrument switch out |
DataStreams: | oliaoscpcfM1.00, oliaoscpcfM1.b1, oliaoscpcfM1.a1 |
Description: | During this time period, the CPCF was switched with the CPCU. The data during this time period was collected from a different model CPC with different sampling conditions and size cutoffs. |
Suggestions: | Don't use data |
Measurements: | oliaoscpcfM1.b1: more
oliaoscpcfM1.a1: more oliaoscpcfM1.00: more |