Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
01/11/2017 | 0400 | 01/14/2017 | 1739 | Suspect |
02/04/2017 | 2310 | 02/04/2017 | 0509 | Suspect |
12/29/2016 | 0835 | 12/29/2016 | 1134 | Suspect |
02/09/2017 | 0000 | 02/24/2017 | 2159 | Suspect |
Subject: | SGP/SMPS/E13 - Impactor pressure elevated |
DataStreams: | sgpaossmpsE13.00, sgpaossmpsE13.a1 |
Description: | The impactor flow rate was reading higher values due to an obstruction in the impactor orifice. This affects the concentrations reported. The data is likely recoverable by re-inversion of the aerosol size distribution with the corrected flow rates. Data during these times should be used for qualitative purposes only. |
Suggestions: | Use with caution |
Measurements: | sgpaossmpsE13.a1: more
sgpaossmpsE13.00: more |