Subject: | SGP/CO2FLX/C1 - Incorrect raw file header |
DataStreams: | sgp30co2flx4mC1.a1, sgpco2flx4mC1.00, sgpco2flx60mC1.00, sgpco2flxsoilC1.00, sgp30co2flx60mC1.a1, sgpco2flxwind60mC1.a1, sgp30co2flx25mC1.00, sgpco2flx25mC1.b1, sgpco2flx4mC1.b1, sgpco2flx60mC1.b1, sgpco2flxsoilauxC1.b1, sgpco2flxradC1.00, sgpco2flxwind25mC1.a1, sgpco2flx25mC1.00, sgpco2flxsoilC1.b1, sgp30co2flx25mC1.b1, sgpco2flxrad4mC1.b1, sgpco2flxwind4mC1.a1, sgp30co2flx4mC1.b1, sgp30co2flx4mC1.00, sgp30co2flx25mC1.a1, sgp30co2flx60mC1.00, sgpco2flxgas4mC1.a1, sgp30co2flx4mmetC1.b1, sgp30co2flx60mC1.b1 |
Description: | An ingest failure caused by a change in the header of the instrument output file generated by EddyPro resulted in missing data. This is a result of a bug in the EddyPro software. This only occurs when the all sensor (e.g., gas analyzer) data is screened/flagged for the first averaging period of the processing run. This should be a rare occurrence. |
Suggestions: | |
Measurements: | sgpco2flx60mC1.b1: more
sgpco2flx25mC1.00: more sgp30co2flx4mC1.b1: more sgpco2flxwind4mC1.a1: more sgp30co2flx25mC1.a1: more sgp30co2flx60mC1.a1: more sgp30co2flx4mmetC1.b1: more sgpco2flxrad4mC1.b1: more sgpco2flxsoilauxC1.b1: more sgpco2flx4mC1.00: more sgpco2flxradC1.00: more sgp30co2flx4mC1.00: more sgpco2flxsoilC1.b1: more sgp30co2flx25mC1.00: more sgp30co2flx60mC1.00: more sgpco2flx4mC1.b1: more sgpco2flxsoilC1.00: more sgp30co2flx4mC1.a1: more sgp30co2flx25mC1.b1: more sgp30co2flx60mC1.b1: more sgpco2flx60mC1.00: more sgpco2flx25mC1.b1: more sgpco2flxgas4mC1.a1: more sgpco2flxwind25mC1.a1: more sgpco2flxwind60mC1.a1: more |