Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
07/08/2005 | 2300 | 09/13/2005 | 2100 | Suspect |
Subject: | PYE/AIP/M1 - Bad T/RH values from thermal gradients |
DataStreams: | pyeaipfitrh1ogrenM1.c1, pyeaipfitrh1ogrenM1.s1 |
Description: | Due to the thermal inertia of the instrument wall, slow signal response and a radial temperature gradient inside the instrument cavity, the reported RH and temperature inside the wet nephelometer are off. The radial temperature gradients need to be characterized inside the instrument at several locations within the instrument with respect to a calibrated Vaisala T and RH sensor at the instrument exit. The resulting fRH VAP needs to be recalculated and analyzed. |
Suggestions: | Do not use data. Data may be reposted in another data archive. Check with mentor. |
Measurements: | pyeaipfitrh1ogrenM1.c1: more
pyeaipfitrh1ogrenM1.s1: more |