Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
06/19/2015 | 0200 | 07/07/2015 | 1600 | Suspect |
Subject: | MAO/AOS/S1 - Blocked 1 micron impactor state |
DataStreams: | maoaosimpactorS1.a1 |
Description: | A bug or clump of dirt stuck on the inlet of the 1 micron impactor. This increased the air flow restriction in the 1 micron impactor, changing the cut off point. |
Suggestions: | The 10 micron data is correct. The 1 micron is incorrect. Do not use 1 micron data. |
Measurements: | maoaosimpactorS1.a1: more
|