Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
09/17/2015 | 1934 | 09/17/2015 | 2035 | Incorrect |
Subject: | MAO/ACSM/AETH/CPC/NEPHELOMETER/PSAP/SMPS/UHSAS/S1 Calibration of the Nephelometer |
DataStreams: | maoaosnephdryS1.a1, maoaosaethS1.a1, maoaospsap3wS1.b1, maoaosnephwetS1.b1, maoaospsap3wS1.01, maoaosuhsasS1.a1, maoaoscpcuS1.a1, maoaosaethS1.00, maoaossmpsS1.a1, maoaoscpcfS1.a1, maoaosacsmS1.a1, maoaossmpstrhS1.a1, maoaospsap3wS1.a1, maoaosnephdryS1.b1 |
Description: | Inlet line for these instruments was used during Nephelometer calibration compromising the data. |
Suggestions: | Do not use this data |
Measurements: | maoaosuhsasS1.a1: more
maoaossmpstrhS1.a1: more maoaossmpsS1.a1: more maoaospsap3wS1.01: more maoaospsap3wS1.b1: more maoaosaethS1.a1: more maoaosnephdryS1.b1: more maoaosacsmS1.a1: more maoaosnephdryS1.a1: more maoaosnephwetS1.b1: more maoaoscpcfS1.a1: more maoaospsap3wS1.a1: more maoaoscpcuS1.a1: more maoaosaethS1.00: more |