Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
04/29/2015 | 1700 | 05/05/2015 | 1529 | Missing |
04/10/2015 | 2000 | 04/13/2015 | 1509 | Missing |
04/25/2015 | 1700 | 04/27/2015 | 1429 | Missing |
05/26/2015 | 1600 | 07/18/2015 | 2359 | Missing |
05/15/2015 | 2000 | 05/19/2015 | 1559 | Missing |
03/28/2015 | 1800 | 03/30/2015 | 1459 | Missing |
Subject: | SGP/IRSI/C1 - CPU overheated |
DataStreams: | sgpirsiirC1.b1, sgpirsivisC1.b1 |
Description: | The CPU intermittently overheated, interrupting the acquisition and processing of images. This caused periods of missing data, generally from early afternoon until late morning (restarted during daily rounds). A cooling unit was added to the system on 6/8 and the IRSI was returned to the vendor for repair on 6/19. The IRSI was re-installed on 7/19. |
Suggestions: | |
Measurements: | sgpirsiirC1.b1: more
sgpirsivisC1.b1: more |