Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
02/17/2015 | 0000 | 04/29/2015 | 2246 | Suspect |
Subject: | SGP/AERI/C1 - Cryocooler failure |
DataStreams: | sgpaerich2C1.a2, sgpaeriC1.00, sgpaeri01ch2C1.a1, sgpaeri01C1.00, sgpaerixC1.00, sgpaerisummaryC1.a1, sgpaeriengineerC1.b1, sgpaerisummaryC1.a2, sgpaeri01engineerC1.a1, sgpaerich1C1.b1, sgpaeri01summaryC1.a1, sgpaerisummaryC1.b1, sgpaerich1C1.a1, sgpaeri01ch1C1.a1, sgpaerich2C1.b1, sgpaerich2C1.a1, sgpaerich1C1.a2 |
Description: | During this time period the cryocooler began to fail and was eventually replaced. |
Suggestions: | During this time period check the detector temperature. Data collected with temperatures above 90k are likely unusable especially in the channel 1 long wave data. |
Measurements: | sgpaerixC1.00: more
sgpaerich1C1.b1: more sgpaerich1C1.a1: more sgpaerich1C1.a2: more sgpaeri01ch2C1.a1: more sgpaeri01engineerC1.a1: more sgpaeriC1.00: more sgpaerisummaryC1.a1: more sgpaerisummaryC1.b1: more sgpaerisummaryC1.a2: more sgpaerich2C1.b1: more sgpaeri01ch1C1.a1: more sgpaeri01summaryC1.a1: more sgpaeriengineerC1.b1: more sgpaerich2C1.a2: more sgpaerich2C1.a1: more sgpaeri01C1.00: more |