Subject: | SGP/RL/C1 - Alignment malfunction resulting in poor signal-to-noise ratio |
DataStreams: | sgprlC1.a0, sgprlC1.00
|
Description: | The signal-to-noise ratio was poor during the effected period due to a malfunction of the
alignment system. This was probably caused by a problem in the stepper motor that
controls the beam steering mirror. |
Suggestions: | Users should avoid using data during this period. |
Measurements: | sgprlC1.00: - Raw data stream - documentation not supported
more
sgprlC1.a0: - t2_analog_high
- water_counts_low
- water_analog_low
- depolarization_analog_high
- nitrogen_counts_high
- liquid_analog_high
- t2_counts_high
- depolarization_counts_high
- elastic_counts_high
- water_counts_high
- t1_counts_high
- nitrogen_analog_low
- water_analog_high
- elastic_analog_high
- nitrogen_analog_high
- liquid_counts_high
- elastic_counts_low
- elastic_analog_low
- t1_analog_high
- nitrogen_counts_low
more
|