Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
01/25/2014 | 0000 | 03/06/2015 | 1619 | Incorrect |
Subject: | MAO/CCN/M1 - Instrument undercounting of lowest super saturation values |
DataStreams: | maoaosccn100M1.a1 |
Description: | The slow temperature decay of the OPC induced droplet evaporation and caused an under counting of the droplets at the lowest %ss, 0.15. |
Suggestions: | Use the next highest setting of 0.25 %ss. |
Measurements: | maoaosccn100M1.a1: more
|