Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
05/28/2014 | 1844 | 07/16/2014 | 1602 | Incorrect |
Subject: | SGP/MFRSR/E34 - Heater Problem |
DataStreams: | sgpmfrsrE34.b1, sgpmfrsrE34.a0 |
Description: | During the period covered by this DQR there is an intermittent problem with the head temperature. Typically this problem has been associated with a failing, or failed, heater controller board. In this case, however, it appears there was an issue with the signal cable. When that was replaced, the problem went away. |
Suggestions: | Most of the data covered by this DQR are likely fine. It would straight forward to remove those instances when the heater was not functioning properly. Consult the mentor with any questions. |
Measurements: | sgpmfrsrE34.b1: more
sgpmfrsrE34.a0: more |