Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
02/16/2014 | 1400 | 02/24/2014 | 1300 | Suspect |
Subject: | MAO/CPC/M1 - Low CPC signal |
DataStreams: | maoaoscpcM1.a1 |
Description: | The Condensation Particle Counter (CPC) was clogged and needed cleaning. After cleaning signal was still low with a leak in the CPC sample line. |
Suggestions: | Instrument measured low but still tracking correctly. Use with caution. |
Measurements: | maoaoscpcM1.a1: more
|