Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
02/10/2014 | 1309 | 03/01/2014 | 0000 | Does not affect quality |
Subject: | MAO/TRACEGAS/S1 - NOy spurious signals |
DataStreams: | maoaosnoxS1.a1 |
Description: | NOy MFM (and the Pressure sensor) began to report spurious signals. This appears related to water infiltration. The Operator installed a 'beer can' rain diverter in early March to remedy the problem. The loss of this sensor is inconsequential as the flow is controlled and measured by an upstream Mass Flow Controller in the inlet. |
Suggestions: | |
Measurements: | maoaosnoxS1.a1: more
|