Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
12/20/2013 | 1200 | 01/14/2014 | 2000 | Suspect |
Subject: | SGP/SASHE/C1 - Temperature controller disabled |
DataStreams: | sgpsashemfrC1.b1, sgpsashevisC1.b1, sgpsashevishisunC1.a0, sgpsashenirC1.b1, sgpsashenirhisunC1.a0 |
Description: | After servicing of the SASHe on Dec 20, 2013 a temperature controller was left in a stand-by mode. The situation was corrected slightly before 20:00 UT on Jan 14, 2014. Both the "vis" and "nir" spectrometers have some potential for temperature sensitivity but the impact should not be severe given that the system is housed within a climate controlled building and the temperature gradients are small. |
Suggestions: | The MFRSR, NIMFR, and RSS are possible substitutes for the SASHe measurement. |
Measurements: | sgpsashemfrC1.b1: more
sgpsashevisC1.b1: more sgpsashenirhisunC1.a0: more sgpsashenirC1.b1: more sgpsashevishisunC1.a0: more |