Subject: | SGP/AOS/C1 - PSAP data removed due to high RH |
DataStreams: | sgpnoaaaosavgC1.b1
|
Description: | PSAP data was removed from b1 level data and VAPS because of high noise and negative
values from high RH on filter. |
Suggestions: | Use the CLAP3w data as this instrument is heated and less susceptible to water
condensation on the filter substrate. |
Measurements: | sgpnoaaaosavgC1.b1: - Bbs_R_Dry_1um_Neph3W_1
- Ba_G_Dry_1um_PSAP1W_1_std
- Bs_B_Dry_10um_Neph3W_1
- Bbs_B_Dry_10um_Neph3W_1
- Bbs_B_1um_PSAP1W_1_std
- Ba_G_Dry_10um_PSAP1W_1_N
- Bs_G_Dry_1um_Neph3W_1
- Bs_G_Dry_10um_PSAP1W_1_std
- Bs_B_Dry_1um_PSAP1W_1_N
- Bbs_G_Dry_1um_Neph3W_1
- Bs_G_1um_PSAP1W_1_std
- Bbs_G_Dry_1um_PSAP1W_1_N
- Bs_G_Dry_10um_Neph3W_1
- Bbs_G_Dry_10um_PSAP1W_1_std
- Bbs_B_Dry_10um_PSAP1W_1_std
- Bs_G_Dry_10um_PSAP1W_1_N
- Bs_B_1um_PSAP1W_1_std
- Ba_G_Dry_1um_PSAP1W_1_N
- Bbs_R_1um_PSAP1W_1_std
- Bs_B_Dry_10um_PSAP1W_1_N
- Bbs_R_Dry_10um_Neph3W_1
- Bs_R_Dry_10um_PSAP1W_1_std
- Bbs_R_Dry_10um_PSAP1W_1_N
- Bbs_R_Dry_10um_PSAP1W_1_std
- Bbs_B_Dry_1um_PSAP1W_1_N
- Ba_G_Dry_1um_PSAP1W_1
- Bs_B_Dry_10um_PSAP1W_1_std
- Bs_R_Dry_1um_Neph3W_1
- Bs_R_1um_PSAP1W_1_std
- Bbs_B_Dry_1um_Neph3W_1
- Bbs_R_Dry_1um_PSAP1W_1_N
- Bs_R_Dry_1um_PSAP1W_1_N
- Bs_G_Dry_1um_PSAP1W_1_N
- Bbs_G_1um_PSAP1W_1_std
- Bs_R_Dry_10um_Neph3W_1
- Ba_G_Dry_10um_PSAP1W_1
- Bs_B_Dry_1um_Neph3W_1
- Ba_G_Dry_10um_PSAP1W_1_std
- Bbs_G_Dry_10um_Neph3W_1
- Bbs_B_Dry_10um_PSAP1W_1_N
- Bbs_G_Dry_10um_PSAP1W_1_N
- Bs_R_Dry_10um_PSAP1W_1_N
more
|