Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
10/26/2012 | 0000 | 09/19/2013 | 1500 | Incorrect |
Subject: | SGP/IRT/C1 - Degraded hardware |
DataStreams: | sgpirt25mC1.b1 |
Description: | The surface temperature from the 25m IRT was ~10K less than measured from the PIR. The lens was replaced on 2/8/2013 but there was still a bias. The problem was corrected when the IRT (and that at 10m) was upgraded to a newer model. It may be possible to determine the calibration offset, using the PIR as reference, and correct the surface temperature measurements. |
Suggestions: | |
Measurements: | sgpirt25mC1.b1: more
|