Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
07/17/2012 | 0000 | 06/24/2013 | 1500 | Incorrect |
Subject: | PVC/NEPHELOMETER/M1 - Incorrect dryneph RH data |
DataStreams: | pvcaosnephdryM1.a1 |
Description: | The RH sensor inside the dry nephelometer was found to be off by as much as 10%. The RH_nephvol_dry was calculated in the edited b1 data from the internal T_nephvol_dry and the dew point at the sample inlet. |
Suggestions: | Use corrected pvcnoaaaosM1.b1 data |
Measurements: | pvcaosnephdryM1.a1: more
|