Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
01/21/2013 | 0000 | 07/13/2013 | 2359 | Suspect |
Subject: | TWP/CSPHOT/C1 - Obstruction in collimator tubes |
DataStreams: | twpcsphotalmC1.a1, twpcsphotppC1.a1, twpcsphotaotfiltC1.a1, twpcsphotaotC1.a1 |
Description: | During this time period, data were suspect due to obstructions in the collimator tubes. The instrument was also due for calibration. |
Suggestions: | Use the quality assured level 2.0 AOT data from AERONET when available(twpcsphotaotfiltqaC1.a1). Level 2.0 data and post-calibration level 1.5 data should become available in 2014. Can also use MFRSR data if available. |
Measurements: | twpcsphotalmC1.a1: more
twpcsphotaotC1.a1: more twpcsphotppC1.a1: more twpcsphotaotfiltC1.a1: more |