Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
11/27/2012 | 0311 | 06/01/2013 | 2036 | Suspect |
Subject: | TWP/CEIL/C2 - poor receiver |
DataStreams: | twpceilC2.b1 |
Description: | The receiver optical cross-talk compensation was poor and could not be adjusted. It appeared the receiver had failed so the instrument was replaced. |
Suggestions: | |
Measurements: | twpceilC2.b1: more
|