Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
03/06/1999 | 0000 | 03/07/1999 | 0900 | Suspect |
02/01/2001 | 0000 | 03/15/2001 | 0000 | Suspect |
Subject: | TWP/QCRAD/C2 - Air temperature noisy or bad |
DataStreams: | twpqcrad1longC2.s2, twpqcrad1longC2.s1, twpqcrad1longC2.c1, twpqcrad1longC2.c2 |
Description: | Air temperature is bad or noisy during this time. Air temperature data is needed for tests on LW up and LW down pyrgeometer case and dome temperatures, and tests on the irradiances from upwelling LW/SW and downwelling LW. There were no instrument DQRs on this problem at the time this report was submitted. |
Suggestions: | Use caution with LW up and down pyrgeometer irradiance, dome temperature, case temperature, SW up data and diffuse irradiance data because all quality control checks could not be run. |
Measurements: | twpqcrad1longC2.c2: more
twpqcrad1longC2.s2: more twpqcrad1longC2.s1: more twpqcrad1longC2.c1: more |