Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
01/01/2010 | 0000 | 06/02/2010 | 1800 | Incorrect |
Subject: | GRW/AOS/M1 - CCN signal decline |
DataStreams: | grwaosccnM1.a1 |
Description: | There was a slow decline in the CCN signal from January to June of 2010 at GRW. The reason for this decline is unknown as all the instrument operating parameters appear normal. The decline was likely from a change in the CCN column thermal properties or a clogging of the detector inlet. |
Suggestions: | Best not to use the data. |
Measurements: | grwaosccnM1.a1: more
|